WETEC

Senseca

HP4SC Hydrostatic Level Transmitter


Slim hydrostatic level transmitter for water and industrial liquids.

  • 0…100 mbar to 0…20 bar / 0…1 to 0…100 mH₂O, up to ±0.05% FSI

  • 4…20 mA FSK or RS485 Modbus-RTU output

  • Optional temperature and 0…50 mS/cm conductivity measurement

Quantity
1

Overview

The Hydrocont HP4SC is a hydrostatic level transmitter for continuous monitoring and control of liquid levels in fresh water, wastewater and saltwater. A capacitive 99.9% Al₂O₃ ceramic cell measures the hydrostatic pressure directly through the process diaphragm. Its slim Ø22 mm stainless steel design fits confined spaces such as small-diameter bore holes and sounding tubes, and the support cable with steel core keeps it reliable under low temperatures, shock and vibration loads.

It covers ranges from 0…100 mbar to 0…20 bar, or 0…1 to 0…100 mH₂O, with accuracy options down to ≤ ±0.05% FSI, and is rated IP68 up to 100 m / 20 bar. Output is 4…20 mA with FSK communication or RS485 Modbus-RTU, both with integrated overvoltage protection. Optional Pt100 / Pt1000 temperature measurement and a 4-electrode conductivity cell up to 0…50 mS/cm add more water parameters, and ATEX/IECEx versions are available.

Typical applications

  • Water and waste water sector

  • Fresh water, wastewater and saltwater level measurement

  • Bore holes, wells and small-diameter sounding tubes

  • Long-term water level monitoring at hydroelectric power plants

  • Weir control

  • Stockpile management of storage tanks

  • Environmental technology and process industry

Key Features

  • Capacitive 99.9% Al₂O₃ ceramic sensing cell, dry system

  • Slim Ø22 mm stainless steel design

  • Ranges 0…100 mbar to 0…20 bar, or 0…1 to 0…100 mH₂O

  • Accuracy options ≤ ±0.05%, ±0.1% or ±0.2% FSI

  • 4…20 mA with FSK or RS485 Modbus-RTU output

  • Integrated overvoltage protection

  • Optional Pt100 / Pt1000 temperature measurement

  • Optional 4-electrode conductivity cell, 0…50 mS/cm

  • IP68 up to 100 m / 20 bar

  • Support cable with steel core, up to 300 m

  • Duplex steel version for sea water; ATEX/IECEx versions

Benefits

  • Fits confined spaces such as small bore holes and sounding tubes

  • Reliable under low temperatures, shock and vibration loads

  • High accuracy and long-term stability

  • Level, temperature and conductivity from one device

  • Easy integration into plant systems

Specifications

PRESSURE / LEVEL INPUT

 

Sensor type

Capacitive ceramic cell (dry system, 99.9% Al₂O₃)

Measuring ranges (relative)

0…100 mbar up to 0…20 bar, or 0…1 mH₂O up to 0…100 mH₂O

Overload / burst pressure

5…40 bar, depending on measuring range

Characteristic curve deviation

≤ ±0.05% FSI / ±0.1% FSI / ±0.2% FSI

Temperature effect

Tk zero ≤ ±0.015% FSI/K, ≤ ±0.75% FSI; Tk span ≤ ±0.015% FSI/K, ≤ ±0.5% FSI (≥ 0.4 bar / 4 mH₂O), ≤ ±0.8% FSI (< 0.4 bar / 4 mH₂O)

Effect of auxiliary power

≤ ±0.002% FSI/V

Long-term zero drift

≤ ±0.15% FSI/year

Resolution

≥ 16 bits

 

 

TEMPERATURE INPUT (OPTIONAL)

 

Pt100, 3-wire (4…20 mA version)

Class B, IEC 60751; -20…+70 °C; t90 ≤ 60 s; error ≤ ±0.3 K + 0.005 x |Tp|

Pt1000 (RS485 version)

IEC 60751; -20…+70 °C; resolution ≤ ±0.01 K; deviation ≤ ±0.2 K + 0.005 x |Tp|; long-term drift ≤ ±0.1 K/year

 

 

CONDUCTIVITY INPUT (OPTIONAL)

 

Sensor type

Conductive 4-electrode cell

Measuring range

0…50 mS/cm

Resolution

≤ 1 µS/cm

Calibration deviation

≤ ±1% of measured value (≥ ±1 µS/cm)

Temperature compensation

-2 %/K, -5…+45 °C; reference temperature +25 °C

 

 

OUTPUTS

 

4…20 mA, FSK, 2-wire

Signal range 4…20 mA, limit/error 3.9…20.5 mA; FSK 1200 bit/s; resolution ≤ 1 µA; load RL ≤ (Us - 9 V) / 20.5 mA

RS485 Modbus-RTU, 4-wire

9600 baud (4800…38400 baud); t90 ≤ 2 ms pressure/level, ≤ 60 s temperature, ≤ 2 s conductivity

Surge protection

Integrated (EN/IEC 61000-4-5), rated discharge current 10 kA (8/20 µs)

 

 

POWER REQUIREMENT

 

4…20 mA version

9…35 Vdc (Ex: 9…30 Vdc), reverse-polarity protected

RS485 version

6…35 Vdc, reverse-polarity protected; ≤ 10 mA (≤ 15 mA with temperature/conductivity)

 

 

ENVIRONMENTAL

 

Process / ambient temperature

-20…+70 °C (ATEX/IECEx: see certificate)

Protection degree

IP68 [≤ 100 m / ≤ 20 bar] (EN/IEC 60529)

Pressure cycles

≥ 100 million (1.2 x PN)

Impact / vibration resistance

50 g [11 ms] (EN/IEC 60068-2-27) / 20 g [10…2000 Hz] (EN/IEC 60068-2-6)

Climate class

4K4H (EN/IEC 60721-3-4)

EMC

Class B equipment / industrial environment (EN/IEC 61326)

 

 

MATERIALS

 

Process wetted

Ceramic Al₂O₃ 99.9%; steel 1.4404/316L, 1.4571/316Ti or 1.4462/318LN (duplex, sea water resistant); FKM/FPM, EPDM, PE, PUR, epoxy

Support cable

Steel shield breaking strength ≥ 920 N; bending radius ≥ 100 mm; PE or PUR sheath; length up to 300 m

 

 

MECHANICAL

 

Diameter

Ø 22 mm

Weight

0.3 kg + 0.068 kg/m cable (0.375 kg + 0.068 kg/m with temperature/conductivity)

 

 

APPROVALS

 

Variants

Standard; ATEX/IECEx II 1 G Ex ia IIC T6…T1 Ga / II 1 D Ex ia IIIC Da; ATEX/IECEx II 2 G Ex ib IIC T6…T1 Gb / II 2 D Ex ib IIIC T80°C Db

Drawing

HP4SC-Dimensional-Drawing

Downloads

Hydrocont-HP4-Datasheet-EN DownloadHydrocont-HP4-Datasheet-EN
Hydrocont-HP4-User-Manual-EN DownloadHydrocont-HP4-User-Manual-EN