WETEC

ECD

S80 Ion Selective Sensors


Intelligent digital sensors for the measurement of specific ions in solution.

  • Measurement and operating temperature ranges dependent on electrode type

  • Replaceable, application-specific ion selective electrode cartridges

  • Two-way digital communication with onboard sensor identity and calibration data storage

  • Insertion/submersion and valve retractable designs

  • Standard 316 stainless steel housing with optional construction materials

Quantity
1

Overview

The ECD S80 Ion Selective Sensors are intelligent digital sensors designed to measure the activity of specific free ions in solution. They feature replaceable, application-specific electrode cartridges, with measurement, pH and operating temperature ranges depending on the selected electrode type.

The sensors convert analogue electrode signals into digital signals, enabling two-way communication with compatible ECD transmitters and controllers. Sensor identity, measurement type, serial number and calibration data are stored in the onboard memory, allowing automatic configuration of the connected instrument.

The S80 is available in insertion/submersion and valve retractable configurations. The insertion design uses a ¾″ MNPT compression fitting with adjustable insertion depth, while the valve retractable design uses a 1″ ball valve with a 1″ NPT process connection. The standard housing is constructed from 316 stainless steel, with alternative materials available to suit different process requirements.

Typical Applications

  • Measurement of specific free ions in aqueous solutions

  • Continuous online ion monitoring using suitable electrode cartridges

  • In-line ion measurement in process pipes and flow cells

  • Submersion ion monitoring in process tanks

Key Features

  • Measurement of specific free ion activity in solution

  • Replaceable, application-specific ion selective electrode cartridges

  • Measurement, pH and operating temperature ranges dependent on electrode type

  • Two-way digital communication with compatible ECD transmitters and controllers

  • Onboard storage of sensor identity, measurement type, serial number and calibration data

  • Factory-calibrated sensors

  • Insertion/submersion design with adjustable insertion depth and ¾″ MNPT compression fitting

  • Valve retractable design with 1″ ball valve, 1″ NPT process connection and flared end to prevent blowout

  • Standard 316 stainless steel housing with alternative construction materials available

  • Standard 10 ft cable with optional cable lengths and detachable cable connection

Benefits

  • Application-specific electrode selection to suit different ion measurement requirements

  • Reduced operating costs through replaceable electrode cartridges

  • Simplified installation with factory-calibrated sensors and onboard calibration data

  • Automatic transmitter configuration through digital sensor recognition

  • Flexible installation to accommodate different process arrangements

  • Sensor removal and maintenance without shutting down the process line when using the valve retractable design

  • Choice of housing materials to suit different process fluid compatibility requirements

Specifications

Measurement

Measurement parameter

Specific ion (pION)

Measurement principle

Ion-selective electrode

Measured quantity

Free ion activity in solution

Electrode design

Replaceable, application-specific electrode cartridges

 

 

Electrode type

Part number

Measurement range

pH range

Temperature range

Ammonium

2005083

0.05–18,000 ppm

2–10

0–40 °C

Cadmium

2005140

0.1–11,200 ppm

3–9

0–80 °C

Calcium

2005143

0.1–40,000 ppm

2.5–10

0–40 °C

Cyanide

2005142

0.1–260 ppm

11–13

0–80 °C

Cupric

2005058

1.0 ppb–6,300 ppm

2–6

0–80 °C

Fluoride

2005163

0.02–2,000 ppm

5–8

0–80 °C

Lead

2005141

2–20,700 ppm

4–8

0–80 °C

Potassium

2005034

0.1–40,000 ppm

2–12

0–40 °C

Sodium

2005031

0.2–23,000 ppm

11–14

0–80 °C

 

 

Digital Communication and Data Storage

Signal conversion

Analogue electrode signals to digital protocol

Communication

Two-way digital communication

Compatible instruments

ECD T80 Transmitter and LQ800 Controller

Stored sensor information

Sensor identity, measurement type and serial number

Calibration memory

Three calibration registers

Instrument configuration

Automatic configuration through sensor recognition

Optional diagnostics

SENTINEL remaining-life diagnostics

 

 

Mechanical

Sensor configurations

Insertion/submersion and valve retractable

Standard insertion sensor dimensions

¾″ OD × 10″ length

Standard valve retractable sensor dimensions

¾″ OD × 17″ length

Standard housing material

316 stainless steel

Optional housing materials

Titanium, Hastelloy C-22, polypropylene and PVDF, depending on configuration

Standard O-ring material

Viton®

Optional O-ring materials

Ethylene propylene, Viton® 75, Kalrez® and CV75

 

 

Process Connections

Insertion/submersion connection

¾″ MNPT compression fitting

Insertion length

Adjustable

Valve retractable connection

1″ NPT

Ball valve size

1″

Blowout protection

Flared end on valve retractable sensor

Available gland fitting materials

Stainless steel, Hastelloy, titanium, polypropylene, PVDF and Teflon®

Available valve assembly materials

Stainless steel, Hastelloy, titanium, PVDF and polypropylene

 

 

Cable and Connector

Standard cable length

10 ft

Optional cable lengths

20, 30, 40, 50 and 100 ft

Other cable lengths

Consult manufacturer

Standard cable connection

Integral cable

Optional cable connection

Detachable straight (axial) connector

Detachable connector protection

IP68

 

 

Calibration

Initial calibration

Factory calibrated

Calibration requirement

Periodic calibration

Standard specific ion calibration solutions

10 ppm and 100 ppm

Alternative calibration solution concentrations

Available upon request

Downloads

S80 - Sensor Manual DownloadS80 - Sensor Manual