ISE Ion-Selective Electrodes
Ion-selective electrodes that measure specific ion concentrations in water, in mg/l.
- Ammonium, nitrate, chloride, sodium, calcium and bromide
- Accuracy of ±10% of reading or 2 mg/l, whichever is greater
- Resolution 0.1
- Corrected for ionic strength using conductivity readings
Overview
Solinst Eureka ISE Ion-Selective Electrodes are sensors that respond to the presence of specific ions and report ion concentrations in mg/l. They are used in conjunction with a reference electrode, and readings are corrected for ionic strength using conductivity readings. Available ions are ammonium, nitrate, chloride, sodium, calcium and bromide. Two types are used: solid-state ISEs, made of a solid billet of material, and liquid membrane ISEs, made of a thin doped polymer membrane covering an electrolyte with an electrode wire.
Typical applications
- Measuring ammonium and nitrate as indicators of nitrogen-based nutrients
- Detecting contamination from wastewater treatment plants or agricultural activities
- Measuring chloride as an indicator of contamination by sea water or other sources with higher chloride levels than expected in natural waters
- Tracking long-term specific-ion trends, which can signal the need for more detailed chemical study of the water and its contamination sources
- Calculating ammonia and total ammonia (with pH measurement) when operating at high pH
Key Features
- Ions: ammonium, nitrate, chloride, sodium, calcium and bromide
- Ammonium and nitrate: 0 to 100 mg/l as nitrogen
- Chloride: 0.5 to 18,000 mg/l; sodium: 0.05 to 20,000 mg/l
- Calcium: 0 to 40,000 mg/l; bromide: 0 to 80,000 mg/l
- Accuracy of ±10% of reading or 2 mg/l, whichever is greater; resolution 0.1
- Liquid or solid membrane sensor types
- Corrected for ionic strength using conductivity readings
- Two-point calibration
- Replaceable sensor tips that screw into a receiver with a simple o-ring seal
- Maximum depth of 15 m
Benefits
- Nutrient monitoring: Ammonium and nitrate are indicators of nitrogen-based nutrients from wastewater or agricultural sources.
- Contamination screening: Chloride measurements indicate contamination by sea water or other high-chloride sources.
- Concentration readings: ISE readings are corrected for ionic strength using conductivity, giving concentrations rather than raw activity.
- Easy tip replacement: Most ISEs screw into a receiver with a simple o-ring seal, making it easy to change tips when their lifetime is exhausted.
- Ammonia calculation: Measuring ammonium with an ISE together with pH allows ammonia and total ammonia to be calculated.
Specifications
GENERAL SPECIFICATIONS |
|
Sensor type | Liquid or solid membrane |
Parameters | Ammonium, nitrate, chloride, sodium, calcium, bromide |
Units | mg/l |
Calibration | Two point |
Maintenance | Cleaning, periodic replacement of sensor tips |
Sensor life | Six months |
Maximum depth | 15 m |
Reference electrode | Used in conjunction with a reference electrode |
Concentration correction | Corrected for ionic strength (via conductivity readings) |
Compatible probes | Fits the small sensor ports on the Manta+30B, Manta+35A and Manta+40; a small sensor may be substituted for any medium sensor |
ION RANGES |
|
Ammonium | Range: 0 to 100 mg/l as nitrogen |
Nitrate | Range: 0 to 100 mg/l as nitrogen |
Chloride | Range: 0.5 to 18,000 mg/l |
Sodium | Range: 0.05 to 20,000 mg/l |
Calcium | Range: 0 to 40,000 mg/l |
Bromide | Range: 0 to 80,000 mg/l |
Notes | The accuracy specification relies on non-trivial maintenance practice and frequent calibration near the temperature of measurement; sensors require periodic tip replacement |
COMPENSATION REQUIREMENTS |
|
Ammonium | Requires temperature, pH, conductivity/salinity and reference sensor compensation |
Nitrate | Requires temperature, conductivity/salinity and reference sensor compensation |
Chloride | Requires temperature, conductivity/salinity and reference sensor compensation |
USE AND CARE |
|
Sea water | Ammonium and nitrate sensors do not work in sea water |
Sensor tips | All liquid-junction ISEs have a life of about six months; any ISE replaceable tip older than six months should be replaced. Most ISEs screw into a receiver with a simple o-ring seal |
Storage | Do not let the ISE dry out; place a small amount of tap water in the sensor storage cup to ensure 100% humidity |
Downloads
| Water Quality Sensor Configurations Datasheet (EN) | DownloadWater Quality Sensor Configurations Datasheet (EN) |
|---|