ECD
X80 Hazardous Location Transmitter
Universal transmitter for continuous analytical measurement in hazardous industrial locations.
Single or dual-channel measurement of pH, ORP, specific ion, dissolved oxygen, turbidity, conductivity, resistivity and chlorine
Digital communication with ECD Intelligent Sensors and automatic parameter configuration
4–20 mA output with Modbus RTU or optional HART® communication, depending on configuration
Graphic LCD display and non-intrusive calibration
Corrosion-resistant 316 stainless steel enclosure
ATEX/IECEx or FM-approved configurations
Overview
The ECD X80 Series is a universal transmitter designed for continuous analytical measurement in hazardous industrial locations. Available in single or dual-channel configurations, it supports the measurement of pH, ORP, specific ion (pION), dissolved oxygen, turbidity, conductivity, resistivity and chlorine. The transmitter communicates digitally with compatible ECD Intelligent Sensors and automatically configures its menus and display screens according to the measured parameter.
The X80 features a graphic LCD display, magnetic switches for menu navigation and non-intrusive calibration without opening the enclosure. It is available in single-channel loop-powered or 24 Vdc-powered configurations, with 4–20 mA outputs, Modbus RTU, optional HART® communication and optional alarm relays, depending on the selected configuration. Its corrosion-resistant 316 stainless steel enclosure is available in ATEX/IECEx or FM-approved configurations.
Typical Applications
Continuous process monitoring in hazardous industrial locations
pH, ORP and specific ion measurement
Dissolved oxygen, turbidity and chlorine monitoring
Conductivity and resistivity measurement
Analogue and digital signal integration with industrial control systems
Key Features
Single or dual-channel universal transmitter
Supports pH, ORP, specific ion (pION), dissolved oxygen, turbidity, conductivity, resistivity and chlorine measurement
Digital communication with compatible ECD Intelligent Sensors and automatic parameter configuration
128 × 64 pixel graphic LCD with numerical and graphical display modes
Magnetic switches for menu navigation and non-intrusive calibration
4–20 mA output with Modbus RTU or optional HART® communication, depending on configuration
Three optional configurable SPDT alarm relays for 24 Vdc-powered versions
Loop-powered (single-channel only) or 24 Vdc-powered versions
Corrosion-resistant, electropolished 316 stainless steel enclosure
Pipe or wall mounting options
ATEX/IECEx or FM-approved configurations for hazardous locations
Benefits
Reduces the need for multiple instrument types through universal measurement capability
Simplifies setup through automatic sensor recognition and configuration
Reduces initial calibration requirements with factory-calibrated intelligent sensors
Allows calibration without opening the transmitter enclosure
Simplifies measurement interpretation through numerical and graphical displays
Facilitates integration with industrial control systems through flexible output options
Provides durability through corrosion-resistant 316 stainless steel construction
Accommodates different hazardous location requirements through available approval configurations
Offers installation flexibility with pipe or wall mounting options
Specifications
Input | |
|---|---|
Input protocol | ECD digital protocol |
Compatible sensors | ECD S80/S88 Intelligent Sensors, subject to approval configuration |
Sensor types | Liquid, gas and process sensors |
Number of channels | Single or dual channel |
Measurement Ranges | |
|---|---|
pH | −1.00 to 15.00 pH |
ORP | −1500 to +1500 mV |
Specific ion (pION) | 000.1–999.9, auto-ranging: ppb, ppm, parts per thousand |
Dissolved oxygen | 000.1–999.9, auto-ranging: ppb, ppm, % saturation, mg/L |
Conductivity | 0.055 µS to 2.00 S, auto-ranging: µS, mS, S |
Resistivity | 0.001–50.00 MΩ |
Turbidity | 000.0–4000 NTU, auto-ranging: NTU, FNU, mg/L, ppm, % solids (ATEX/IECEx datasheet) |
Chlorine (ATEX/IECEx datasheet) | Free or total chlorine, auto-ranging: ppb to ppm |
Chlorine (FM datasheet) | Free chlorine, auto-ranging: ppb to ppm |
Temperature | −30 to +140 °C |
Mechanical | |
|---|---|
Housing material | Electropolished 316 stainless steel |
Mounting provisions | 2 × M4 and 3 × ¾″ FNPT |
Environmental Conditions | |
|---|---|
Ambient temperature (ATEX/IECEx datasheet) | −20 to +70 °C |
Ambient temperature (FM datasheet) | −20 to +85 °C |
Storage temperature | −30 to +85 °C |
Relative humidity | 0–90%, non-condensing |
Display and User Interface | |
|---|---|
Display type | Graphic LCD |
Display resolution | 128 × 64 pixels |
Display dimensions | 2.75″ × 1.5″ |
Loop-powered display | Black lettering on grey background |
24 Vdc-powered display | Blue lettering on white background with LED backlight |
Main display screens | Data, millivolt and graphical display |
Graphical display styles | Trend line, bar graph and gauge |
Menu navigation | Magnetic switches |
Input Power | |
|---|---|
Loop-powered version | 24 Vdc, single channel only |
Loop-powered supply range | 18–36 Vdc |
Maximum loop load | 600 Ω |
Minimum loop-powered power requirement | 35 mW |
Line-powered version | 24 Vdc |
Line-powered supply range | 18–36 Vdc |
Minimum line-powered power requirement | 250 mW |
Outputs and Communication | |
|---|---|
Analogue output | 4–20 mA |
Analogue output configuration | One standard output or optional second output |
Loop-powered communication | Optional HART® |
Line-powered communication | Modbus RTU (standard) or HART® (optional) |
Alarm Relays | |
|---|---|
Availability | Optional for 24 Vdc-powered versions |
Number of relays | 3 |
Relay type | SPDT, Form 1C |
Relay rating | 250 Vac, 3 A maximum, resistive load |
Relay configuration | Setpoint alarms, timed relays or fault alarm |
Calibration | |
|---|---|
Sensor calibration | Factory-calibrated intelligent sensors |
Auto Cal | Two-point calibration |
Standardize | Single-point calibration |
Manual calibration | Entry of previously determined offset and slope values |
ATEX/IECEx Hazardous Location Approvals | |
|---|---|
ATEX certification number | ITS16ATEX101458X |
IECEx certification number | ETL 16.0049X |
X80 transmitter marking | II 2 G Ex db mb [ia IIC Ga] IIC T4 Gb |
Hazardous location classification | Zone 1, Ex db mb Gb |
Certified ambient temperature | −20 °C < Ta < +55 °C |
FM Hazardous Location Approvals | |
|---|---|
Approval | FM-approved X80 transmitter |
Zone classification | Class I, Zone 1, IIB+H₂, T4 |
Temperature range stated for Zone 1 approval | −40 to +85 °C |
Enclosure protection | Type 4X, IP66 |
Explosion-Proof Enclosure Approvals | |
|---|---|
CSA-certified enclosure classification | Class I, Zone 1, Ex d IIB+H₂ |
CSA enclosure protection | IP68 |
FM-approved enclosure classification | Class I, Division 1, Groups B, C and D; Class II, Division 1, Groups E, F and G; Class III, Division 1 |
FM enclosure protection | NEMA 4X, IP66 |
Shipping | |
|---|---|
Shipping dimensions | 8″ × 8″ × 5″ (20.5 × 20.5 × 12.7 cm) |
Weight | 8.0 lbs (3.65 kg) |
Downloads
| X80 – Haz Location Brochure ATEX & IECEx Approvals | DownloadX80 – Haz Location Brochure ATEX & IECEx Approvals |
|---|---|
| X80 – Haz Location Brochure FM Approvals | DownloadX80 – Haz Location Brochure FM Approvals |
Related Products

ECD
DCA-23 Seawater Dechlorination Analyzer
The DCA-23 monitors seawater chlorination and dechlorination from 0.00–2.00 ppm chlorine equivalent, featuring reagent-free measurement, flow-independent…
Learn More
ECD
DC80 Dechlorination Analyzer
The DC80 Dechlorination Analyzer uses a zero-shift measurement strategy, automatic flow control and pH compensation to monitor very low total chlorine levels…
Learn More
ECD
CD80 Chlorine Dioxide Analyzer
The CD80 is a reagentless, panel-mounted analyzer for continuous chlorine dioxide (ClO₂) monitoring from 0.05 to 20 ppm, featuring automatic flow control, a…
Learn More
ECD
TC80 Total Chlorine Analyzer
The TC80 is a reagentless, panel-mounted analyzer for total chlorine monitoring in water, featuring amperometric sensing, automatic pH compensation, automatic…
Learn More