WETEC

ECD

CD80 Chlorine Dioxide Analyzer


Reagentless, panel-mounted chlorine dioxide analyzer for continuous water quality monitoring.

  • Measurement range: 0.05…20 ppm (standard) or 0.01…5 ppm (low range)

  • Automatic flow control and integrated T80 transmitter

  • 4…20 mA outputs and Modbus RTU communication

  • Optional pH sensor and automatic cleaning system

Quantity
1

Overview

The ECD CD80 is a reagentless, panel-mounted analyzer designed for continuous chlorine dioxide (ClO₂) measurement in drinking water, industrial cooling water, rinse water and other freshwater samples. It uses an amperometric sensor with a PTFE membrane, gold cathode and silver/silver chloride anode, offering a standard measurement range of 0.05…20 ppm or a low-range option of 0.01…5 ppm.

The system incorporates a constant head flow controller, flow cell and T80 transmitter to simplify installation and maintain stable sample flow. It supports 4…20 mA outputs, Modbus RTU communication and three configurable relays, with 24 Vdc or 100…240 Vac power supply options. An optional pH sensor and automatic spray cleaning system are available.

Typical Applications

  • Chlorine dioxide monitoring in drinking water

  • Disinfectant monitoring in industrial cooling water

  • Chlorine dioxide measurement in industrial rinse water

  • Chlorine dioxide monitoring in wastewater

  • Continuous monitoring of chlorine dioxide in other freshwater samples

Key Features

  • Reagentless amperometric chlorine dioxide sensor

  • Standard measurement range of 0.05…20 ppm or low range of 0.01…5 ppm

  • PTFE membrane with gold cathode and silver/silver chloride anode

  • Panel-mounted, plumb-and-play design

  • Automatic flow control with a constant head flow controller

  • Sample flow range of 38…300 L/h

  • Sample operating temperature range of 0…50 °C

  • Response time of T90 in 2 minutes

  • Electrolyte life of up to 12 months

  • Integrated T80 transmitter with graphical LCD display

  • 4…20 mA outputs and Modbus RTU communication

  • Three user-configurable alarm relays

  • IP65 transmitter enclosure

  • 24 Vdc or 100…240 Vac power supply options

  • Optional pH sensor and automatic spray cleaning system

Benefits

  • Reduced operating costs through reagentless measurement

  • Simplified installation with a preassembled, panel-mounted system

  • Stable sample flow without external pressure regulators or rotameters

  • Reduced risk of clogging and blockages with large flow tubes and cells

  • Extended maintenance intervals with electrolyte life of up to 12 months

  • Easy membrane replacement and electrolyte replenishment without tools

  • Flexible integration with external control and monitoring systems

  • Reduced sensor fouling with the optional automatic spray cleaning system

Specifications

Measurement

Measuring principle

Amperometric (polarographic)

Measured parameter

Chlorine dioxide (ClO₂)

Standard measurement range

0.05…20 ppm

Low measurement range

0.01…5.00 ppm

pH measurement range

0…14 pH (with pH sensor)

pH compensation range

pH 5…10 (accuracy degrades rapidly above pH 9)

Response time

T90 in 2 minutes

 

 

Sensor

Cathode

Gold

Anode

Silver/silver chloride

Membrane

PTFE

Electrolyte life

Up to 12 months

pH sensor protocol

Digital S80 protocol

pH sensor body

316L stainless steel

pH electrode

Replaceable electrode cartridge

 

 

Sample and Flow Requirements

Flow control

Constant Head Flow Controller (CHFC)

Minimum sample flow

38 L/h (10 gal/h)

Maximum sample flow

300 L/h (80 gal/h)

Sample pressure

1…30 psig (0.1…2 bar)

Sample operating temperature

0…50 °C (32…122 °F)

Wetted materials

PVC, PP, PVDF, PTFE, glass and 316 stainless steel

Sample inlet connection

¼″ FNPT with barb fitting

Drain connection

¾″ FNPT

 

 

Display

Transmitter

T80

Display type

LCD

Display resolution

128 × 64 pixels

Display dimensions

2.75″ × 1.5″

Display colour

Black on grey background

Backlit display colour

Blue on white background

Graphical display modes

Line, gauge and bar

 

 

Outputs and Communication

Chlorine dioxide analogue output

1 × 4…20 mA, set to sensor range

pH analogue output

1 × 4…20 mA, set to 0…14 pH (optional)

Digital communication

Modbus RTU (standard)

 

 

Alarm Relays

Number of relays

3

Relay type

SPDT, Form C

Maximum relay voltage

250 Vac

Maximum relay current

10 A (resistive load)

Relay functions

High/low alarms with expiration timer, periodic timers or fault alarms

 

 

Power Supply

DC power supply

24 Vdc

DC input voltage range

18…36 Vdc

DC power requirement

250 mW minimum

AC power supply

100…240 Vac

AC frequency

50/60 Hz

AC power consumption

4 W

Fuse protection

250 V, 1 A, slow-blow fuse (AC model)

 

 

Mechanical

System mounting

Wall-mounted panel

Transmitter enclosure material

Polycarbonate

Enclosure colour

Beige

Enclosure protection

IP65, weatherproof

Enclosure format

½ DIN

Transmitter dimensions (L × W × D)

14.4 × 14.4 × 9.0 cm (5.7″ × 5.7″ × 3.5″)

 

 

Environmental Conditions

Ambient temperature (DC model)

−20…+70 °C

Ambient temperature (AC model)

−20…+60 °C

Storage temperature

−30…+85 °C

Relative humidity

0…80% up to 31 °C, decreasing linearly to 50% at 40 °C

Maximum operating altitude

2,000 m (6,500 ft)

Mains supply voltage fluctuation

Up to ±10% of nominal voltage

Transient overvoltage category

CAT II

Pollution degree

2

Outdoor installation

Protective cover or sunshield required

 

 

Optional Configurations

pH sensor configuration

With or without stainless steel pH sensor

Automatic cleaning

With or without spray cleaner

Flow controller configuration

Standard CHFC or high CHFC

Downloads

CD80 Instruction Manual DownloadCD80 Instruction Manual