ECD
S88 Hazardous Location Sensors
Intelligent digital pH sensors with replaceable electrode cartridges for industrial process monitoring.
Measurement range: 0.00–14.00 pH
Temperature range: −10…+90 °C (0…+130 °C for high-temperature sensors)
Two-way digital communication with onboard sensor identity and calibration data storage
Insertion, submersion and valve-retractable configurations
Standard 316 stainless steel housing with alternative materials available
Overview
The ECD S88 pH Sensors are intelligent digital sensors designed for industrial pH measurement, covering a measurement range of 0.00–14.00 pH. They feature replaceable, application-specific electrode cartridges and operate at temperatures from −10 to +90 °C, with high-temperature versions supporting 0 to +130 °C. The standard housing is constructed from 316 stainless steel, with alternative materials available to suit different process conditions.
The sensors convert analogue electrode signals into a digital protocol, enabling two-way communication with compatible ECD transmitters. Sensor identity, measurement type, serial number and calibration information are stored in the onboard memory, allowing automatic transmitter configuration upon connection. The sensors are available in insertion/submersion and valve-retractable designs, with adjustable insertion lengths to accommodate different installation requirements.
Typical Applications
In-line pH monitoring in pipelines and flow cells
Submersible pH measurement in process tanks
High-temperature process pH monitoring using suitable electrode cartridges
Key Features
pH measurement range of 0.00–14.00 pH
Temperature range of −10…+90 °C (0…+130 °C for high-temperature sensors)
Two-way digital communication with compatible ECD transmitters
Onboard storage of sensor identity, measurement type, serial number and three calibration registers
Replaceable, application-specific electrode cartridges
Insertion/submersion and valve-retractable sensor designs
Adjustable insertion length with ¾″ MNPT compression fitting for insertion sensors
1″ ball valve with 1″ NPT process connection for valve-retractable sensors
Flared end on valve-retractable sensors to prevent blowout
Standard 316 stainless steel housing with alternative materials available
Wetted materials: 316L stainless steel, PEEK or PES (electrode body), Teflon and Viton
Benefits
Simplified sensor setup through onboard data storage and automatic transmitter configuration
Application-specific electrode cartridges suit different measurement requirements
Reduced cost of ownership through replaceable electrode cartridges
Flexible installation in pipelines, flow cells and process tanks
Alternative construction materials accommodate different process conditions
Flared-end design helps prevent blowout in valve-retractable configurations
Specifications
Measurement | |
|---|---|
Measured parameter | pH |
Measurement range | 0.00–14.00 pH |
Temperature range | −10…+90 °C |
Temperature range (high-temperature sensors) | 0…+130 °C |
Digital Communication and Data Storage | |
|---|---|
Signal conversion | Analogue electrode signals to digital protocol |
Communication | Two-way digital communication |
Stored sensor information | Sensor identity, measurement type and serial number |
Calibration registers | 3 |
Analyzer compatibility | ECD digital analyzers |
Automatic configuration | Transmitter displays and outputs configured according to the measured parameter |
Sensor Design and Construction | |
|---|---|
Sensor designs | Insertion/submersion or valve-retractable |
Electrode cartridges | Replaceable, application-specific |
Standard housing material | 316 stainless steel |
Alternative construction materials | Available upon request |
Wetted materials | 316L stainless steel, PEEK or PES (electrode body), Teflon and Viton |
Insertion/Submersion Sensor | |
|---|---|
Process connection | ¾″ MNPT compression fitting |
Insertion length | Adjustable |
Insertion installation | Pipe tees, flow cells or through tank walls |
Submersion installation | Standpipe using a reversed compression fitting |
Valve-Retractable Sensor | |
|---|---|
Ball valve size | 1″ |
Process connection | 1″ NPT |
Insertion/retraction mechanism | Sliding sensor through the ball valve by loosening the rear compression fitting |
Blowout prevention | Flared sensor end |
Downloads
| S80 - Sensor Manual | DownloadS80 - Sensor Manual |
|---|
Related Products

ECD
T80 Universal Transmitter
The ECD T80 Universal Transmitter provides continuous industrial process monitoring with single- or dual-channel operation, automatic S80 sensor recognition…
Learn More
ECD
S80 Ion Selective Sensors
The S80 Ion Selective Sensors measure specific ions in solution, featuring digital communication, onboard calibration storage, replaceable electrodes, and…
Learn More
ECD
S80 Inductive Conductivity Sensors
The S80 Inductive Conductivity Sensors feature non-contact conductivity measurement, chemical-resistant PVDF construction, intelligent digital communication…
Learn More
ECD
S80 Resistivity Sensors
S80 Resistivity Sensors measure from 0.00 to 50.00 MΩ, featuring digital communication, onboard identity and calibration data storage, and…
Learn More