Chloride Ion Selective Electrode
Combination Chloride ISE with Intermediate Junction reference system.
Concentration Range | 1.8 to 35,000 mg/L Cl- |
Linear Range | 18 to 3,500 mg/L Cl- |
Slope | -57 mV/decade +/-3 mV at 25 oC |
Response Time | <30 seconds for 90% of final value from 18 to 35,000 mg/L Cl- |
pH Range | 2 to 14 (Note: An excess of OH - of 80 times the Br- concentration produces a 10% error) |
Main Interferences | OH-, NH4+, S2O32-, Br -, I-, CN-, S2- |
Temperature Range | -0 to 60 oC (32 to 140 oF) |
Sensor Material | Polycrystalline solid state membrane |
Body and Sleeve | Polypropylene/PEEK |
Reference System | Intermediate Junction (IJ) reference system is a double junction design. Ag/AgCl reference. |