WETEC

Senseca

LPS12 Class B pyranometer with Shadow Ring


Spectrally Flat Class B pyranometer with adjustable shadow ring for diffuse solar irradiance measurement.

  • Thermopile sensor, −200…4000 W/m² measuring range and 283…2800 nm spectral range

  • RS485 Modbus-RTU, configurable analog, 2-wire 4…20 mA or passive mV outputs, depending on model

  • IP67 protection, integrated bubble level and adjustable feet for outdoor installation

Quantity
1

Overview

The Senseca LPS12 is a Spectrally Flat Class B pyranometer equipped with an adjustable shadow ring for measuring diffuse solar irradiance on a flat surface. The shadow ring blocks direct solar radiation from reaching the thermopile sensor, allowing diffuse irradiance measurement. Classified according to ISO 9060:2018, the pyranometer provides a measuring range of −200…4000 W/m² and a spectral range of 283…2800 nm.

Depending on the model, the LPS12 offers RS485 Modbus-RTU, configurable analog, 2-wire 4…20 mA or passive mV outputs. Digital models include internal temperature, relative humidity and pressure sensors for monitoring operating conditions, with optional tilt monitoring. The instrument features a goniometer, adjustable feet and integrated bubble level for positioning and alignment. Its IP67 protection supports outdoor operation, while factory calibration according to ISO 9847:2023 provides an individual calibration report.

Typical Applications

  • Diffuse solar irradiance measurement and monitoring

  • Cloud cover and atmospheric transparency assessment

  • Weather forecasting and climate modelling

  • Solar energy potential assessment and system optimisation

  • Atmospheric radiation and climate variability studies

Key Features

  • Spectrally Flat Class B pyranometer according to ISO 9060:2018

  • Adjustable shadow ring for diffuse solar irradiance measurement

  • Thermopile sensor with black matt coated sensitive surface

  • Measuring range −200…4000 W/m²

  • Spectral range 283…2800 nm

  • RS485 Modbus-RTU, configurable analog, 2-wire 4…20 mA or passive mV outputs, depending on model

  • Internal temperature, relative humidity and pressure diagnostics in digital models

  • Optional tilt sensor for installation position monitoring

  • Goniometer, graduated sliding bars, adjustable feet and integrated bubble level

  • IP67 protection with aluminium alloy and stainless steel ring and stand

  • Factory calibrated according to ISO 9847:2023 with individual calibration report

Benefits

  • Diffuse solar irradiance measurement by blocking direct sunlight

  • Class B measurement performance according to ISO 9060:2018

  • Flexible output options for integration with monitoring systems

  • Easier installation and alignment with adjustable positioning components

  • Internal diagnostics for monitoring sensor operating conditions in digital models

  • IP67 protection for outdoor operation

  • Factory calibrated with individual calibration report for measurement confidence

Specifications

General

Product series

PYRAsense LPS12

Instrument type

Pyranometer with adjustable shadow ring

Measurement

Diffuse solar irradiance on a flat surface

Classification

Spectrally Flat Class B

Classification standard

ISO 9060:2018

Standards and recommendations

WMO recommendations and IEC 61724-1 requirements

Mean time between failures (MTBF)

> 10 years

 

 

Irradiance Measurement

Sensor

Thermopile

Measuring range

−200…4000 W/m²

Resolution

0.1 W/m²

Spectral range (50%)

283…2800 nm

Typical sensitivity

6…15 µV/(W/m²)

Viewing angle

2π sr

 

 

Measurement Performance According to ISO 9060:2018

Response time (95%)

< 10 s

Zero offset — response to 200 W/m² thermal radiation

Absolute value < 10 W/m²

Zero offset — response to 5 K/h ambient temperature change

Absolute value < 4 W/m²

Total zero offset, including other sources

Absolute value < 15 W/m²

Long-term instability (1 year)

Absolute value < 1%

Non-linearity

Absolute value < 1%

Directional response (up to 80° with 1000 W/m² beam)

Absolute value < 18 W/m²

Spectral error

Absolute value < 0.5%

Temperature response (−10…+40 °C)

Absolute value < 1.5%

Tilt response

Absolute value < 1%

 

 

Electrical Outputs

(Model)

Output

LPS12M00

Isolated RS485 Modbus-RTU

LPS12M0T

Isolated RS485 Modbus-RTU

LPS12MA0

Isolated RS485 Modbus-RTU + configurable analog output

LPS12MAT

Isolated RS485 Modbus-RTU + configurable analog output

LPS120C0

2-wire (current loop) 4…20 mA

LPS120P0

Passive mV

 

 

Configurable Analog Output — LPS12MAx Models

Default output

4…20 mA

Alternative current output

0…20 mA

Alternative voltage outputs

0…1 V, 0…5 V, 0…10 V

Default analog output irradiance range

0…2000 W/m²

Analog irradiance range configuration

User-configurable within the full measuring range

 

 

Power Supply and Consumption

RS485 output power supply

7…30 Vdc

Analog output power supply (except 0…10 V)

10…30 Vdc

0…10 V output power supply

15…30 Vdc

LPS12M0x consumption

15 mA at 24 Vdc; 21 mA at 12 Vdc

LPS12MAx consumption

37 mA at 24 Vdc; 43 mA at 12 Vdc (Iout = 22 mA)

LPS120P0 power supply

Not required (passive output)

 

 

Internal Diagnostic Sensors — LPS12Mxx Models

Internal temperature measuring range

−40…+80 °C

Internal temperature resolution

0.1 °C

Internal temperature accuracy

±0.5 °C (0…60 °C)

Internal relative humidity measuring range

0…100% RH

Internal relative humidity resolution

0.1% RH

Internal relative humidity accuracy

±3% RH at 25 °C (20…80% RH)

Internal pressure measuring range

300…1100 hPa

Internal pressure resolution

0.1 hPa

Internal pressure accuracy

±1 hPa (0…60 °C)

 

 

Optional Tilt Sensor — LPS12MxT Models

Available models

LPS12M0T and LPS12MAT

Measuring range

0…180°

Resolution

0.1°

Accuracy

< 0.5°

 

 

Mechanical and Installation

Shadow ring

Adjustable

Shadow ring adjustment

Graduated sliding bars

Positioning components

Goniometer, integrated bubble level and adjustable feet

Number of adjustable feet

3

Bubble level accuracy

< 0.2°

Weight

Approximately 6.8 kg (pyranometer, support and shadow ring)

Pyranometer housing

Anodized aluminium

Screen

ASA

Dome

Optical glass

Ring and stand

Aluminium alloy and stainless steel

 

 

Environmental

Operating temperature

−40…+80 °C

Operating relative humidity

0…100% RH

Maximum operating altitude

6000 m

Protection degree

IP67

 

 

Electrical Connection and Communication

Connector — LPS12M0x, LPS120C0 and LPS120P0

5-pole M12

Connector — LPS12MAx

8-pole M12

Default Modbus address

1

Default communication settings

19200 baud, 8E1

Configuration and monitoring software

DATAsense

 

 

Calibration

Factory calibration standard

ISO 9847:2023 (Type A1)

Calibration method

Comparison with a reference pyranometer

Calibration documentation

Individual calibration report

Optional certification

ISO 17025 calibration certificate available upon request

 

 

Drawing

LPS12 Class B pyranometer with Shadow Ring Drawing

Downloads

LPS12 Family Datasheet DownloadLPS12 Family Datasheet
LPS12-LPS13 Family User Manual DownloadLPS12-LPS13 Family User Manual
LPS12 Family CE Declaration of Conformity DownloadLPS12 Family CE Declaration of Conformity