LPPHOT03 Photometric Probe


Outdoor illuminance probe with photopic V(λ) spectral response and cosine correction.

  • Illuminance measurement in lux with <4% calibration uncertainty

  • Passive mV output without external power supply

  • 4…20 mA or 0…10 Vdc outputs, scaled 0…150 klux

  • RS485 Modbus-RTU output with selectable 0…20 or 0…200 klux ranges

  • Operating temperature −20…+60 °C

  • Output configurations vary by model

Quantity
1

Overview

The Senseca LPPHOT03 is an outdoor photometric probe designed to measure illuminance in lux, with a spectral response similar to the standard photopic curve V(λ) of the human eye. It features a diffuser for cosine correction and a protective dome for outdoor measurements. Each probe is individually calibrated against a reference luxmeter calibrated by a Primary Metrological Laboratory, with a calibration uncertainty of less than 4%.

The LPPHOT03 and LPPHOT03BL versions provide passive mV output without requiring an external power supply. The LPPHOT03BLAC and LPPHOT03BLAV versions offer normalized 4…20 mA and 0…10 Vdc outputs, respectively, both scaled from 0…150 klux. The LPPHOT03BLS version provides RS485 Modbus-RTU communication with selectable 0…20 klux and 0…200 klux measuring ranges. BL versions feature a base with a levelling device, and all versions operate from −20…+60 °C.

Typical Applications

  • Outdoor illuminance monitoring

  • Daylight and artificial light level measurement

  • Photometric measurements aligned with human visual response

  • Integration with analogue or RS485 Modbus-RTU data acquisition systems

Key Features

  • Illuminance measurement in lux with photopic V(λ) spectral response

  • Diffuser for cosine correction and protective dome for outdoor use

  • Photopic response error f'1 <6%, cosine response error f2 <3% and linearity error f3 <1%

  • Individual calibration against a reference luxmeter with <4% calibration uncertainty

  • Passive mV/klux output without external power supply

  • Normalized 4…20 mA or 0…10 Vdc outputs, scaled 0…150 klux

  • RS485 Modbus-RTU output with selectable 0…20 klux or 0…200 klux ranges

  • Base with levelling device on BL versions

  • Operating temperature −20…+60 °C

  • 4-pole M12 connector for mV and analogue versions; 8-pole M12 connector for RS485 version

Benefits

  • Illuminance measurements aligned with human visual response

  • Cosine correction supports outdoor light measurements

  • Individual calibration provides a documented calibration reference

  • No external power supply needed for passive mV versions

  • Choice of output configurations for different monitoring systems

  • Easier probe positioning with the levelling device on BL versions

Specifications

Measurement

Measured quantity

Illuminance

Measurement unit

lux

Spectral range

V(λ)

Typical sensitivity

0.5…1.5 mV/(klux)

Calibration uncertainty

<4%

Photopic response error (f'1)

<6%

Cosine response error (f2)

<3%

Linearity error (f3)

<1%

Impedance (non-normalized version)

0.5…1.0 kΩ

 

 

Model

Output

Measuring range / scaling

LPPHOT03

mV/klux

Not specified in datasheet

LPPHOT03BL

mV/klux

Not specified in datasheet

LPPHOT03BLAC

4…20 mA

4 mA = 0 klux; 20 mA = 150 klux

LPPHOT03BLAV

0…10 Vdc

0 V = 0 klux; 10 V = 150 klux

LPPHOT03BLS

RS485 Modbus-RTU

Selectable 0…20 klux or 0…200 klux

 

 

RS485 Modbus-RTU

Communication interface

RS485

Communication protocol

Modbus-RTU

Low measuring range

0…20,000 lux

High measuring range

0…200,000 lux

Range selection

Serial command

Measurement register

Address 2

Measurement data format

16-bit integer

Low-range measurement register scaling

Register value represents illuminance in lux

High-range measurement register scaling

Illuminance (lux) = register value × 10; resolution 10 lux

Status register

Address 3

Status register format

16-bit integer

Status bit 0

Measurement error when set to 1

Status bit 2

Configuration data error when set to 1

Status bit 3

Program memory error when set to 1

Average value register

Address 4

Averaging

Last 4 measurements

Sensor signal register

Address 5

Low-range sensor signal scaling

Sensor signal in µV

High-range sensor signal scaling

Sensor signal (µV) = register value × 10; resolution 10 µV

 

 

Electrical

Passive mV output

No external power supply required

4…20 mA normalized output

10…30 Vdc

0…10 Vdc normalized output

15…30 Vdc

RS485 Modbus-RTU output

5…30 Vdc

 

 

Environmental

Intended use

Outdoor photometric measurements

Operating temperature

−20…+60 °C

 

 

Construction and Connection

Cosine correction component

Diffuser

Protective component

Glass dome

Connector type

M12 male connector

4-pole M12 connector

LPPHOT03, LPPHOT03BL, LPPHOT03BLAC and LPPHOT03BLAV

8-pole M12 connector

LPPHOT03BLS

Levelling device

Base with levelling device on BL versions

Available cable lengths

2 m, 5 m or 10 m

Cable connection

Female M12 connector

Cable ordering

Ordered separately

 

 

Calibration

Calibration

Individually calibrated

Calibration method

Comparison with a reference luxmeter

Reference luxmeter

Calibrated by a Primary Metrological Laboratory

Calibration procedure

CEI Publication No. 69 (1987)

Calibration factor

Shown on the probe housing

Calibration documentation

Calibration report

Drawing

LPPHOT03 Photometric Probe Drawing