90flt
90flt
90-FLT
  • 90FL-T measures pH, ORP, Conductivity, TDS, Salinity, Dissolved Oxygen, Turbidity & Temperature
  • Automatic calibration
  • Automatic datalogging and RS232 Serial Port Computer interface standard (cable sold separately)

TPS range of Field Lab Analysers provide a complete measurement solution for several of the most important water quality parameters.

User-friendly
All measurements are shown simultaneously on the 80 character display. There is no need to change modes to check any of your vital data. The user-friendly menu system guides the operator through all operations such as calibration, setup etc. Full text help and error messages are provided along the way.

Waterproof
The 90 series is designed for rugged field applications in wet environments. The case, keypad and connectors are rated to IP65.

Good Laboratory Practices
To comply with GLP guidelines, the date, time and results of the last calibration are stored in memory, along with the unit's serial number. This data can be displayed or sent directly to the RS232 port. Warning of failed calibration is provided. All readings stored in memory are stamped with the date and time. 

Notepad Function
A standard feature of the 90 series is the Notepad function. Up to 8100 sets of readings (depending on model) can be recorded for later review. These can be downloaded to a printer or computer via the RS232 port. For added functionality, the user is able to enter up to two additional items to be tagged on to the reading. Many operators use this function as a pond or site identifier, depth or readings from other instruments.

Datalogging
Automatic datalogging provides the capability to automatically record up to 8100 sets of readings (depending on model) at user-set intervals. The unit remains dormant between readings and only wakes up whenever a reading is due. Battery life is considerably improved with this facility.

RS232 Serial Port Interface
All recorded data can be downloaded directly to a Printer or Computer with the standard RS232 port (cable sold separately). With the WinTPS communication software, many instrument functions can be controlled from the computer. This makes the units ideal for remote monitoring.

The 90 Series is proudly designed and manufactured in Australia. Our aim is to provide you with the best quality, service and value for money. The TPS Quality System has been certified in accordance with the AS/NZS ISO 9001 standard. Since 1968, TPS has built a reputation for excellent backup service that is second to none.

 

pH  
Range 0.00 to 14.00 pH
Resolution 0.01 pH
Accuracy В±0.01pH
Temperature Compensation Automatic, -5 to 100oC
ORP  
Range 0 to +/-1999 mV
Resolution 1 mV
Accuracy +/-1 mV
Conductivity  
Ranges 4 ranges with auto range selection
k=0.1 Sensor : 
0 to 2.000 uS/cm
0 to 20.00 uS/cm
0 to 200.0 uS/cm
0 to 2000 uS/cm

k=1 Sensor : 
0 to 20.00 uS/cm
0 to 200.0 uS/cm
0 to 2000 uS/cm
0 to 20.00 mS/cm

k=10 Sensor : 
0 to 200.0 uS/cm
0 to 2000 uS/cm
0 to 20.00 mS/cm
0 to 200.0 mS/cm

Resolution 0.05% of full scale of selected range.
Accuracy В±0.5% of full scale of selected range at 25oC.
Temperature Compensation All readings automatically normalised to 25oC at -2.2%/oC.
TDS  
Ranges 4 ranges with auto range selection.
k=0.1 Sensor : 
0 to 1.000 ppM
0 to 10.00 ppM
0 to 100.0 ppM
0 to 1000 ppM

k=1 Sensor : 
0 to 10.00 ppM
0 to 100.0 ppM
0 to 1000 ppM
0 to 10.00 ppK

k=10 Sensor : 
0 to 100.0 ppM
0 to 1000 ppM
0 to 10.00 ppK
0 to 100.0 ppK

Resolution 0.1% of full scale of selected range.
Accuracy В±0.5% of full scale of selected range at 25oC.
Temperature Compensation All readings automatically normalised to 25oC at -2.2%/oC.
Salinity  
Ranges k=0.1 Sensor : 0 to 0.10 % 0 to 1.0 PSU
k=1 Sensor : 0 to 1.19 % 0 to 11.9 PSU
k=10 Sensor : 0 to 8.00 % 0 to 80.0 PSU
Resolution 0.01 % or 0.1 PSU
Accuracy В±0.5% of full scale of selected range at 25oC.
Temperature Compensation All readings automatically normalised to 25oC at -2.2%/oC.
Dissolved Oxygen  
Ranges 0 to 32.00 ppM (mg/L)
0 to 320.0 %Saturation
0 to 66.0 %Gaseous
Resolution 0.01 ppM (mg/L)
0.1 %Saturation
0.1 %Gaseous
Accuracy В±0.02 ppM (mg/L)
В±0.3 %Saturation
В±0.1 %Gaseous
Temperature Compensation Automatic for solubility (ppM only) and membrane permeability.
Turbidity  
Ranges 0.1 to 200.0 NTU
0 to 2000 NTU
(Ranges subject to sensor performance)
Resolution 0.1 NTU
1 NTU
Accuracy +/-1 NTU
Linearity 3%
Temperature  
Range -10.0 to 110.0 oC (Cond. sensor limit 60 oC)
Resolution 0.1 oC
Accuracy +/-0.2 oC
Datalogging 6500 readings (5400 with extra data input active).
Power  7.2V Rechargeable NiCad Battery
Any 12VDC supply will recharge the battery.

 