WETEC

Senseca

HD30.S1 / HD30.S2 Spectroradiometer Probes


Interchangeable spectral probes for visible and UV analysis with the HD30.1 data logger.

  • HD30.S1 covers 380…780 nm (visible) and HD30.S2 covers 220…400 nm (UV)
  • 2,048-element linear CCD detector with transmissive diffraction grating
  • Opaline quartz diffuser for cosine correction, with calibration data stored in each probe
Quantity
1

Overview

The Senseca HD30.S1 and HD30.S2 are interchangeable spectral probes for the HD30.1 portable spectroradiometer and data logger. Connected to the HD30.1 indicator, each probe extends the system to a different spectral range: the HD30.S1 operates in the 380…780 nm visible range, while the HD30.S2 operates in the 220…400 nm ultraviolet range. Together they cover the spectrum from 220 to 780 nm.

Both probes use a 2,048-element linear CCD detector and a spectrometer based on a transmissive diffraction grating. The HD30.S1 is calibrated with a reference halogen lamp and the HD30.S2 with a reference deuterium lamp. An opaline quartz diffuser provides cosine correction for measurements from extended light sources, and each probe stores its own calibration file, so the calibration data stay associated with the optical sensor.

Typical applications

  • Spectral and colorimetric analysis of light sources
  • Quality control of lighting systems
  • Verification of lighting in museums and exhibition spaces
  • Monitoring of blue-violet emissions that may damage sensitive exhibits
  • PAR measurements in agriculture, forestry and oceanography
  • Verification of ultraviolet sterilisation and disinfection systems
  • Dermatological phototherapy and UV medical treatments
  • Verification of blue-light lamps used in neonatal therapies
  • Non-destructive testing using ultraviolet radiation

Key Features

  • Two interchangeable probes: HD30.S1 (380…780 nm) and HD30.S2 (220…400 nm)
  • 2,048-element linear CCD detector with transmissive diffraction grating
  • Calibration file stored in each probe
  • Opaline quartz diffuser for cosine correction (3 mm on HD30.S1, 2 mm on HD30.S2)
  • Automatic or manual integration time, from 1 ms to 4 s
  • HD30.S1 measures spectral irradiance, irradiance, illuminance, PAR, CCT, CIE 1931 and CIE 1976 coordinates, CRI and spectral transmittance
  • HD30.S2 measures spectral irradiance, UVA, UVB and UVC irradiance and spectral transmittance
  • Stray light below 0.03%

Benefits

  • Visible and ultraviolet coverage from 220 to 780 nm with just two probes on one indicator
  • Calibration data stays with the optical sensor, because each probe stores its own calibration file
  • Calibrated against reference lamps: halogen for HD30.S1 and deuterium for HD30.S2
  • Fine spectral resolution, with a 4.5 nm pass band (HD30.S1) or 2.5 nm (HD30.S2) and 0.3 nm wavelength accuracy

Specifications

MEASUREMENT

 

Spectral range

HD30.S1: 380 nm…780 nm

HD30.S2: 220 nm…400 nm

Sensor

Linear CCD (2048 elements)

Type of spectrometer

Based on transmission diffraction grating

Numerical aperture

0.16

Input slit

HD30.S1: 125 µm

HD30.S2: 70 µm

Pass band

HD30.S1: 4.5 nm

HD30.S2: 2.5 nm

Wavelength accuracy

0.3 nm

Wavelength reproducibility

0.1 nm

Integration time

1 ms to 4 s (automatic or manual)

Stray light

<0.03%

Measuring mode

HD30.S1: Spectral irradiance, irradiance, illuminance [lux], PAR, proximal colour temperature, CIE 1931 (x,y) and CIE 1976 (u’,v’) coordinates, trichromatic coordinates, CRI, spectral transmittance

HD30.S2: Spectral irradiance, UV irradiance, UVB irradiance, UVC irradiance, spectral transmittance

Range of use (HD30.S1)

Illuminance 5…70,000 lux

Input optics

Φ 11.8 mm quartz diffuser

Cosine correction

HD30.S1: opaline quartz diffuser (3 mm)

HD30.S2: opaline quartz diffuser (2 mm)

 

 

ACCURACY

 

Calibration

HD30.S1: halogen lamp

HD30.S2: deuterium lamp

Incertitude (HD30.S1)

Spectral irradiance ±5%, illuminance ±4%, PAR ±4%, CCT ±45 K, x,y ±0.002, CRI ±1.5

Incertitude (HD30.S2)

Spectral irradiance ±15%, UV irradiance ±6%, UVB irradiance ±8%, UVC irradiance ±10%

 

 

ENVIRONMENTAL

 

Working temperature

0…40°C

 

 

MECHANICAL

 

Probe dimensions / weight

75 x 150 x H 74 mm / 370 g

Cable length

1.5 m

 

 

ORDERING CODES

 

HD30.S1

Spectral probe for the 380…780 nm visible range. Measures spectral irradiance, irradiance, illuminance, PAR and spectral transmittance, and calculates CCT, chromaticity coordinates and CRI

HD30.S2

Spectral probe for the 220…400 nm ultraviolet range. Measures spectral irradiance, UVA, UVB and UVC irradiance and spectral transmittance

System requirement

A complete measuring system consists of the HD30.1 indicator and at least one interchangeable spectral probe

Downloads

HD30.1 Datasheet (EN) DownloadHD30.1 Datasheet (EN)
HD30.1 User Manual (EN) DownloadHD30.1 User Manual (EN)