WETEC
LPS03 Class C Pyranometer thumbnail 1
LPS03 Class C Pyranometer thumbnail 2
LPS03 Class C Pyranometer thumbnail 3
LPS03 Class C Pyranometer thumbnail 4

Senseca

LPS03 Class C Pyranometer


Spectrally Flat Class C thermopile pyranometer for global solar irradiance measurement and monitoring.

  • Measuring range −200…4000 W/m² and spectral range 300…2800 nm

  • ISO 9060:2018 classification and factory calibration according to ISO 9847:2023

  • RS485 Modbus-RTU, configurable analog, 2-wire 4…20 mA or passive mV outputs (model-dependent)

  • IP67 protection with integrated bubble level for horizontal positioning

Quantity
1

Overview

The Senseca LPS03 series comprises Spectrally Flat Class C pyranometers designed to measure global solar irradiance on a flat surface, including direct and diffuse solar radiation. Based on the thermopile measuring principle, the pyranometers cover a measuring range of −200…4000 W/m² and a spectral range of 300…2800 nm. Classified according to ISO 9060:2018, they are suitable for research, meteorology and photovoltaic (PV) monitoring.

Depending on the model, the LPS03 pyranometers provide RS485 Modbus-RTU, configurable analog, 2-wire 4…20 mA or passive mV outputs. Digital models incorporate internal temperature, relative humidity and pressure sensors for monitoring operating conditions, with tilt sensing available on selected models. The instruments feature an integrated bubble level and adjustable feet for horizontal positioning, along with IP67 protection, an anodized aluminium housing, an ASA protection screen and an optical glass dome. Each pyranometer is factory calibrated according to ISO 9847:2023 (Type A1) and supplied with an individual calibration report.

Typical Applications

  • Global solar irradiance measurement for scientific research

  • Solar radiation monitoring in meteorological applications

  • Photovoltaic (PV) solar irradiance monitoring

Key Features

  • Spectrally Flat Class C classification according to ISO 9060:2018

  • Thermopile sensor for global solar irradiance measurement

  • Measuring range −200…4000 W/m²

  • Spectral range (50%) 300…2800 nm

  • Resolution 0.1 W/m²

  • Response time (95%) < 18 s

  • RS485 Modbus-RTU, configurable analog (4…20 mA, 0…20 mA, 0…1 V, 0…5 V or 0…10 V), 2-wire 4…20 mA or passive mV outputs (model-dependent)

  • Internal temperature, relative humidity and pressure diagnostic sensors (LPS03Mxx models)

  • Optional tilt sensor (LPS03M0T and LPS03MAT models)

  • Integrated bubble level and adjustable feet for horizontal positioning

  • IP67 protection with anodized aluminium housing, ASA protection screen and optical glass dome

  • Operating temperature −40…+80 °C

  • Factory calibrated according to ISO 9847:2023 (Type A1), with individual calibration report

Benefits

  • Reliable global solar irradiance measurement for research, meteorology and PV monitoring

  • Flexible output options for integration with different monitoring and data acquisition systems

  • Easier installation with integrated bubble level and adjustable feet

  • Continuous monitoring of operating conditions through built-in diagnostic sensors (digital models)

  • Optional tilt monitoring to help maintain correct sensor positioning

  • Durable outdoor operation with IP67 protection and weather-resistant materials

  • Convenient configuration and real-time monitoring using DATAsense software (digital models)

  • Supplied factory calibrated with an individual calibration report

Specifications

General Measurement

Sensor and measuring principle

Thermopile (Seebeck effect)

Typical sensitivity

5…15 µV/(W/m²)

Measuring range

−200…4000 W/m²

Resolution

0.1 W/m²

Viewing angle

2π sr

Spectral range (50%)

300…2800 nm

 

 

Technical According to ISO 9060:2018

Classification

Spectrally Flat Class C

Response time (95%)

<18 s

Zero offset — response to 200 W/m² thermal radiation

Magnitude <15 W/m²

Zero offset — response to 5 K/h ambient temperature change

Magnitude <4 W/m²

Total zero offset, including the above effects and other sources

Magnitude <20 W/m²

Long-term instability (1 year)

Magnitude <1%

Non-linearity

Magnitude <1%

Directional response (up to 80° with 1000 W/m² beam)

Magnitude <20 W/m²

Spectral error

Magnitude <1%

Temperature response (−10…+40 °C)

Magnitude <2%

Tilt response

Magnitude <1.5%

 

 

Model

Output

LPS03M00

Isolated RS485 Modbus-RTU

LPS03M0T

Isolated RS485 Modbus-RTU

LPS03MA0

Isolated RS485 Modbus-RTU + configurable analog output

LPS03MAT

Isolated RS485 Modbus-RTU + configurable analog output

LPS030C0

2-wire current loop, 4…20 mA

LPS030P0

Passive mV

 

 

Configurable Analog Output — LPS03MA0 and LPS03MAT

Default analog output

4…20 mA

Other selectable outputs

0…20 mA, 0…1 V, 0…5 V, 0…10 V

Default analog output irradiance range

0…2000 W/m²

Analog output irradiance range

User-configurable

 

 

Electrical

Power supply — RS485 output

7…30 Vdc

Power supply — analog output (except 0…10 V)

10…30 Vdc

Power supply — 0…10 V output

15…30 Vdc

Consumption — LPS03M0x

15 mA at 24 Vdc; 21 mA at 12 Vdc

Consumption — LPS03MAx

37 mA at 24 Vdc and Iout = 22 mA; 43 mA at 12 Vdc and Iout = 22 mA

Connection — except LPS03MAx

5-pole M12

Connection — LPS03MAx

8-pole M12

 

 

Internal Diagnostic Sensors — LPS03Mxx Models (Internal Temperature)

Measuring range

−40…+80 °C

Resolution

0.1 °C

Accuracy

±0.5 °C (0…60 °C)

 

 

Internal Diagnostic Sensors — LPS03Mxx Models (Internal Relative Humidity)

Measuring range

0…100% RH

Resolution

0.1% RH

Accuracy

±3% RH at 25 °C (20…80% RH)

 

 

Internal Diagnostic Sensors — LPS03Mxx Models (Internal Pressure)

Measuring range

300…1100 hPa

Resolution

0.1 hPa

Accuracy

±1 hPa (0…60 °C)

 

 

Tilt Sensor — LPS03M0T and LPS03MAT Models

Measuring range

0…180°

Resolution

0.1°

Accuracy

<0.5°

 

 

Environmental and Mechanical

Operating temperature

−40…+80 °C

Operating relative humidity

0…100% RH

Maximum operating altitude

6000 m

Protection degree

IP67

Housing material

Anodized aluminium

Protection screen material

ASA

Dome material

Optical glass

Dome outer diameter

32 mm

Dome thickness

4 mm

Weight

Approximately 230 g

Bubble level accuracy

<0.2°

Mean time between failures (MTBF)

>10 years

 

 

Calibration and Other Standards

Calibration standard

ISO 9847:2023 (Type A1)

Calibration method

Comparison with a reference pyranometer

Reference calibration

Reference sample calibrated annually at WRC (World Radiation Center)

Calibration documentation

Individual calibration report supplied

Additional certification

ISO 17025 Calibration Certificate available upon request

Other stated compliance

WMO recommendations and IEC 61724-1 requirements

 

 

Configuration and Monitoring — LPS03Mxx Models

PC application software

DATAsense

Sensor configuration

Modbus parameters and analog output measuring range

Measurement monitoring

Real-time monitoring, including graphical irradiance display

Data recording

Measurement values can be saved to a file during connection

Additional monitoring function

Built-in days-of-operation counter

Drawing

LPS03 Class C Pyranometer for Accurate Global Solar Irradiance Measurement and Monitoring Drawing

Downloads

LPS03 Family Datasheet DownloadLPS03 Family Datasheet
LPS03 Family User Manual DownloadLPS03 Family User Manual
LPS03 Family CE Declaration of Conformity DownloadLPS03 Family CE Declaration of Conformity