WETEC
LPPIRG01 Pyrgeometer thumbnail 1
LPPIRG01 Pyrgeometer thumbnail 2

Senseca

LPPIRG01 Pyrgeometer


Pyrgeometer for far infrared radiation measurement and energy balance studies.

  • Thermopile sensor with −300…+300 W/m² measuring range

  • Spectral range 5.5…45 µm (50%) and 160° viewing angle

  • Passive mV output (LPPIRG01) or RS485 Modbus-RTU output (LPPIRG01S)

  • IP67-rated housing with anodized aluminium body, ASA shield and silicon window

Quantity
1

Overview

The Senseca LPPIRG01 / LPPIRG01S pyrgeometer measures far infrared radiation at wavelengths greater than 4.5 µm for atmospheric radiation monitoring and energy balance studies. It uses a blackened thermopile sensor and an integrated NTC temperature sensor to determine infrared irradiance. The instrument provides a measuring range of −300…+300 W/m², a spectral range of 5.5…45 µm (50%) and a 160° viewing angle.

The LPPIRG01 features a passive mV output without requiring an external power supply, while the LPPIRG01S provides RS485 Modbus-RTU communication and operates on 5…30 Vdc. Both models feature IP67-rated protection, an anodized aluminium housing, an ASA radiation protection shield and a silicon window with a scratch-resistant DLC coating. An integrated bubble level assists horizontal positioning, while a silica-gel cartridge helps prevent internal condensation. Each instrument is factory calibrated outdoors against a reference pyrgeometer calibrated at the World Radiation Center (WRC) and supplied with a calibration report.

Typical Applications

  • Far infrared radiation measurement for atmospheric studies

  • Downward and upward long-wave radiation monitoring

  • Energy balance studies using pyrgeometers and an albedometer

  • Infrared irradiance data acquisition using data loggers

Key Features

  • Blackened thermopile sensor for far infrared radiation measurement

  • Integrated NTC sensor for instrument temperature measurement

  • Measuring range −300…+300 W/m²

  • Spectral range 5.5…45 µm (50% transmission)

  • 160° viewing angle

  • Typical sensitivity 5…10 µV/(W/m²)

  • Passive mV output without external power supply (LPPIRG01)

  • RS485 Modbus-RTU output with 5…30 Vdc power supply (LPPIRG01S)

  • IP67 protection with anodized aluminium housing and ASA solar radiation protection screen

  • Silicon window with scratch-resistant DLC coating

  • Integrated bubble level with accuracy better than 0.1°

  • Silica-gel cartridge to absorb internal moisture and help prevent condensation

  • Operating temperature −40…+80 °C

  • Factory calibrated outdoors against a reference pyrgeometer calibrated at the World Radiation Center (WRC), with calibration report supplied

Benefits

  • Supports far infrared radiation monitoring and energy balance studies

  • Passive mV (LPPIRG01) or RS485 Modbus-RTU (LPPIRG01S) outputs to suit different monitoring systems

  • Passive model requires no external power supply

  • IP67 protection and weather-resistant construction support outdoor installations

  • DLC-coated silicon window reduces the risk of scratching during cleaning

  • Silica-gel cartridge absorbs internal moisture to help prevent condensation on the silicon window

  • Integrated bubble level assists precise horizontal positioning

  • Factory calibrated outdoors against a reference pyrgeometer calibrated at the World Radiation Center (WRC), with calibration report supplied

Specifications

Measurement

Sensor

Thermopile

Measuring range

−300…+300 W/m²

Spectral range (50%)

5.5…45 µm

Viewing angle

160°

Typical sensitivity

5…10 µV/(W/m²)

Impedance

33…45 Ω

 

 

Performance According to ISO 9060:2018

Response time (95%)

<28 s

Zero offset (Type B): response to a 5 K/h change in ambient temperature

<4 W/m² (absolute magnitude)

Long-term instability (1 year)

<1.5% (absolute magnitude)

Non-linearity

<1% (absolute magnitude)

Spectral selectivity

<5% (absolute magnitude)

Temperature response

<3%

Tilt response

<2% (absolute magnitude)

 

 

Electrical

LPPIRG01

LPPIRG01S

Output

Passive mV

RS485 Modbus-RTU

Power supply

No power required

5…30 Vdc

Current consumption

Not specified

2 mA @ 24 Vdc

 

 

Electrical Connection

Connection

8-pole M12 connector

 

 

Environmental

Operating temperature

−40…+80 °C

Operating relative humidity

0…100% RH

Maximum operating altitude

6000 m

Protection degree

IP67

 

 

Mechanical

Housing material

Anodized aluminium

Screen material

ASA

Sensitive area window material

Silicon

Weight

Approximately 0.9 kg

Bubble level accuracy

<0.1°

 

 

Reliability

Mean time between failures (MTBF)

>10 years

 

 

RS485 Modbus-RTU Communication – LPPIRG01S Only

Communication protocol

Modbus-RTU

Communication interface

RS485

Modbus address range

1…247

Default Modbus address

1

Supported baud rates

9600, 19200, 38400, 57600, 115200 bps

Default baud rate

19200 bps

Data bits

8

Supported communication formats

8N1, 8N2, 8E1, 8E2, 8O1, 8O2

Default communication format

8E1

RS485 output isolation

Not isolated

 

 

Calibration

Calibration

Factory calibrated

Calibration method

Outdoor comparison with a reference pyrgeometer

Reference instrument

Pyrgeometer calibrated at the World Radiation Center (WRC)

Calibration conditions

Instruments kept outdoors for at least one night under clear skies

Calibration documentation

Calibration report supplied

 

 

Drawing

LPSILICON-PYRA04 Silicon Pyranometer Drawing

Downloads

LPPIRG01 Family Datasheet DownloadLPPIRG01 Family Datasheet
LPPIRG01 Family User Manual DownloadLPPIRG01 Family User Manual
LPPIRG01 Family CE Declaration of Conformity DownloadLPPIRG01 Family CE Declaration of Conformity