TPS Bromide Ion Selective Electrode
Bromide Ion Selective Electrode

Combination Bromide ISE with Intermediate Junction reference system.

Concentration Range 0.4 to 79,000 mg/L Br -
Linear Range 1 to 7,900 mg/L Br -
Slope -57 mV/decade +/-3 mV at 25 oC
Response Time <30 seconds for 90% of final value from 4.0 to 79,000 mg/L Br -
pH Range 2 to 14   (Note: An excess of OH - of 80 times the Br- concentration produces a 10% error)
Main Interferences OH-, Cl-, I-, CN-, S2-
Temperature Range -0 to 60 oC (32 to 140 oF)
Sensor Material Polycrystalline solid state membrane
Body and Sleeve Polypropylene/PEEK
Reference System Intermediate Junction (IJ) reference system is a double junction design. Ag/AgCl reference.