WETEC

BARANI DESIGN

MeteoTemp RH+T+PRESSURE 3-in-1 Probe


3-in-1 temperature, humidity and barometric pressure probe with ultra-low power consumption.

  • Temperature, humidity, dew point, frost point and pressure outputs

  • Sealed pressure sensor, 300…1100 hPa, adds only approx. 20 µA

  • RS485 Modbus-RTU / ASCII with 3-stage lightning protection

Quantity
1

Overview

The BARANI MeteoTemp RH+T+PRESSURE is a 3-in-1 sensor probe that combines a sealed barometric pressure sensor with the WMO temperature and humidity performance of the MeteoTemp RH+T. Combined output features temperature, relative humidity, dew point, frost point and barometric pressure. It has the same form factor as the 2-in-1 probe, and adding the pressure sensor does not change its ultra-low power performance: total consumption stays around 800 µA, with only approx. 20 µA added by the pressure sensor.

Temperature is measured from -45 to +105 °C with ± 0.10 °C accuracy between 0 and +65 °C, humidity across 0…100 %RH, and pressure from 300 to 1100 hPa. A replaceable, self-cleaning PTFE filter cap keeps the sensing elements clean and speeds humidity recovery after rain. Output is RS485 with Modbus-RTU or ASCII protocol, with 3-stage lightning, surge and ESD protection. Calibration coefficients are stored in the sensor for easy recalibration and traceability. Combined with the helical MeteoShield, MeteoTemp achieves some of the lowest air temperature uncertainties available.

Typical applications

  • Meteorology and weather stations

  • Airports

  • Agriculture

  • Hydrology

  • Road management and roadside sensors

  • Buoy and marine

  • Ships

  • Coastal sites

  • Polar and winter sites

  • Ski lift and snow making

  • Smart cities

  • Industrial SCADA and PLC systems

Key Features

  • Temperature, humidity, dew point, frost point and barometric pressure outputs

  • Integrated sealed barometric pressure sensor, 300…1100 hPa

  • Meets WMO requirements for air temperature and humidity

  • ± 0.10 °C temperature accuracy (0…+65 °C)

  • Ultra-low power consumption, pressure sensor adds only approx. 20 µA

  • Replaceable, self-cleaning PTFE filter cap

  • Corrosion-resistant, hydrophobic white sensor body

  • RS485 output with Modbus-RTU or ASCII protocol

  • 3-stage lightning, surge, transient and ESD protection

  • Calibration coefficients stored in the sensor

  • Waterproof bayonet locking connector

Benefits

  • 3 sensors in 1 simplify your systems

  • Low measurement uncertainty for accurate air temperature

  • Extends weather station battery life

  • Smaller solar panel for lower weather station cost

  • Quick humidity recovery after rain or wave impact on buoys

  • Easy recalibration and long-term traceability

Specifications

TEMPERATURE

 

Sensor type

Silicon

Measuring range

-45…+105 °C

Accuracy

± 0.10 °C (0…+65 °C); below 0 °C increases linearly to ± 0.20 °C at -45 °C; ± 0.20 °C maximum over full range

Resolution

0.01 °C

Stability

< 0.02 °C per year

Response time

5…30 s (τ63%, excludes filter-cap effects)

Meets WMO

Yes

 

 

RELATIVE HUMIDITY

 

Sensor type

Capacitive

Measuring range

0…100 %RH

Accuracy (0…80 %RH)

-45…0 °C: ± 3.95 → ± 2.05 %RH; 0…+25 °C: ± 2.05 → ± 1.30 %RH; +25…+50 °C: ± 1.30 → ± 2.05 %RH

Accuracy (80…100 %RH)

Add ± 0.025 %RH for each 1 %RH above 80 %RH (up to + 0.50 %RH at 100 %RH)

Hysteresis

Short-term ≤ ± 1.0 %RH additional

Resolution

0.01 %RH

Stability

< 0.25 %RH per year

Response time

8…30 s (τ63%, excludes filter-cap effects)

Meets WMO

Yes

 

 

DEW POINT / FROST POINT

 

Accuracy (calculated at 90 %RH)

± 0.39 °C DPw at +25 °C; ± 0.46 °C DPw / ± 0.40 °C FPi at 0 °C; road-NaCl reference: ± 0.44 °C FPi at -7 °C

Resolution

0.01 °C

Output range

Calculated, dependent on temperature and RH

Response time

8…30 s

Meets WMO

Yes

 

 

BAROMETRIC PRESSURE

 

Sensor type

Piezo-resistive, sealed

Measuring range

300…1100 hPa

Accuracy

± 0.5 hPa (0…55 °C and 750…1100 hPa); else ± 0.8 hPa

Resolution

0.02 hPa

Stability

0.5 hPa per year

Response time

0.1 s

Meets WMO

No

 

 

OUTPUTS

 

Digital output

RS485 with Modbus-RTU or ASCII protocol

Connection

Bayonet connector with silicone O-ring

Cable wiring

White = GND; Brown = VCC; Green = A (non-inverting); Yellow = B (inverting)

 

 

POWER REQUIREMENT

 

Power supply

5…24 Vdc with reverse polarity protection (startup in-rush < 100 mA for 1 ms)

Power consumption

800 µA at 1 Hz output incl. RS485 communication (pressure sensor adds only approx. 20 µA); 25 µA quiescent

 

 

PROTECTION

 

Lightning and surge protection

3-stage, per IEC EN 61000-4-2, EN 61000-4-4, EN 61000-4-5 on both data and power lines; Surge, EFT/Burst

Surge test

4 kV / 2 kA acc. to EN50082-2 / IEC61000-6-2

ESD

15 kV

 

 

ENVIRONMENTAL

 

Operating conditions

-45…+105 °C / 0…100 %RH

Protection degree

IP66W (DIN 40050)

 

 

MECHANICAL

 

Dimensions

Length 133 mm (164 mm with mating connector); Ø 16 mm base; Ø 12 mm PTFE sensor cap

Filter cap

Replaceable screw-on PTFE (Teflon) filter cap

Sensor body

Corrosion-resistant, hydrophobic white body

 

 

CALIBRATION

 

Calibration equation

From a simple offset value up to x⁴ calibration equation

Coefficients

Up to 9 significant digits, stored in the sensor; read out and programmed via Modbus

Drawing

MeteoTemp-RH-T-Pressure-Dimensional-Drawing

Downloads

MeteoTemp-RH-T-Pressure-Datasheet-EN DownloadMeteoTemp-RH-T-Pressure-Datasheet-EN