ECD
CA6 Aluminum Analyzer
Online colorimetric analyzer for automatic measurement of aluminum in water.
Measuring ranges: 5–150 ppb (26 mm cell), 10–500 ppb (16 mm cell), and up to 20 ppm with internal dilution
Pyrocatechol violet method with absorbance measurement at 572 nm
Automatic calibration, validation and cleaning
Two 4–20 mA outputs and Modbus RTU RS485 communication
Color touchscreen interface and IP54 dual-compartment enclosure
Overview
The ECD CA-6 Aluminum Analyzer is a compact online colorimetric instrument designed for the automatic measurement of aluminum in water. It uses the pyrocatechol violet method with differential photometric absorbance measurement at 572 nm. The measuring ranges are 5–150 ppb Al³⁺ with a 26 mm cell and 10–500 ppb Al³⁺ with a 16 mm cell, extending up to 20 ppm Al³⁺ when equipped with an internal dilution module.
The analyzer features a heated colorimetric reaction cell, automatic calibration, validation and cleaning, and a programmable analysis frequency. Its IP54 dual-compartment enclosure separates the electronic and liquid components, while the color touchscreen interface provides access to measured values, operating status and configuration settings. It also includes an integrated data logger with USB download, two 4–20 mA outputs and Modbus RTU RS485 communication.
Typical Applications
Aluminum monitoring in drinking water
Aluminum measurement in industrial wastewater
Aluminum monitoring in municipal wastewater
Aluminum concentration measurement in surface water
Key Features
Automatic online sequential sampling and colorimetric analysis
Pyrocatechol violet method with absorbance measurement at 572 nm
Measuring ranges of 5–150 ppb Al³⁺ (26 mm cell) and 10–500 ppb Al³⁺ (16 mm cell), extending up to 20 ppm with internal dilution
LED light source and heated colorimetric reaction cell
Programmable analysis frequency with an 8–10 minute analysis cycle
Automatic calibration, validation and cleaning
Color touchscreen interface with multilingual support
Integrated data logger with USB download
Two 4–20 mA outputs and Modbus RTU RS485 communication
Four programmable SPDT potential-free alarm relays
Single or dual sample streams, with up to four streams using an external sequencer
IP54 dual-compartment enclosure separating electronic and liquid components
Rugged epoxy powder-coated, cold-rolled steel cabinet
Loss of sample and low reagent alarms
Benefits
Automated online analysis reduces the need for manual sampling and testing
Low reagent consumption and adjustable analysis frequency help reduce operating costs
Automatic calibration, validation and cleaning reduce downtime and operator intervention
Reference measurement compensates for sample color, turbidity and other optical interferences
Wide measuring range accommodates different aluminum concentrations
User-friendly interface simplifies operation and process configuration
Separate compartments isolate electronic components from the liquid-handling system
Factory-tested system simplifies installation and commissioning
Multiple sample stream configurations provide monitoring flexibility
Sample loss and low reagent alarms enable timely operator intervention
Specifications
Measurement | |
|---|---|
Measured parameter | Aluminum (Al³⁺) |
Measurement units | ppb, ppm, mg/L |
Measuring principle | Differential photometric absorbance |
Analytical method | Pyrocatechol violet method |
Light source | LED |
Measurement wavelength | 572 nm |
Measuring range (26 mm cell) | 5–150 ppb Al³⁺ |
Measuring range (16 mm cell) | 10–500 ppb Al³⁺ |
Extended measuring range | Up to 20 ppm Al³⁺ with internal dilution |
Reproducibility (26 mm cell) | ±5 ppb or ±5%, whichever is greater |
Reproducibility (16 mm cell) | ±10 ppb or ±5%, whichever is greater, up to 250 ppb; ±20 ppb or ±5%, whichever is greater, from 250–500 ppb |
Analysis System | |
|---|---|
Analysis frequency | Freely programmable, batch near-continuous analysis |
Cycle time | 8–10 minutes, including conditioning before analysis and rinsing after measurement |
Reaction cell | Heated colorimetric reaction cell |
Calibration, validation and cleaning | Automatic, with freely selectable intervals |
Reagent consumption (16 mm cell) | 1.7 L (0.45 US gal) of each reagent per 30 days at a 15-minute analysis frequency |
Reagent consumption (26 mm cell) | 2.5 L (0.66 US gal) of each reagent per 30 days at a 15-minute analysis frequency |
Sample Requirements | |
|---|---|
Sample pressure | Pressure-free from overflow vessel |
Sample temperature | 5–50 °C (41–122 °F) |
Sample flow rate | 80–500 mL/min |
Sample connection | 6 mm (¼ in.) |
Number of sample streams | 1 stream; 2 streams with integrated switching valve; 3 or 4 streams with external sequencer |
Drain Requirements | |
|---|---|
Drain type | Pressure-free, atmospheric drain |
Drain connection | 12 mm (½ in.) |
Display and Data Management | |
|---|---|
User interface | Color graphic touchscreen |
Display information | Measured values and status information |
Language support | Multiple languages |
Data logging | Integrated data logger with USB download |
Outputs, Communication and Alarms | |
|---|---|
Analogue outputs | 2 × 4–20 mA outputs for measured data |
Digital communication | Modbus RTU RS485 |
Alarm relays | 4 programmable SPDT potential-free relays |
Alarm functions | Loss of sample and low reagent alarms |
Digital input | Remote start/stop |
Electrical | |
|---|---|
Power supply | 100–240 VAC, 50/60 Hz (standard); 24 VDC (optional) |
Maximum power consumption | 80 VA |
Environmental Conditions | |
|---|---|
Operating temperature | 5–45 °C (41–113 °F) |
Relative humidity | 10–90%, non-condensing |
Operating environment | Indoor use; outdoor installation requires a protective cabinet or shelter (not included) |
Physical and Mechanical | |
|---|---|
Cabinet material | Cold-rolled steel |
Cabinet finish | Epoxy powder coating |
Enclosure design | Two separate compartments for electronic and liquid-handling components |
Access doors | Two lockable doors |
Dimensions (H × W × D) | 606 × 380 × 209 mm (23.6 × 15.0 × 8.2 in.) |
Weight | Approximately 20 kg (44 lbs) |
Ingress protection | IP54 |
Installation | Wall mounting (standard); benchtop support or panel mounting (optional) |
Drawing

Downloads
| CA6 - Aluminum Analyzer Brochure | DownloadCA6 - Aluminum Analyzer Brochure |
|---|
Related Products

ECD
DCA-23 Seawater Dechlorination Analyzer
The DCA-23 monitors seawater chlorination and dechlorination from 0.00–2.00 ppm chlorine equivalent, featuring reagent-free measurement, flow-independent…
Learn More
ECD
DC80 Dechlorination Analyzer
The DC80 Dechlorination Analyzer uses a zero-shift measurement strategy, automatic flow control and pH compensation to monitor very low total chlorine levels…
Learn More
ECD
CD80 Chlorine Dioxide Analyzer
The CD80 is a reagentless, panel-mounted analyzer for continuous chlorine dioxide (ClO₂) monitoring from 0.05 to 20 ppm, featuring automatic flow control, a…
Learn More
ECD
TC80 Total Chlorine Analyzer
The TC80 is a reagentless, panel-mounted analyzer for total chlorine monitoring in water, featuring amperometric sensing, automatic pH compensation, automatic…
Learn More