ECD
S80 Inductive Conductivity Sensors
Intelligent inductive conductivity sensors with non-contact toroidal measurement technology and chemical-resistant PVDF construction.
Conductivity measurement up to 1000 mS
Two-way digital communication with stored sensor identity and calibration information
Insertion/submersion and valve-retractable designs
Overview
The ECD S80 Inductive Conductivity Sensors are intelligent digital sensors that perform non-contact conductivity measurements using two inductively coupled toroidal coils. Featuring a chemically resistant PVDF (KYNAR) body and a measurement capability of up to 1000 mS, they are suitable for high-conductivity solutions and applications where electrode coating or corrosion is a concern.
The sensors convert analogue signals into a digital protocol that enables two-way communication with compatible ECD analysers. The measurement type, serial number and three calibration registers are stored in the internal sensor memory, enabling automatic configuration of the analyser’s displays and outputs upon connection.
Two installation designs are available: insertion/submersion and valve-retractable. The former uses a ¾-inch MNPT compression fitting for adjustable insertion depth, while the latter incorporates a 1-inch ball valve with a 1-inch NPT process connection, allowing sensor removal and servicing without interrupting the sample flow.
Typical Applications
Conductivity measurement in high-conductivity solutions
Percentage concentration measurement of solutions
Conductivity monitoring in corrosive environments
Measurement in applications prone to electrode coating
In-line conductivity monitoring in pipes and flow cells
Submersible conductivity monitoring in tanks
Key Features
Non-contact conductivity measurement using inductively coupled toroidal coils
Conductivity measurement up to 1000 mS
Chemically resistant PVDF (KYNAR) sensor body
Two-way digital communication with compatible ECD analysers
Internal memory storing measurement type, serial number and three calibration registers
Automatic configuration of analyser displays and outputs upon connection
Insertion/submersion design with a ¾-inch MNPT compression fitting for adjustable insertion depth
Valve-retractable design with a 1-inch ball valve, 1-inch NPT process connection and flared sensor end for blowout prevention
Benefits
Wide conductivity measurement capability allows one sensor to serve multiple applications
Non-contact toroidal measurement with chemically resistant PVDF construction minimises problems associated with electrode coating and corrosion
Stored sensor identity and calibration information enables automatic analyser configuration, reducing manual setup requirements
Flexible installation options accommodate pipes, flow cells and tanks
Valve-retractable design allows sensor removal and servicing without interrupting sample flow
Specifications
Measurement | |
|---|---|
Measurement type | Inductive conductivity |
Measuring principle | Non-contact measurement using inductively coupled toroidal coils |
Maximum conductivity measurement | 1000 mS |
Temperature Compensation | |
|---|---|
Temperature compensation | Automatic, 0…100 °C |
Temperature accuracy | ±0.2 °C |
Temperature sensing element | 100 kΩ thermistor |
Operating Conditions | |
|---|---|
Pressure range | 0…100 psig |
Digital Communication and Calibration | |
|---|---|
Communication | Two-way digital communication |
Compatible instruments | ECD T80 Transmitter and LQ800 Controller |
Internal memory | Sensor identity, measurement type, serial number and calibration registers |
Calibration | Factory calibrated; periodic calibration required |
Mechanical | |
|---|---|
Sensor body material | KYNAR (PVDF) |
Standard insertion/submersion dimensions | ¾-inch OD × 10-inch length |
Standard valve-retractable dimensions | ¾-inch OD × 17-inch length |
Alternative housing dimensions | 1-inch OD × 10-inch length |
Installation and Process Connections | |
|---|---|
Installation configurations | Insertion/submersion and valve-retractable |
Insertion/submersion process connection | ¾-inch MNPT compression fitting |
Insertion depth | Adjustable |
Valve-retractable process connection | 1-inch ball valve with 1-inch NPT process connection |
Blowout protection | Flared sensor end for valve-retractable design |
Cable and Connector | |
|---|---|
Cable configuration | Standard integral cable or optional detachable cable connection |
Standard cable length | 10 ft |
Optional cable lengths | 20, 30, 40, 50 and 100 ft |
Detachable connector type | Straight (axial) |
Detachable connector protection rating | IP68 |
Valve Retraction Assembly Options Model | |
|---|---|
VSSE | 1-inch NPT, 316 stainless steel valve retraction assembly for inductive conductivity sensors |
VKYE | 1-inch KYNAR valve assembly for ¾-inch inductive conductivity sensors |
Downloads
| S80 - Sensors Manual | DownloadS80 - Sensors Manual |
|---|
Related Products

ECD
T80 Universal Transmitter
The ECD T80 Universal Transmitter provides continuous industrial process monitoring with single- or dual-channel operation, automatic S80 sensor recognition…
Learn More
ECD
S88 Hazardous Location Sensors
The S88 pH Sensors offer digital pH measurement from 0–14 pH, built-in calibration data storage, replaceable electrode cartridges, and insertion, submersion…
Learn More
ECD
S80 Ion Selective Sensors
The S80 Ion Selective Sensors measure specific ions in solution, featuring digital communication, onboard calibration storage, replaceable electrodes, and…
Learn More
ECD
S80 Resistivity Sensors
S80 Resistivity Sensors measure from 0.00 to 50.00 MΩ, featuring digital communication, onboard identity and calibration data storage, and…
Learn More