CO2 Sensor
- Nondispersive infrared (NDIR) technology measures CO2 levels in ppm. May be used as a handheld portable (internal, rechargeable battery) or wall-mounted for continous recording and data logging with a WatchDog Data Logger (one CO2 meter per logger).
- Includes 6-volt DC transformer. Range: 0-4,000 ppm. Accuracy: +/- 5% of reading or 50 ppm, whichever is greater.
- Spectrum CO2 Sensor for connection with WatchDog Weather Stations
| General Specification | |
|---|---|
| Method | Dual Beam Absorption Infrared™ |
| Display-LCD | Independent CO2 and Temperature readings. Calculates and Displays Ventilation Rates |
| Sample Method | Diffusion or flow through (50 - 100 ml/min) |
| Operating Conditions | 32-122°F (0-50°C) 0-95% RH, non-condensing |
| Storage Temperatures | -4 to 140°F (-20 to 60°C) |
| Certifications | FCC Class 15 Part B CE |
| Warranty | 1 year against manufacturer defect |
| Performance - CO2 Channel | |
|---|---|
| Measurement Range | 0-10,000 ppm display 0-4,000 ppm voltage output |
| Display Resolution | ± 1 ppm |
| Accuracy | ±50 ppm or ±5% of reading up to 5,000 PPM (Above 5,000 PPM not specified) |
| Repeatability | ±20 ppm |
| Temperature Dependence | ±0.1% of reading per °C or ±2 ppm per °C, whichever is greater, referenced to 25°C |
| Pressure Dependence | 0.13% of reading per mm Hg (Corrected via user input for elevation) |
| Response Time | <60 seconds for 90% of step change |
| Warm-Up Time | <60 seconds at 22°C |
| Calibration Interval | 12 months, offset adjustment using single gas at 0-1000 ppm CO2. Full factory calibration available |
| Performance - Temperature Channel | |
|---|---|
| Temperature Range | Voltage output 32 to 104°F (0 to 40°C) Display 32 to 122°F (0 to 50°C) |
| Display Resolution | 0.1°F (0.1°C) |
| Display Options | °F, °C, or Off. Set with panel button. |
| Accuracy | ±2°F (±1°C) |
| Response Time | 20-30 minutes (case must equilibrate with environment) |
| Calibration Interval | 12 months, offset adjustment using temperature standard at 50 to 86°F (10 to 30°C). Full factory calibration available |
| Outputs | |
|---|---|
| (Analog) CO2 | 0-4 VDC, 1mV/ppm (4,000 ppm max) |
| Temperature | 0-4 VDC linear, 32-104°F (0-40°C) |
| Output Impedance | 100 Ohms |
| (Digital) | RS232 for use with Telaire® CO2View™ Graphing Software |
| (Wiring Connection) | One RJ-45 Connector Dual Analog output plus digital output |
| Power Supply | |
|---|---|
| Battery Type | Four AA batteries, not included |
| Battery Operation | 80 hours (alkaline) |
| External | 6 VDC from external AC/DC adapter, included |
| Power Requirements | 100 mA Peak, 20 mA average from 6V |