ECD
FC80 Free Chlorine Analyzer
Reagentless, panel-mounted analyzer for continuous free chlorine monitoring in water.
Free chlorine measurement ranges: 0.05–20 ppm (high range) or 0.01–5.00 ppm (low range)
Automatic pH compensation and flow control
Integrated free chlorine sensor, pH sensor and T80 transmitter
4–20 mA output and Modbus RTU communication
Compliant with EPA Method 334.0
Optional automatic cleaning system
Overview
The ECD FC80 Free Chlorine Analyzer is a reagentless, panel-mounted system designed for continuous monitoring of free chlorine in drinking water, industrial cooling water, rinse water, wastewater and other freshwater applications. It uses an amperometric chlorine sensor with a PTFE membrane and an S80 pH sensor for automatic pH compensation. High- and low-range configurations cover 0.05–20 ppm and 0.01–5.00 ppm, respectively.
The system incorporates a constant head flow controller, sensor flow cells and a T80 transmitter. It supports 4–20 mA output and Modbus RTU communication, with 24 VDC or 100–240 VAC power supply options. The analyzer is factory calibrated, complies with EPA Method 334.0 and is available with an optional automatic cleaning system.
Typical Applications
Drinking water disinfection monitoring
Industrial cooling water monitoring
Industrial rinse water monitoring
Wastewater free chlorine monitoring
Chlorine monitoring in other freshwater systems
Key Features
Reagentless free chlorine measurement
Panel-mounted system with integrated T80 transmitter
Amperometric chlorine sensor with gold cathode, silver/silver chloride anode and PTFE membrane
High-range configuration: 0.05–20 ppm
Low-range configuration: 0.01–5.00 ppm
pH measurement range of 0–14 pH
Automatic pH and temperature compensation
Constant head flow controller for automatic sample flow regulation
Sample operating temperature range of 0–50 °C
4–20 mA output and Modbus RTU communication
24 VDC or 100–240 VAC power supply options
IP65 transmitter enclosure
Factory-calibrated system
Compliant with EPA Method 334.0
Optional automatic cleaning system
Benefits
Lower operating costs by eliminating the need for measurement reagents
Simplified installation with a preassembled, panel-mounted design
Improved measurement consistency through automatic pH and temperature compensation
Stable sample flow without additional pressure regulators or rotameters
Reduced risk of clogging and blockages with large flow tubes and cells
Easy integration with monitoring and control systems
Simplified commissioning through factory calibration
Extended intervals between electrolyte refills of up to 12 months
Easy, tool-free replacement of the chlorine sensor membrane
Reduced manual cleaning requirements with the optional automatic cleaning system
Specifications
Sensors and Flow Train | |
|---|---|
Measuring principle | Three-electrode amperometric sensing |
Chlorine sensor | Potentiostatic sensor with digital protocol |
Chlorine sensor electrodes | Gold cathode, silver/silver halide anode and stainless steel counter electrode |
Sensor membrane | Microporous PTFE |
pH sensor | Digital S80 protocol, 316L stainless steel body with replaceable electrode cartridge |
Chlorine measurement range (high range) | 0.05–20 ppm |
Chlorine measurement range (low range) | 0.005–2.00 ppm |
pH measurement range | pH 4–14 |
Operating temperature | 0–50 °C (32–122 °F) |
Sample flow rate | 38–300 L/h (10–80 gal/h) |
Wetted materials | PVC, PP, PVDF, PTFE, glass and 316 stainless steel |
Process inlet connection | ¼″ FNPT with barb fitting |
Process drain connection | ¾″ FNPT |
Response time | T90 in 2 minutes |
Electrolyte life | 3–6 months |
TC80 Analyzer – Measurements and Display | |
|---|---|
Chlorine measurement range | 0.00 ppb–22.00 ppm |
pH measurement range | pH 0.00–14.00 |
pH compensation of total chlorine | pH 4–12 |
Display | LCD with LED backlight |
Display resolution | 128 × 64 pixels |
Display dimensions | 2.75″ × 1.5″ |
Display colour | Black on grey background; blue on white background with LED backlight on |
Graphical display options | Bar, gauge or line |
Outputs and Communication | |
|---|---|
Total chlorine output | One 4–20 mA output, set to the sensor measurement range |
pH output | One optional 4–20 mA output, set to pH 0–14 |
Digital communication | Modbus RTU via RS485 (standard) |
Alarm relays | Three SPDT, Form C relays |
Relay configuration | High/low alarms with expiration timer, periodic timers or fault alarms |
Input Power | |
|---|---|
DC power supply | 24 VDC (18–36 VDC) |
DC power requirement | 250 mW minimum |
AC power supply | 100–240 VAC, 50/60 Hz |
AC power consumption | 4 W |
Fuse protection | 250 V, 1 A, slow-blow fuse |
Enclosure | |
|---|---|
Enclosure material | Polycarbonate |
Enclosure colour | Beige |
Protection rating | IP65, weatherproof |
Enclosure format | ½ DIN |
Dimensions (L × W × D) | 14.4 × 14.4 × 9.0 cm (5.7″ × 5.7″ × 3.5″) |
Environmental Conditions | |
|---|---|
Installation environment | Outdoor use with protective cover or sunshield |
Ambient temperature (24 VDC models) | −20 to +70 °C |
Ambient temperature (100–240 VAC models) | −20 to +60 °C |
Storage temperature | −30 to +85 °C |
Relative humidity | 0–80% up to 31 °C, decreasing linearly to 50% RH at 40 °C |
Maximum altitude | 2,000 m (6,500 ft) |
Mains supply voltage fluctuation | Up to ±10% of nominal voltage |
Transient overvoltage category | CAT II |
Pollution degree | 2 |
Sample Requirements | |
|---|---|
Sample pressure | 1–30 psig (0.1–2 bar) |
System Configuration and Features | |
|---|---|
System configuration | Panel-mounted system |
Integrated components | Total chlorine sensor, pH sensor, constant head flow controller, flow cells and T80 transmitter |
Measurement technology | Reagentless |
Flow control | Automatic |
Calibration | Factory calibrated |
Compliance | EPA Method 334.0 |
Automatic cleaning | Optional spray cleaner for the chlorine flow cell |
Downloads
| FC80 Instruction Manual | DownloadFC80 Instruction Manual |
|---|
Related Products

ECD
DCA-23 Seawater Dechlorination Analyzer
The DCA-23 monitors seawater chlorination and dechlorination from 0.00–2.00 ppm chlorine equivalent, featuring reagent-free measurement, flow-independent…
Learn More
ECD
DC80 Dechlorination Analyzer
The DC80 Dechlorination Analyzer uses a zero-shift measurement strategy, automatic flow control and pH compensation to monitor very low total chlorine levels…
Learn More
ECD
CD80 Chlorine Dioxide Analyzer
The CD80 is a reagentless, panel-mounted analyzer for continuous chlorine dioxide (ClO₂) monitoring from 0.05 to 20 ppm, featuring automatic flow control, a…
Learn More
ECD
TC80 Total Chlorine Analyzer
The TC80 is a reagentless, panel-mounted analyzer for total chlorine monitoring in water, featuring amperometric sensing, automatic pH compensation, automatic…
Learn More