WETEC

ECD

FC80 Free Chlorine Analyzer


Reagentless, panel-mounted analyzer for continuous free chlorine monitoring in water.

  • Free chlorine measurement ranges: 0.05–20 ppm (high range) or 0.01–5.00 ppm (low range)

  • Automatic pH compensation and flow control

  • Integrated free chlorine sensor, pH sensor and T80 transmitter

  • 4–20 mA output and Modbus RTU communication

  • Compliant with EPA Method 334.0

  • Optional automatic cleaning system

Quantity
1

Overview

The ECD FC80 Free Chlorine Analyzer is a reagentless, panel-mounted system designed for continuous monitoring of free chlorine in drinking water, industrial cooling water, rinse water, wastewater and other freshwater applications. It uses an amperometric chlorine sensor with a PTFE membrane and an S80 pH sensor for automatic pH compensation. High- and low-range configurations cover 0.05–20 ppm and 0.01–5.00 ppm, respectively.

The system incorporates a constant head flow controller, sensor flow cells and a T80 transmitter. It supports 4–20 mA output and Modbus RTU communication, with 24 VDC or 100–240 VAC power supply options. The analyzer is factory calibrated, complies with EPA Method 334.0 and is available with an optional automatic cleaning system.

Typical Applications

  • Drinking water disinfection monitoring

  • Industrial cooling water monitoring

  • Industrial rinse water monitoring

  • Wastewater free chlorine monitoring

  • Chlorine monitoring in other freshwater systems

Key Features

  • Reagentless free chlorine measurement

  • Panel-mounted system with integrated T80 transmitter

  • Amperometric chlorine sensor with gold cathode, silver/silver chloride anode and PTFE membrane

  • High-range configuration: 0.05–20 ppm

  • Low-range configuration: 0.01–5.00 ppm

  • pH measurement range of 0–14 pH

  • Automatic pH and temperature compensation

  • Constant head flow controller for automatic sample flow regulation

  • Sample operating temperature range of 0–50 °C

  • 4–20 mA output and Modbus RTU communication

  • 24 VDC or 100–240 VAC power supply options

  • IP65 transmitter enclosure

  • Factory-calibrated system

  • Compliant with EPA Method 334.0

  • Optional automatic cleaning system

Benefits

  • Lower operating costs by eliminating the need for measurement reagents

  • Simplified installation with a preassembled, panel-mounted design

  • Improved measurement consistency through automatic pH and temperature compensation

  • Stable sample flow without additional pressure regulators or rotameters

  • Reduced risk of clogging and blockages with large flow tubes and cells

  • Easy integration with monitoring and control systems

  • Simplified commissioning through factory calibration

  • Extended intervals between electrolyte refills of up to 12 months

  • Easy, tool-free replacement of the chlorine sensor membrane

  • Reduced manual cleaning requirements with the optional automatic cleaning system

Specifications

Sensors and Flow Train

Measuring principle

Three-electrode amperometric sensing

Chlorine sensor

Potentiostatic sensor with digital protocol

Chlorine sensor electrodes

Gold cathode, silver/silver halide anode and stainless steel counter electrode

Sensor membrane

Microporous PTFE

pH sensor

Digital S80 protocol, 316L stainless steel body with replaceable electrode cartridge

Chlorine measurement range (high range)

0.05–20 ppm

Chlorine measurement range (low range)

0.005–2.00 ppm

pH measurement range

pH 4–14

Operating temperature

0–50 °C (32–122 °F)

Sample flow rate

38–300 L/h (10–80 gal/h)

Wetted materials

PVC, PP, PVDF, PTFE, glass and 316 stainless steel

Process inlet connection

¼″ FNPT with barb fitting

Process drain connection

¾″ FNPT

Response time

T90 in 2 minutes

Electrolyte life

3–6 months

 

 

TC80 Analyzer – Measurements and Display

Chlorine measurement range

0.00 ppb–22.00 ppm

pH measurement range

pH 0.00–14.00

pH compensation of total chlorine

pH 4–12

Display

LCD with LED backlight

Display resolution

128 × 64 pixels

Display dimensions

2.75″ × 1.5″

Display colour

Black on grey background; blue on white background with LED backlight on

Graphical display options

Bar, gauge or line

 

 

Outputs and Communication

Total chlorine output

One 4–20 mA output, set to the sensor measurement range

pH output

One optional 4–20 mA output, set to pH 0–14

Digital communication

Modbus RTU via RS485 (standard)

Alarm relays

Three SPDT, Form C relays

Relay configuration

High/low alarms with expiration timer, periodic timers or fault alarms

 

 

Input Power

DC power supply

24 VDC (18–36 VDC)

DC power requirement

250 mW minimum

AC power supply

100–240 VAC, 50/60 Hz

AC power consumption

4 W

Fuse protection

250 V, 1 A, slow-blow fuse

 

 

Enclosure

Enclosure material

Polycarbonate

Enclosure colour

Beige

Protection rating

IP65, weatherproof

Enclosure format

½ DIN

Dimensions (L × W × D)

14.4 × 14.4 × 9.0 cm (5.7″ × 5.7″ × 3.5″)

 

 

Environmental Conditions

Installation environment

Outdoor use with protective cover or sunshield

Ambient temperature (24 VDC models)

−20 to +70 °C

Ambient temperature (100–240 VAC models)

−20 to +60 °C

Storage temperature

−30 to +85 °C

Relative humidity

0–80% up to 31 °C, decreasing linearly to 50% RH at 40 °C

Maximum altitude

2,000 m (6,500 ft)

Mains supply voltage fluctuation

Up to ±10% of nominal voltage

Transient overvoltage category

CAT II

Pollution degree

2

 

 

Sample Requirements

Sample pressure

1–30 psig (0.1–2 bar)

 

 

System Configuration and Features

System configuration

Panel-mounted system

Integrated components

Total chlorine sensor, pH sensor, constant head flow controller, flow cells and T80 transmitter

Measurement technology

Reagentless

Flow control

Automatic

Calibration

Factory calibrated

Compliance

EPA Method 334.0

Automatic cleaning

Optional spray cleaner for the chlorine flow cell

Downloads

FC80 Instruction Manual DownloadFC80 Instruction Manual