ECD
DCA-23 Seawater Dechlorination Analyzer
Seawater chlorination and dechlorination analyzer with dual-input potentiometric measurement.
Chlorine equivalent measurement range of 0.00…2.00 ppm
Minimum detection limit of 0.014 ppm
No added chemicals required
Flow-independent measurement
4…20 mA output with optional alarm relays
Rack-mounted system for easy installation
Overview
The ECD DCA-23 is a seawater chlorination and dechlorination analyzer featuring a dual-input potentiometric design. It measures pH, temperature and total residual oxidant (TRO) potential to determine the equivalent chlorine concentration over a range of 0.00…2.00 ppm, with a minimum detection limit of 0.014 ppm and a limit of quantification of 0.044 ppm.
The analyzer requires no added chemicals and is insensitive to flow variations. It is available as a complete rack-mounted system incorporating a T23 transmitter, sensors, piping and valves, or as separate components for custom installations. It provides 4…20 mA output with optional alarm relays and supports loop-powered, 24 Vdc, 110 Vac or 220 Vac configurations. The analyzer is specifically designed for seawater and is not suitable for freshwater applications.
Typical Applications
Seawater cooling systems in power plants and refineries
Seawater chlorination and dechlorination at desalination plants
Saltwater air conditioning (SWAC) systems
Seawater heat exchangers at LNG regasification terminals
Residual oxidant monitoring for seawater dechlorination compliance
Key Features
Dual-input potentiometric measurement design
Chlorine equivalent measurement range of 0.00…2.00 ppm
Minimum detection limit (MDL) of 0.014 ppm
Limit of quantification (LOQ) of 0.044 ppm
Measurement of pH, temperature and total residual oxidant (TRO) potential
No added chemicals required
Flow-independent measurement
T23 transmitter with 2 × 16-character LCD
4…20 mA analogue output with optional dual SPDT alarm relays
Loop-powered, 24 Vdc, 110 Vac or 220 Vac configurations
NEMA 4X transmitter enclosure
Complete Uni-Strut rack-mounted system
Schedule 80 PVC piping
Benefits
No added chemicals required during measurement
Measurement unaffected by flow variations
Low detection limit for monitoring low residual oxidant concentrations
Easy installation with a preassembled rack-mounted system
Simple calibration by comparison with a free chlorine grab sample test
Flexible power supply and output configurations to suit different installation requirements
Separate components available for custom installations
Minimal maintenance requirements
Specifications
Measurement Performance | |
|---|---|
Measured parameters | pH, TRO potential and temperature |
Chlorine equivalent measurement range | 0.00…2.00 ppm |
Minimum detection limit (MDL) | 0.014 ppm |
Limit of quantification (LOQ) | 0.044 ppm |
Flow sensitivity | None |
Added chemicals | Not required |
Calibration method | Comparison with a free chlorine grab sample test |
Measurement medium | Seawater only |
pH Sensor | |
|---|---|
Sensor model | PHS10-T23-CBL-EG-75PP |
Measurement range | 0…14 pH |
Temperature range | 0…80 °C |
Pressure range | 0…100 psig |
Wetted materials | 316 stainless steel, PES, Teflon, glass, Viton |
TRO Sensor | |
|---|---|
Sensor model | MVS10-T23-CBL-EG-75PP |
Measurement range | ±1500 mV |
Temperature range | 0…80 °C |
Pressure range | 0…100 psig |
Wetted materials (listed under TRS10 in datasheet) | 316 stainless steel, PES, Teflon, platinum, Viton |
Temperature Measurement | |
|---|---|
Measurement range | 0…100 °C |
T23 Transmitter | |
|---|---|
Display | LCD, 2 × 16 characters |
Display size | 4 × 1 in |
Enclosure rating | NEMA 4X |
Enclosure dimensions (L × W × D) | 5.7 × 5.7 × 3.5 in |
Keypad | Membrane switch keypad |
Cable glands | 2 × ½ in |
Power Supply Part Number | |
|---|---|
1290220-1 | Loop powered |
1290200-2 | 24 Vdc (12…36 Vdc) at 0.05 A |
1290200-3 | 24 Vdc (12…36 Vdc) at 0.10 A |
1290200-4 | 110 Vac, 50/60 Hz |
1290200-5 | 220 Vac, 50/60 Hz |
Analogue Outputs | |
|---|---|
Output signal | 4…20 mA |
Maximum load | 525 Ω |
Output 1 | Chlorine, 0.00…2.00 ppm (all models) |
Output 2 | pH, 0…14 pH (1290200-2, -3, -4 and -5) |
Alarm Relays | |
|---|---|
Number of relays | 2 (optional) |
Applicable models | 1290200-3, -4 and -5 |
Relay type | SPDT |
Maximum AC contact rating | 230 Vac / 5 A |
Maximum DC contact rating | 30 Vdc / 5 A |
Load type | Resistive |
Flow Train and Process Fittings | |
|---|---|
Compression fitting size | ¾ in |
Compression fitting materials | Polypropylene (PP), Viton |
Pipe size | ¾ in |
Pipe tee size | 1 in |
Pipe and tee material | Schedule 80 PVC |
Inlet and drain connections | Slip fittings |
Rack-Mounted System | |
|---|---|
System components | T23 analyzer, two sensors, pH flow cell and TRO flow cell |
Mounting rail | Uni-Strut rail |
Mounting rail size | 1⅝ × 1⅝ in |
Maximum system pressure and temperature rating | 50 psig at 60 °C |
System dimensions (L × H × D) | 26 × 20 × 8.5 in |
Shipping Weight | |
|---|---|
T23 transmitter | 4.8 lb |
PHS10 or TRS10 sensor | 2.5 lb |
Complete rack-mounted system | 18 lb |
Drawing

Downloads
| DCA-23 Instruction Manual | DownloadDCA-23 Instruction Manual |
|---|
Related Products

ECD
DC80 Dechlorination Analyzer
The DC80 Dechlorination Analyzer uses a zero-shift measurement strategy, automatic flow control and pH compensation to monitor very low total chlorine levels…
Learn More
ECD
CD80 Chlorine Dioxide Analyzer
The CD80 is a reagentless, panel-mounted analyzer for continuous chlorine dioxide (ClO₂) monitoring from 0.05 to 20 ppm, featuring automatic flow control, a…
Learn More
ECD
TC80 Total Chlorine Analyzer
The TC80 is a reagentless, panel-mounted analyzer for total chlorine monitoring in water, featuring amperometric sensing, automatic pH compensation, automatic…
Learn More
ECD
FC80 Free Chlorine Analyzer
The FC80 Free Chlorine Analyzer is a reagentless, panel-mounted system for continuous free chlorine monitoring in water, featuring automatic pH compensation…
Learn More