Anti-fouling multiparameter water quality sonde with a copper-infused silicone nose cone and a central wiper.

  • Standard sensors: temperature, pH, conductivity, optical dissolved oxygen and turbidity
  • Optional ORP, depth and up to three fluorometers
  • 200 m depth rating with over 1,000,000 logged readings at 1 Hz
Quantity
1

Overview

The Manta F35 is an anti-fouling multiparameter water quality sonde from the Manta+ family, designed for monitoring water quality in fresh, brackish and marine environments. It has the same standard sensors as the Manta+35, with a copper-infused silicone nose cone covering the sensor bodies and a central wiper that keeps the exposed sensor surfaces clean during continuous deployments.

Typical applications

  • Environmental monitoring: baseline water quality assessments, long-term monitoring of aquatic ecosystems, wetland health and restoration projects, and the effects of agricultural runoff
  • Drinking water monitoring: source water protection and monitoring, treatment plant optimization and process control, distribution system monitoring, and detecting and responding to contamination events
  • Stormwater monitoring: the effects of storm events on local waterways and the effectiveness of stormwater controls and treatment systems
  • Hydrology and hydrogeology: groundwater monitoring for aquifer characterization, surface water and groundwater interaction studies, and monitoring saltwater intrusion in coastal aquifers
  • Dredging and construction: monitoring turbidity during dredging operations, assessing the impact of construction projects, and monitoring runoff from construction sites
  • Mining: monitoring runoff and discharge, the effects of mining on ground and surface water, and the effects of acid mine drainage
  • Wastewater management: monitoring effluent quality for compliance, optimizing wastewater treatment processes, and tracking the impact of wastewater discharges
  • Oceanography and marine research: monitoring coastal water quality and ocean acidification, studying the health of coral reefs, marine ecosystems and biodiversity, and tracking the movement of water masses and currents
  • Aquaculture: maintaining optimal water quality for fish and shellfish farming, preventing disease outbreaks and optimizing growth
  • Industrial applications: process water monitoring and control, cooling water monitoring and treatment, and monitoring the environmental impact of industrial activities and effluents
  • Research and education: conducting scientific research on water quality and aquatic ecosystems, and providing hands-on learning experiences for students in environmental science

Key Features

  • Copper-infused silicone nose cone covers the sensor bodies and fits snugly over the sensors
  • Works in fresh and salt waters, for biological and non-biological foulants
  • Nose cone is easily removed if damaged or during sensor maintenance, and reduces the volume of solution needed for calibration
  • Central wiper keeps the exposed sensor surfaces clean during continuous deployments
  • Copper-less and silver-impregnated nose cones also available
  • Standard sensors: temperature, pH, conductivity, optical dissolved oxygen and turbidity
  • Optional ORP and depth sensors, plus up to three fluorometers (medium sensors)
  • Optical dissolved oxygen sensor with a replaceable cap and pH sensor with a refillable reference solution
  • Over 1,000,000 logged readings at a rate of 1 Hz
  • Communicates via RS-232, SDI-12, MODBUS, USB or Bluetooth
  • 200 m depth rating
  • LED status indicators for diagnostic information

Benefits

  • Better Data, Less Cleaning: the copper-infused nose cone and central wiper mean better data and less overall cleaning during continuous deployments.
  • Comprehensive Monitoring: measure multiple water quality parameters simultaneously, providing a complete picture of water health.
  • Versatile Use: suitable for applications from spot checking to continuous monitoring.
  • Real-time Data: connect to telemetry stations for immediate access to water quality data in the cloud.
  • Durable: anti-corrosive housing and sensors and robust marine bulkhead connectors.
  • Easy to Use: one touch and automatic data capture, simple calibrations, and mobile apps to collect and manage data.

Specifications

GENERAL SPECIFICATIONS

 

Standard sensors

Temperature, pH, conductivity, dissolved oxygen (optical), turbidity, anti-fouling wiper

Optional sensors

ORP (combined with pH sensor); depth, vented or absolute (occupies one medium port); up to three fluorometers (medium sensors)

Not available

PAR, CO2, TDG and ISE sensors

Number of sensors

Up to 11

Diameter

3.5"

Length (with guard)

No battery pack: 21"
With internal battery pack (IBP): 22.5"
With external Li-ion battery pack (EBP): 28.5"

Weight (with guard)

No battery pack: 7.75 lbs
With IBP: 8.75 lbs
With EBP: 10 lbs

Internal battery pack

6 "C" size batteries

Operating temperature

-5 to 50°C

Depth rating

200 m

Communication

RS-232, SDI-12, MODBUS, USB or Bluetooth®

Sample rate

1 Hz

Data memory

>1,000,000 logged readings

 

 

STANDARD SENSOR SPECIFICATIONS

 

Temperature

Range: -5 to 50°C
Resolution: 0.01
Accuracy: ±0.1
Calibration not required

pH

Range: 0 to 14 units
Resolution: 0.01
Accuracy: ±0.1 within 10°C of calibration; or 0.2
Refillable reference electrode; corrected for temperature; typical sensor life >6 years

Turbidity

Range: 0 to 1000 FNU; 1000 to 4000 FNU
Resolution: 0.01
Accuracy: ±0.3 FNU or ±2% of reading, whichever is greater (0 to 1000 FNU); ±4% of reading (1000 to 4000 FNU)
Filtered for non-turbidity spikes; includes wiper to clean the optics; FNU and NTU are interchangeable

Dissolved oxygen (optical) - concentration

Range: 0 to 20 mg/l, ±0.1; 20 to 30 mg/l, ±0.15; 30 to 50 mg/l, ±5% of reading
Resolution: 0.01
Compensated for temperature and salinity; EPA approved "lifetime" luminescence method; typical sensor cap life >6 years

Dissolved oxygen (optical) - % saturation

Range: 0 to 500% saturation
Resolution: 0.1
Accuracy: corresponds with the accuracy of the concentration reading

Conductivity - specific conductance (µS/cm)

Range: 0 to 5000 µS/cm
Resolution: 0.1
Accuracy: ±0.5% of reading or ±1, whichever is greater
Corrected for temperature; four easy-to-clean graphite electrodes; optional sensor provides ±0.5% of reading accuracy to 100 mS/cm

Conductivity - specific conductance (mS/cm)

Range: 0 to 100 mS/cm; 100 to 275 mS/cm
Resolution: 0.001
Accuracy: ±1% of reading ±0.001 (0 to 100 mS/cm); ±2% of reading (100 to 275 mS/cm)

Conductivity - salinity

Range: 0 to 70 PSU
Resolution: 0.01
Accuracy: ±2% of reading
Calculated from conductivity and temperature; PSU is equivalent to ppt

Conductivity - total dissolved solids (TDS)

Range: 0 to 65 g/l
Resolution: 0.1
Accuracy: ±5% of reading
Calculated from conductivity and temperature

 

 

OPTIONAL SENSOR SPECIFICATIONS

 

ORP

Range: -999 to 999 mV
Resolution: 0.1
Accuracy: ±20 mV
Optional ORP sensor is combined with pH sensor

Depth (absolute)

Range: 0 to 25 m, accuracy ±0.05; 0 to 200 m, accuracy ±0.4
Resolution: 0.01
Compensated for temperature and salinity

Depth (vented)

Range: 0 to 10 m
Resolution: 0.001
Accuracy: ±0.003
Compensated for temperature, salinity and barometric pressure

Barometric pressure

Range: 400 to 900 mm Hg
Resolution: 0.1
Accuracy: ±1.5
Included with depth sensor

 

 

FLUOROMETERS (MEDIUM SENSORS, UP TO THREE)

 

Resolution and linearity

Resolution: 0.01; linearity of 0.99 R²

Chlorophyll a - blue

0 to 100 µg/l

Chlorophyll a - red

0 to 500 µg/l

Rhodamine dye

0 to 200 ppb

Phycocyanin (freshwater BGA)

0 to 4500 ppb

Phycoerythrin (marine BGA)

0 to 700 ppb

CDOM/FDOM

0 to 500 ppb

Optical brightener

0 to 300 ppb

Tryptophan

0 to 5000 ppb

Fluorescein dye

0 to 150 ppb

PTSA

0 to 650 ppb

Refined oil

0 to 20 ppm

Crude oil

0 to 300 ppb

Caution

Never look directly at a fluorometer sensor. The UV rays emitted by the sensor can cause eye damage.

 

 

SENSOR CONFIGURATION

 

Sensor sizes

pH, pH reference, conductivity and DO are "small" sensors; fluorometers are "medium" sensors

Wiper

For a probe with a turbidity sensor, the turbidity wiper extension brush wipes the fluorometer and DO sensor. For probes with no turbidity, a wiper system is available to wipe DO and fluorometer sensors

Fluorometer substitution

Any fluorometer may be substituted for turbidity

Port plugs

Any sensor may be removed and the port plugged

Optional sensors

ORP is combined with pH and is optional; depth is optional and occupies one medium port

 

 

SUPPLIED WITH

 

Included

Weighted sensor guard, storage/calibration cup, temperature sensor, internal data logging memory, marine connector, electronic manual, Windows-compatible Manta Control Software, maintenance kit, data cable, RS-232 to USB adaptor

 

 

OPTIONS AND ACCESSORIES

 

Nose cone options

Copper-less nose cone; silver-impregnated nose cone

Underwater cables

3 to 200 meters in length (sold separately)

Battery packs

Attachable rechargeable battery packs; adding a battery pack converts the probe to a logging device

Wireless communication

mantaMobile™: rechargeable, high-capacity lithium battery with a Bluetooth® transceiver in a water-resistant case; communicates with Android and Apple® devices through the MantaLink™ app (free download for iOS and Android)

Other accessories

Flow-cells, cable reels for extra-long cables, SDI-12 and MODBUS adaptor cables, carry cases, pipe kits for continuous deployments, calibration standards including secondary calibration standards for fluorometers

Drawing

Manta-F35-Anti-Fouling-Water-Quality-Sonde-Diagram

Downloads

Manta+ Water Quality Probes Family Datasheet (EN) DownloadManta+ Water Quality Probes Family Datasheet (EN)
Manta+ Water Quality Probes Quick Reference Guide (EN) DownloadManta+ Water Quality Probes Quick Reference Guide (EN)