WETEC

Eureka

Manta+ 40 Multiparameter Sonde


Multiparameter water quality sonde with room for up to 14 sensors.

  • Standard turbidity, temperature, pH, conductivity and optical dissolved oxygen sensors, with optional ORP and depth
  • Accommodates three fluorometers (medium sensors) plus three ISE sensors (small sensors)
  • Add a battery pack for autonomous self-powered logging
Quantity
1

Overview

The Solinst Eureka Manta+ 40 is a multiparameter sonde for monitoring water quality in fresh, brackish and marine environments. It comes standard with turbidity, temperature, pH, conductivity and optical dissolved oxygen sensors, with the option to add ORP and depth sensors, and it accommodates three fluorometers plus three ISE sensors. Connect it to any of the Solinst Eureka field display options for sampling in the field, add an internal battery pack for autonomous self-powered logging, or connect to a telemetry station to view real-time data in the cloud.

Typical applications

  • Environmental monitoring: baseline water quality assessments and long-term monitoring of aquatic ecosystems
  • Environmental monitoring: monitoring wetland health and restoration projects
  • Environmental monitoring: assessing the effects of agricultural runoff
  • Drinking water monitoring: source water protection and monitoring
  • Drinking water monitoring: treatment plant optimization and process control
  • Drinking water monitoring: distribution system monitoring
  • Drinking water monitoring: detecting and responding to contamination events
  • Stormwater monitoring: monitoring the effects of storm events on local waterways
  • Stormwater monitoring: monitoring the effectiveness of stormwater controls and treatment systems
  • Hydrology and hydrogeology: groundwater monitoring for aquifer characterization
  • Hydrology and hydrogeology: surface water and groundwater interaction studies
  • Hydrology and hydrogeology: monitoring saltwater intrusion in coastal aquifers
  • Dredging and construction: monitoring turbidity during dredging operations
  • Dredging and construction: assessing the impact of construction projects
  • Dredging and construction: monitoring runoff from construction sites
  • Mining: monitoring runoff and discharge
  • Mining: monitoring the effects of mining on ground and surface water
  • Mining: monitoring the effects of acid mine drainage
  • Wastewater management: monitoring effluent quality for compliance
  • Wastewater management: optimizing wastewater treatment processes
  • Wastewater management: tracking the impact of wastewater discharges
  • Oceanography and marine research: monitoring coastal water quality and ocean acidification
  • Oceanography and marine research: studying the health of coral reefs, marine ecosystems and biodiversity
  • Oceanography and marine research: tracking the movement of water masses and currents
  • Aquaculture: maintaining optimal water quality for fish and shellfish farming
  • Aquaculture: preventing disease outbreaks and optimizing growth
  • Industrial applications: process water monitoring and control
  • Industrial applications: cooling water monitoring and treatment
  • Industrial applications: monitoring environmental impact of industrial activities and effluents
  • Research and education: conducting scientific research on water quality and aquatic ecosystems
  • Research and education: providing hands-on learning experiences for students in environmental science

Key Features

  • Standard turbidity, temperature, pH, conductivity and optical dissolved oxygen sensors, with optional ORP and depth sensors
  • Accommodates three fluorometers (medium sensors) plus three ISE sensors (small sensors)
  • Optical DO sensor with long-life (5+ years) replaceable cap, and refillable pH reference
  • Anti-fouling wiper, anti-corrosive housings and sensors, and robust marine bulkhead connectors
  • Greater than 1,000,000 logged readings at a sample rate of 1 Hz
  • 200 m depth rating (15 m maximum for ISE and TDG sensors)
  • RS-232, SDI-12, MODBUS, USB or Bluetooth® communication
  • LED status indicators on each sonde provide essential diagnostic information

Benefits

  • Comprehensive Monitoring: Measure multiple water quality parameters simultaneously, providing a complete picture of water health.
  • Versatile Use: Suitable for various applications, from spot checking to continuous monitoring.
  • Real-time Data: Connect to telemetry stations for immediate access to water quality data in the cloud.
  • Durable: Features anti-corrosive housings and sensors, robust marine bulkhead connectors, and anti-fouling options.
  • Easy to Use: One touch and automatic data capture, simple calibrations, collect and manage data with mobile apps, LED status indicators.

Specifications

GENERAL SPECIFICATIONS

 

Standard sensors

Temperature, pH, conductivity, dissolved oxygen (optical), turbidity, anti-fouling wiper

Optional sensors

ORP (combined with pH sensor); depth, vented or absolute; up to three fluorometers (medium sensors); up to three ISE or TDG sensors (small sensors)

Number of sensors

Up to 14

Diameter

4.0"

Length (with guard)

No battery pack: 21"
With internal battery pack (IBP): 22.5"
With external Li-ion battery pack (EBP): 28.5"

Weight (with guard)

No battery pack: 10.65 lbs
With IBP: 11.65 lbs
With EBP: 12.9 lbs

Internal battery pack

6 "C" size batteries

Operating temperature

-5 to 50°C

Depth rating

200 m (15 m maximum for ISE and TDG sensors)

Communication

RS-232, SDI-12, MODBUS, USB or Bluetooth®

Sample rate

1 Hz

Data memory

>1,000,000 logged readings

 

 

STANDARD SENSOR SPECIFICATIONS

 

Temperature

Range: -5 to 50°C
Resolution: 0.01
Accuracy: ±0.1
Calibration not required

pH

Range: 0 to 14 units
Resolution: 0.01
Accuracy: ±0.1 within 10°C of calibration; or 0.2
Refillable reference electrode; corrected for temperature; typical sensor life >6 years

Turbidity

Range: 0 to 1000 FNU; 1000 to 4000 FNU
Resolution: 0.01
Accuracy: ±0.3 FNU or ±2% of reading, whichever is greater (0 to 1000 FNU); ±4% of reading (1000 to 4000 FNU)
Filtered for non-turbidity spikes; includes wiper to clean the optics; FNU and NTU are interchangeable

Dissolved oxygen (optical) - concentration

Range: 0 to 20 mg/l, ±0.1; 20 to 30 mg/l, ±0.15; 30 to 50 mg/l, ±5% of reading
Resolution: 0.01
Compensated for temperature and salinity; EPA approved "lifetime" luminescence method; typical sensor cap life >6 years

Dissolved oxygen (optical) - % saturation

Range: 0 to 500% saturation
Resolution: 0.1
Accuracy: corresponds with the accuracy of the concentration reading

Conductivity - specific conductance (µS/cm)

Range: 0 to 5000 µS/cm
Resolution: 0.1
Accuracy: ±0.5% of reading or ±1, whichever is greater
Corrected for temperature; four easy-to-clean graphite electrodes; optional sensor provides ±0.5% of reading accuracy to 100 mS/cm

Conductivity - specific conductance (mS/cm)

Range: 0 to 100 mS/cm; 100 to 275 mS/cm
Resolution: 0.001
Accuracy: ±1% of reading ±0.001 (0 to 100 mS/cm); ±2% of reading (100 to 275 mS/cm)

Conductivity - salinity

Range: 0 to 70 PSU
Resolution: 0.01
Accuracy: ±2% of reading
Calculated from conductivity and temperature; PSU is equivalent to ppt

Conductivity - total dissolved solids (TDS)

Range: 0 to 65 g/l
Resolution: 0.1
Accuracy: ±5% of reading
Calculated from conductivity and temperature

 

 

OPTIONAL SENSOR SPECIFICATIONS

 

ORP

Range: -999 to 999 mV
Resolution: 0.1
Accuracy: ±20 mV
Optional ORP sensor is combined with pH sensor

Depth (absolute)

Range: 0 to 25 m, accuracy ±0.05; 0 to 200 m, accuracy ±0.4
Resolution: 0.01
Compensated for temperature and salinity

Depth (vented)

Range: 0 to 10 m
Resolution: 0.001
Accuracy: ±0.003
Compensated for temperature, salinity and barometric pressure

Barometric pressure

Range: 400 to 900 mm Hg
Resolution: 0.1
Accuracy: ±1.5
Included with depth sensor

Total dissolved gas (TDG)

Range: 400 to 1,400 mm Hg
Resolution: 0.1
Accuracy: ±1
Compensated for temperature; maximum depth 15 m; cannot be wiped

 

 

FLUOROMETERS (MEDIUM SENSORS, UP TO THREE)

 

Resolution and linearity

Resolution: 0.01; linearity of 0.99 R²

Chlorophyll a - blue

0 to 100 µg/l

Chlorophyll a - red

0 to 500 µg/l

Rhodamine dye

0 to 200 ppb

Phycocyanin (freshwater BGA)

0 to 4500 ppb

Phycoerythrin (marine BGA)

0 to 700 ppb

CDOM/FDOM

0 to 500 ppb

Optical brightener

0 to 300 ppb

Tryptophan

0 to 5000 ppb

Fluorescein dye

0 to 150 ppb

PTSA

0 to 650 ppb

Refined oil

0 to 20 ppm

Crude oil

0 to 300 ppb

Caution

Never look directly at a fluorometer sensor. The UV rays emitted by the sensor can cause eye damage.

 

 

ION-SELECTIVE ELECTRODES (SMALL SENSORS, UP TO THREE)

 

Resolution and accuracy

Resolution: 0.1
Accuracy: ±10% of reading or 2 mg/l, whichever is greater

Ammonium

0 to 100 mg/l as nitrogen

Nitrate

0 to 100 mg/l as nitrogen

Chloride

0.5 to 18,000 mg/l

Sodium

0.05 to 20,000 mg/l

Calcium

0 to 40,000 mg/l

Bromide

0 to 80,000 mg/l

Notes

Corrected for ionic strength (via conductivity readings); the accuracy specification relies on non-trivial maintenance practice and frequent calibration near the temperature of measurement; sensors require periodic tip replacement; maximum depth 15 m

 

 

SENSOR CONFIGURATION

 

Port numbers (see drawing)

1. Temperature
2. Wiper
3. Depth (optional)
4. Conductivity
5. Dissolved oxygen (DO)
6. pH reference
7. pH / ORP
8. Turbidity
9, 10, 11. Medium sensors (up to three)
12, 13, 14. Small sensors (up to three)

Sensor sizes

pH, pH reference, conductivity and DO are "small" sensors; fluorometers are "medium" sensors

Wiper

For a probe with a turbidity sensor, the turbidity wiper extension brush wipes the fluorometer and DO sensor. For probes with no turbidity, a wiper system is available to wipe DO and fluorometer sensors

Fluorometer substitution

Any fluorometer may be substituted for turbidity

Small sensor substitution

A small sensor may be substituted for any medium sensor

Port plugs

Any sensor may be removed and the port plugged

Optional sensors

ORP is combined with pH and is optional; depth is optional

 

 

SUPPLIED WITH

 

Included

Weighted sensor guard, storage/calibration cup, temperature sensor, internal data logging memory, marine connector, electronic manual, Windows-compatible Manta Control Software, maintenance kit, data cable, RS-232 to USB adaptor

 

 

OPTIONS AND ACCESSORIES

 

Anti-fouling sensor guards

Copper-gauze anti-fouling sensor guards

Underwater cables

3 to 200 meters in length (sold separately)

Battery packs

Attachable rechargeable battery packs; adding a battery pack converts the probe to a logging device

Wireless communication

mantaMobile™: rechargeable, high-capacity lithium battery with a Bluetooth® transceiver in a water-resistant case; communicates with Android and Apple® devices through the MantaLink™ app (free download for iOS and Android)

Other accessories

Flow-cells, cable reels for extra-long cables, SDI-12 and MODBUS adaptor cables, carry cases, pipe kits for continuous deployments, calibration standards including secondary calibration standards for fluorometers

Drawing

Manta-Plus-40-Multiparameter-Sonde-Diagram

Downloads

Manta+ Water Quality Probes Quick Reference Guide (EN) DownloadManta+ Water Quality Probes Quick Reference Guide (EN)
Manta+ Water Quality Probes Family Datasheet (EN) DownloadManta+ Water Quality Probes Family Datasheet (EN)