Hydrostatic level sensor with dry capacitive ceramic measuring cell.

  • 4…20 mA or 0…10 V output, optional 2x PNP switching outputs

  • Optional LED display and keypad

  • Ranges from 0…50 mbar to 0…20 bar, ATEX versions available

Quantity
1

Overview

The Hydrocont S50 is a compact hydrostatic level sensor for measuring and monitoring fill levels in tanks, basins and channels. It is installed in the wall of the container, where the liquid acts directly on a dry capacitive ceramic membrane without a pressure mediator liquid. High strength against pressure blows, resistance to chemicals and corrosion, and low sensitivity to temperature shocks and EM interference make it suitable for liquids such as water, waste water, solvents, oil, sludge, grease and cleaning agents.

The integrated digital electronics convert the membrane signal into a high resolution 4…20 mA or 0…10 V output. Depending on the version, a bright LED display, 3-key keypad and 2 PNP switching outputs are built in, allowing the measuring range, display, switching outputs and damping to be set on site. It is available in many housing, process connection and construction forms, including tube extension and cable versions for mounting from the top, and ATEX versions for hazardous areas.

Typical applications

  • Fill level measurement in basins, channels and tanks

  • Pressureless containers with water and waste water

  • Solvents, oil, sludge, grease and cleaning agents

  • Hygienic applications with hygienic process connections

  • Hazardous areas (ATEX versions)

Key Features

  • Dry capacitive ceramic measuring cell, 99.9% purity

  • Medium contacts the ceramic membrane directly, no mediator liquid

  • 4…20 mA (2-wire) or 0…10 V (3-wire) output

  • Optional 2x PNP switching outputs

  • Optional high-brightness LED display and 3-key keypad

  • Measuring ranges from 0…50 mbar to 0…20 bar

  • 80-fold overload resistance

  • Accuracy down to ≤ 0.05% with Xcellence membrane

  • Stainless steel or plastic housings, M12 plug, cable or terminal box

  • Tube extension and cable versions for mounting from the top

  • ATEX versions for hazardous areas

Benefits

  • Measurement independent of foam, dielectric value and tank internals

  • High resistance to pressure blows, chemicals and corrosion

  • Low sensitivity to temperature shocks and EM interference

  • High accuracy and long-term stability

  • Easy commissioning with good display readability

  • Process connections to meet all requirements, including hygienic

Specifications

MEASUREMENT

 

Measuring principle

Hydrostatic, dry capacitive ceramic measuring cell (no pressure mediator liquid)

Measuring ranges

0…50 mbar, 0…100 mbar, 0…200 mbar, 0…400 mbar, 0…1 bar, 0…2 bar, 0…4 bar, 0…10 bar, 0…20 bar, -1…+1 bar; special ranges on request

Accuracy (deviation in characteristics)

≤ 0.2% or ≤ 0.1% of nominal range (depending on order code); ≤ 0.05% with Xcellence membrane

Temperature deviation

≤ 0.1% / 10 K of the nominal range

Calibration deviation

≤ 0.05% of the nominal range

Long term drift

≤ 0.1% / year of the nominal range

Influence of supply voltage

≤ 0.02% / 10 V of the nominal range

Resolution

Better than 1 µA or 0.5 mV (16 bit = 65,536 steps)

 

 

OUTPUTS

 

Variants A / B / C / D

4…20 mA, 2-wire

Variants E / F / G / H

0…10 V, 3-wire

Switching outputs (S1 / S2)

2x PNP switching on +VS (variants A and E); output current > 250 mA, current limited, short circuit protected

Display and keypad

LED display and keypad (A, B, E, F); keypad only (C, G); none (D, H)

 

 

POWER REQUIREMENT

 

Supply voltage

Variants C / D: 10.5…45 Vdc; variants A / B / E / F / G / H: 14.5…45 Vdc

Residual ripple

≤ 2 Vss

 

 

MATERIALS

 

Membrane

Ceramic Al₂O₃ 96% or 99.9%

Process connection

Steel 1.4404 (316L); PEEK; others on request

Gaskets

Viton, EPDM, Neoprene, perfluoroelastomer (FFKM)

Connection housing

Steel 1.4301, POM (Delrin), PBT

Extension cable

PE or FEP

 

 

ENVIRONMENTAL

 

Medium temperature

-40…+125 °C (140 °C for 1 h); with extension cable -20…+70 °C

Ambient / storage temperature

-40…+85 °C; with extension cable -20…+70 °C

Protection degree

IP65 / IP67

 

 

VERSIONS

 

Construction forms

Standard, short form flush-mounted, extension cable, tube extension, flush-mounted, high temperature (-10…+200 °C, metallic membrane), extension cable probe / field enclosure

Process connections

G¾", G½", G1½" (ISO 228-1), G1½" PEEK, milk tube DN40/DN50 (DIN 11851), Varivent N, DRD 65 mm, Tri-Clamp 2", flanges DN25…DN80 (DIN EN 1092-1), nut groove adapter Ø44 mm

Electrical connection

M12 plug, cable (2 m) or terminal box

ATEX

ExS50: ATEX II 1/2 G Ex ia IIC T4 Ga/Gb; XDS50: ATEX II 1/2 D Ex ia IIIC T60°C/T102°C Da/Db + ATEX II 1/2 G Ex ia IIC T4 Ga/Gb