R3-50
R3-50

3-Axis Ultrasonic Anemometer

Gill 3-axis Research range of ultrasonic anemometers offers users a number of different instruments. This range is valued for the quality of its measurements and is used by recognised experts to gather data they reply upon for their research and publications. 

The Research range includes models featuring horizontal head configurations for highly accurate vertical flow analysis, with minimal interruption from anemometer geometry. These models have a built in inclinometer to assist set-up on towers or masts.

The R3-50™ 3-axis anemometer has been designed for scientific research, particularly suitable for determining fluxes using the eddy covariance technique. This precision ultrasonic anemometer will monitor wind speeds up to 45m/s, with a fast data output rate of 50Hz. U, V, W and SOS (sonic temperature) outputs are available, and data can be logged using the supplied Gill WindCom™ software.

The R3-50 is of aluminium/carbon fibre construction and will operate effectively in environmental temperatures from -40°C to +60°C. Optional accessories are available including cables, sensor input units (analogue inputs) and power supplies.

  • 3-axis research anemometer
  • 50Hz output rate
  • 0-45 m/s wind speed
  • 0-360º wind direction
  • U, V, W vector outputs
  • Sonic temperature output
  • Aluminium/carbon fibre construction
  • Custom calibration provided as standard

 

 

WIND SPEED  
Range 0 - 45 m/s
Resolution 0.01 m/s
Accuracy <1% RMS
   
DIRECTION  
Range 0 - 360°
Resolution
Accuracy* <±1° RMS
   
SPEED OF SOUND  
Range 300-370 m/s
Resolution 0.01 m/s
Accuracy < ±0.5% @ 20ºC
   
MEASUREMENT  
Ultrasonic output rate 50 Hz
Parameters UVW, Speed of Sound
   
DIGITAL OUTPUT  
Communication RS422 full duplex, 8 data bits, 1 stop bit, no parity
Baud rates 2400 - 115200
Output parameters Selectable 0.4 to 50 Hz
   
PRT INPUT  
Input resolution 0.01°C
Input accuracy <0.01°C (0°C to +50°C) <0.15°C (-40°C to +60°C)
   
ANALOGUE OUTPUTS (REQUIRES PCIA UNIT)  
Quantity 7 (U, V, W, SoS, PRT plus 2 analogue outputs)
Scale ±10, ±20, ±30, ±60 m/s
Update rate 0.4 to 50 Hz
Range and resolution ±2.5 V, 14 bits
Accuracy <0.1% of FSR
   
ANALOGUE INPUTS (REQUIRES SIU)  
Quantity 6 differential inputs
Sampling rate 50 Hz
Range and resolution ±5 V, 14 bits
Accuracy <0.1% of FSR
   
POWER REQUIREMENT  
Anemometer 9-30 VDC (<150mA @ 24 VDC or 300mA @ 12 VDC)
   
MECHANICAL  
Material Aluminium/Carbon Fibre
Size 750mm x 240mm
Weight 1.0 kg
   
ENVIRONMENTAL  
Protection Class IP65
Operating Temp -40°C to +60°C
Precipitation 300mm/hr
EMC BS EN 50081-1: 1992 (Emissions) BS EN 50082-1: 1997 (Immunity
Suitable for exposure to a marine environment  

 

Search