ECD
T80 Universal Transmitter
Universal single- or dual-channel transmitter for continuous industrial process monitoring with ECD S80 Intelligent Sensors.
Supports pH, ORP, pION, dissolved oxygen, turbidity, conductivity, resistivity and other supported measurement parameters
Automatic sensor recognition and configuration
LCD display with numerical and graphical readings
4–20 mA output, with Modbus RTU depending on configuration
Optional HART® communication and alarm relays
Weatherproof NEMA 4X polycarbonate enclosure
Overview
The ECD T80 Universal Transmitter is a single- or dual-channel instrument designed for continuous measurement of pH, ORP, pION, dissolved oxygen, turbidity, conductivity, resistivity and other supported parameters in industrial environments. It communicates digitally with ECD S80 Intelligent Sensors, automatically configuring its menus and display according to the connected sensor. The sensors are factory calibrated and equipped with onboard memory for sensor identification and calibration data.
The transmitter features a graphical LCD with numerical readings, trend lines, bar graphs and gauge displays. It provides 4–20 mA output, with Modbus RTU communication available on line-powered versions. HART® communication and alarm relays are optional. Housed in a weatherproof NEMA 4X polycarbonate enclosure, the T80 is available in loop-powered (single-channel only), 24 VDC and 100–240 VAC versions, with multiple mounting configurations.
Typical Applications
Continuous industrial process monitoring
Monitoring of pH, ORP, specific ions and dissolved oxygen
Measurement of turbidity, conductivity and resistivity
Signal integration with industrial control and monitoring systems
Process alarm and fault monitoring using optional alarm relays
Key Features
Single- or dual-channel transmitter configuration
Supports pH, ORP, pION, dissolved oxygen, turbidity, conductivity, resistivity and other supported measurement parameters
Automatic sensor recognition and configuration with ECD S80 Intelligent Sensors
Compatibility with factory-calibrated S80 sensors featuring onboard data storage
128 × 64 pixel graphical LCD with numerical readings, trend lines, bar graphs and gauge displays
Intuitive menu structure with membrane switch buttons
4–20 mA output, with Modbus RTU communication on line-powered versions
Optional HART® communication
Optional three SPDT alarm relays on line-powered versions
Loop-powered (single-channel only), 24 VDC or 100–240 VAC power supply options
Weatherproof NEMA 4X polycarbonate enclosure
Ambient operating temperature range of −20…+70 °C
Multiple mounting configurations
Benefits
Supports multiple measurement parameters, reducing the need to maintain separate transmitter types
Automatic sensor recognition simplifies instrument configuration
Factory-calibrated S80 sensors enable immediate use upon connection
Numerical and graphical displays facilitate measurement interpretation
Intuitive menu navigation simplifies operation and parameter adjustment
Flexible output and communication options facilitate integration with industrial control systems
Weatherproof NEMA 4X enclosure provides environmental protection
Multiple mounting options accommodate different installation requirements
Specifications
General | |
|---|---|
Model | T80 |
Instrument type | Universal transmitter |
Channel configuration | Single or dual channel |
Measurement operation | Continuous measurement |
Sensor compatibility | ECD S80 Intelligent Sensors |
Input protocol | Digital |
Sensor configuration | Automatic sensor recognition and configuration |
Analogue input | Optional analogue-to-digital input board for mV sensors on line-powered versions |
Measurement Ranges | |
|---|---|
pH | −1.00…15.00 pH |
ORP | −1500…+1500 mV |
pION | 0.1…999.9, auto-ranging (ppb, ppm, parts per thousand) |
Dissolved oxygen | 0.1…999.9, auto-ranging (ppb, ppm, % SAT, mg/L) |
Resistivity | 0.1…50 MΩ |
Turbidity | 0.0…9999 NTU, auto-ranging (NTU, FNU, mg/L, ppm, % solids) |
Total suspended solids (TSS) | 0…150,000 ppm |
Measurement Accuracy | |
|---|---|
ORP | ±1% |
pION | ±1–2% (specific to ion type) |
Dissolved oxygen | ±2% of calibrated range |
Conductivity | ±1% of calibrated range |
Resistivity | ±2% of calibrated range |
Turbidity | ±2% of calibrated range |
Temperature | ±0.3 °C |
Total suspended solids (TSS) | ±2% of calibrated range |
Display and User Interface | |
|---|---|
Display type | Graphical LCD |
Display resolution | 128 × 64 pixels |
Display dimensions | 2.75 × 1.5 inches |
Loop-powered display | Black lettering on grey background |
24 VDC and 100–240 VAC powered display | Blue lettering on white background with LED backlight |
Display modes | Single-channel and dual-channel display |
Display screens | Data screen, millivolt screen and graphical display screen |
Graphical display styles | Trend line, bar graph and gauge |
Menu navigation | Membrane switch buttons |
Calibration | |
|---|---|
Sensor calibration | Factory-calibrated S80 Intelligent Sensors |
Calibration functions | Auto calibration, standardisation, two-point calibration, single-point calibration and manual calibration |
Manual calibration | Entry of previously determined offset and slope values |
Calibration data storage | S80 sensor memory |
Configuration Functions | |
|---|---|
Hold function | Available |
Display adjustment | Backlight and contrast adjustment |
Instrument identification | Labels and tags |
Security | Password protection |
Reset function | Factory default reset |
Simulation function | Input and output signal simulation |
Power Supply | |
|---|---|
Loop-powered version (Code −0) | 24 VDC, single channel only |
Loop power supply range | 18…36 VDC |
Loop power requirement | 35 mW minimum |
Maximum loop load | 600 Ω |
DC-powered version (Code −1) | 24 VDC |
DC power supply range | 18…36 VDC |
DC power requirement | 250 mW minimum |
AC-powered version (Code −2) | 100…240 VAC, 50/60 Hz |
AC power consumption | 4 W |
Outputs and Communication | |
|---|---|
Analogue output | 4…20 mA |
Line-powered analogue output configuration | One output per channel |
Digital communication | Modbus RTU on line-powered versions |
HART® communication | Optional |
Fault condition output | 3.5 mA, 22 mA or none |
Alarm Relays (Optional) | |
|---|---|
Availability | Line-powered versions |
Number of relays | 3 |
Relay type | SPDT, Form 1C |
Maximum resistive current | 3 A |
Voltage rating | 250 VAC |
Relay functions | Alarm (setpoint), timer-activated or fault relays |
Setpoint alarm configuration | User-configurable high/low alarms |
Fault monitoring | T80 self-diagnostics with fault alarms |
Environmental Conditions | |
|---|---|
Ambient operating temperature | −20…+70 °C |
Storage temperature | −30…+85 °C |
Relative humidity | 0…90%, non-condensing |
Mechanical | |
|---|---|
Enclosure material | Polycarbonate |
Enclosure protection | NEMA 4X, weatherproof |
Enclosure format | ½ DIN |
Dimensions (L × W × D) | 14.4 × 14.4 × 9.0 cm (5.7 × 5.7 × 3.5 inches) |
Mounting Configurations | |
|---|---|
Universal mounting bracket | Standard |
Panel mounting | Panel mounting hardware |
Handrail mounting | Handrail mounting plate with 2-inch hardware |
Pole mounting | SSR polyester sun shield with 2-inch mounting hardware |
Rail mounting | SSR polyester sun shield with 2-inch mounting hardware |
Shipping | |
|---|---|
Shipping dimensions | 20.5 × 20.5 × 12.7 cm |
Shipping weight | 1.6 lbs (0.75 kg) |
Drawing

Downloads
| T80 - Universal Transmitter | DownloadT80 - Universal Transmitter |
|---|
Related Products

ECD
S88 Hazardous Location Sensors
The S88 pH Sensors offer digital pH measurement from 0–14 pH, built-in calibration data storage, replaceable electrode cartridges, and insertion, submersion…
Learn More
ECD
S80 Ion Selective Sensors
The S80 Ion Selective Sensors measure specific ions in solution, featuring digital communication, onboard calibration storage, replaceable electrodes, and…
Learn More
ECD
S80 Inductive Conductivity Sensors
The S80 Inductive Conductivity Sensors feature non-contact conductivity measurement, chemical-resistant PVDF construction, intelligent digital communication…
Learn More
ECD
S80 Resistivity Sensors
S80 Resistivity Sensors measure from 0.00 to 50.00 MΩ, featuring digital communication, onboard identity and calibration data storage, and…
Learn More