WETEC

ECD

CA6 Aluminum Analyzer

WaterWater Quality Measurement

Online colorimetric analyzer for automatic measurement of aluminum in water.

  • Measuring ranges: 5–150 ppb (26 mm cell), 10–500 ppb (16 mm cell), and up to 20 ppm with internal dilution

  • Pyrocatechol violet method with absorbance measurement at 572 nm

  • Automatic calibration, validation and cleaning

  • Two 4–20 mA outputs and Modbus RTU RS485 communication

  • Color touchscreen interface and IP54 dual-compartment enclosure

Quantity
1

Overview

The ECD CA-6 Aluminum Analyzer is a compact online colorimetric instrument designed for the automatic measurement of aluminum in water. It uses the pyrocatechol violet method with differential photometric absorbance measurement at 572 nm. The measuring ranges are 5–150 ppb Al³⁺ with a 26 mm cell and 10–500 ppb Al³⁺ with a 16 mm cell, extending up to 20 ppm Al³⁺ when equipped with an internal dilution module.

The analyzer features a heated colorimetric reaction cell, automatic calibration, validation and cleaning, and a programmable analysis frequency. Its IP54 dual-compartment enclosure separates the electronic and liquid components, while the color touchscreen interface provides access to measured values, operating status and configuration settings. It also includes an integrated data logger with USB download, two 4–20 mA outputs and Modbus RTU RS485 communication.

Typical Applications

  • Aluminum monitoring in drinking water

  • Aluminum measurement in industrial wastewater

  • Aluminum monitoring in municipal wastewater

  • Aluminum concentration measurement in surface water

Key Features

  • Automatic online sequential sampling and colorimetric analysis

  • Pyrocatechol violet method with absorbance measurement at 572 nm

  • Measuring ranges of 5–150 ppb Al³⁺ (26 mm cell) and 10–500 ppb Al³⁺ (16 mm cell), extending up to 20 ppm with internal dilution

  • LED light source and heated colorimetric reaction cell

  • Programmable analysis frequency with an 8–10 minute analysis cycle

  • Automatic calibration, validation and cleaning

  • Color touchscreen interface with multilingual support

  • Integrated data logger with USB download

  • Two 4–20 mA outputs and Modbus RTU RS485 communication

  • Four programmable SPDT potential-free alarm relays

  • Single or dual sample streams, with up to four streams using an external sequencer

  • IP54 dual-compartment enclosure separating electronic and liquid components

  • Rugged epoxy powder-coated, cold-rolled steel cabinet

  • Loss of sample and low reagent alarms

Benefits

  • Automated online analysis reduces the need for manual sampling and testing

  • Low reagent consumption and adjustable analysis frequency help reduce operating costs

  • Automatic calibration, validation and cleaning reduce downtime and operator intervention

  • Reference measurement compensates for sample color, turbidity and other optical interferences

  • Wide measuring range accommodates different aluminum concentrations

  • User-friendly interface simplifies operation and process configuration

  • Separate compartments isolate electronic components from the liquid-handling system

  • Factory-tested system simplifies installation and commissioning

  • Multiple sample stream configurations provide monitoring flexibility

  • Sample loss and low reagent alarms enable timely operator intervention

Specifications

Measurement

Measured parameter

Aluminum (Al³⁺)

Measurement units

ppb, ppm, mg/L

Measuring principle

Differential photometric absorbance

Analytical method

Pyrocatechol violet method

Light source

LED

Measurement wavelength

572 nm

Measuring range (26 mm cell)

5–150 ppb Al³⁺

Measuring range (16 mm cell)

10–500 ppb Al³⁺

Extended measuring range

Up to 20 ppm Al³⁺ with internal dilution

Reproducibility (26 mm cell)

±5 ppb or ±5%, whichever is greater

Reproducibility (16 mm cell)

±10 ppb or ±5%, whichever is greater, up to 250 ppb; ±20 ppb or ±5%, whichever is greater, from 250–500 ppb

 

 

Analysis System

Analysis frequency

Freely programmable, batch near-continuous analysis

Cycle time

8–10 minutes, including conditioning before analysis and rinsing after measurement

Reaction cell

Heated colorimetric reaction cell

Calibration, validation and cleaning

Automatic, with freely selectable intervals

Reagent consumption (16 mm cell)

1.7 L (0.45 US gal) of each reagent per 30 days at a 15-minute analysis frequency

Reagent consumption (26 mm cell)

2.5 L (0.66 US gal) of each reagent per 30 days at a 15-minute analysis frequency

 

 

Sample Requirements

Sample pressure

Pressure-free from overflow vessel

Sample temperature

5–50 °C (41–122 °F)

Sample flow rate

80–500 mL/min

Sample connection

6 mm (¼ in.)

Number of sample streams

1 stream; 2 streams with integrated switching valve; 3 or 4 streams with external sequencer

 

 

Drain Requirements

Drain type

Pressure-free, atmospheric drain

Drain connection

12 mm (½ in.)

 

 

Display and Data Management

User interface

Color graphic touchscreen

Display information

Measured values and status information

Language support

Multiple languages

Data logging

Integrated data logger with USB download

 

 

Outputs, Communication and Alarms

Analogue outputs

2 × 4–20 mA outputs for measured data

Digital communication

Modbus RTU RS485

Alarm relays

4 programmable SPDT potential-free relays

Alarm functions

Loss of sample and low reagent alarms

Digital input

Remote start/stop

 

 

Electrical

Power supply

100–240 VAC, 50/60 Hz (standard); 24 VDC (optional)

Maximum power consumption

80 VA

 

 

Environmental Conditions

Operating temperature

5–45 °C (41–113 °F)

Relative humidity

10–90%, non-condensing

Operating environment

Indoor use; outdoor installation requires a protective cabinet or shelter (not included)

 

 

Physical and Mechanical

Cabinet material

Cold-rolled steel

Cabinet finish

Epoxy powder coating

Enclosure design

Two separate compartments for electronic and liquid-handling components

Access doors

Two lockable doors

Dimensions (H × W × D)

606 × 380 × 209 mm (23.6 × 15.0 × 8.2 in.)

Weight

Approximately 20 kg (44 lbs)

Ingress protection

IP54

Installation

Wall mounting (standard); benchtop support or panel mounting (optional)

Drawing

CA6 Aluminum Analyzer Drawing

Downloads

CA6 - Aluminum Analyzer Brochure DownloadCA6 - Aluminum Analyzer Brochure