Thermal flow and temperature transmitter for liquids

  • Calorimetric (thermal) measuring principle with no moving parts and no wear, measuring flow from 2...300 cm/s plus medium temperature, with an integrated volume meter with preset.

  • Fast response time of approx. 1-2 s in water (25 °C), with several sensor and process connection versions and an oil-filled option.

  • 1.2 inch graphic LCD, IO-Link, 1 analog output and 2 configurable switching outputs in a waterproof all-metal housing, IP65 / IP67.

Quantity
1

Overview

The Senseca OMNIPLUS-F is a flow transmitter for water and other liquids with integrated temperature measurement. It works on a calorimetric (thermal) principle, measures flow from 2...300 cm/s and has a fast response time.

It has a 1.2 inch graphic LCD, IO-Link, 1 analog output and 2 configurable switching outputs, and is built in a waterproof all-metal housing with a mineral glass front. Several sensor versions are available, from push-in sensors to threaded and sealing cone connections.

Typical applications

  • Flow measurement of water

  • Flow measurement of other liquids

  • Medium temperature measurement in liquids

  • Flow monitoring with limit switch, frequency or pulse outputs

  • Volume metering with preset counter

  • Installations with high air humidity and heavily fluctuating temperatures (oil-filled version)

  • Flow transmitter for water and other liquids with integrated temperature measurement

  • Measuring range 2...300 cm/s

  • Fast response time

  • Integrated volume meter with preset

  • Measuring method without moving parts, no wear

  • Wetted material stainless steel

  • 1 analog output (20 mA / 10 V)

  • 2 versatile configurable switching outputs

  • IO-Link interface

  • High-resolution graphic LCD

  • Parameters programmable with multifunction ring

  • Waterproof all-metal housing (IP67)

  • Mineral glass front, scratch-proof and chemically resistant

Benefits

  • No moving parts, so no wear

  • Fast response time, approx. 1-2 s in water (25 °C)

  • Flow and medium temperature measured by one device

  • Switching outputs configurable as limit switch, frequency output, pulse output or signal output for preset counter

  • IO-Link with block parameterization and data storage

  • Several sensor and process connection versions to fit different installations

  • Oil-filled version for high air humidity and heavily fluctuating temperatures

Specifications

INPUT

 

Measuring principle

Calorimetric / thermal

Measurement media

Liquids; media configurable on the device

Measuring range, flow

2...300 cm/s

Measuring range, temperature

Corresponding to the specified media temperature range

Signal input

Reset input for volume meter (alternative to switch output at Pin 5)

 

 

MEASURING ACCURACY

 

Measurement uncertainty, flow

±(7 % measurement + 2 % FS)

Measurement uncertainty, temperature

±2 °C

Condition

Flow speed > 2 cm/s

Response time

Approx. 1-2 s in water (25 °C) at an average flow speed

 

 

OUTPUTS

 

Analog output, current

4...20 mA; 0...20 mA

Analog output, voltage

0...10 V; 2...10 V; 0...5 V; 1...5 V; 0.5...4.5 V

Switching outputs

2 transistor outputs push-pull, parameterizable as NPN o.C.; short-circuit and reverse polarity resistant; Iout = 100 mA max per output

Switching output functions

Configurable on the device as limit switch, frequency output, pulse output or signal output for preset counter

 

 

IO-LINK

 

IO-Link revision

V1.1

Bit rate

COM2 (38400 Bit/s)

Minimum cycle time

4 ms

SIO mode

Yes

Port class

A compatible

Block parameterization

Yes

Data storage

Yes

 

 

DISPLAY

 

Display

Graphic 1.2 inch LCD (transflective), 128 x 64 pixel

Backlight

White, red on alarm notification

Operation

Parameters programmable with multifunction ring

 

 

POWER REQUIREMENT

 

Supply voltage

18...30 V DC

Current consumption

< 130 mA (SIO mode, outputs not under load)

 

 

ENVIRONMENTAL

 

Media temperature, connection types 008 / 013 / 015 / 018

-20...+90 °C

Media temperature, connection type 012

-20...+100 °C

Operating pressure, connection types 008 / 015

PN100 (PN200 on request)

Operating pressure, connection types 013 / 018

Up to PN100 (depending on T-piece)

Operating pressure, connection type 012

Up to PN40 (depending on assembly material)

Ambient temperature

-20...+70 °C

Storage temperature

-30...+80 °C

Protection class

IP65 / IP67

 

 

MATERIALS

 

Wetted materials

Stainless steel 1.4571; fluororubber (O-ring for connection type 013)

Housing

Stainless steel 1.4305

Glass

Mineral glass, hardened

Magnet

NeFeB

Ring

POM

 

 

MECHANICAL

 

Electrical connection

M12x1 circular connector, 5-pin

Pin assignment

1 BN: L+; 2 WH: Analog; 3 BU: L-; 4 BK: IO-Link/DO; 5 GY: DI/DO

Housing diameter

Ø50 mm

Weight

Approx. 0.30 kg

Sensor 012VK100 / 012VK150

Push-in sensor Ø12, L = 100 mm / L = 150 mm (other lengths on request)

Sensor 008HK028

Thread G ¼ A, L = 28.0 mm

Sensor 015HK029

Thread G ½ A, L = 29.6 mm

Sensor 015HK045

Thread G ½ A, L = 45.0 mm

Sensor 013TK031

System fastening Ø13.2, L = 31 mm, for T-piece with connection G 3/8 and G ½

Sensor 013TK037

System fastening Ø13.2, L = 37 mm, for T-piece with connection G ¾...G 2

Sensor 018DK045

Sealing cone 24°, M18 x 1.5, L = 45 mm

Oil-filled version (code T)

For applications with high air humidity and heavily fluctuating temperatures

 

 

APPROVALS

 

Conformity

CE

Drawing

Senseca-OMNIPLUS-F-Dimensional-Drawing

Downloads

Senseca-OMNIPLUS-F-Datasheet-EN DownloadSenseca-OMNIPLUS-F-Datasheet-EN