ECD
CA6 Ammonia Analyzer
Online sequential colorimetric analyzer for automated ammonia measurement in water.
Measuring ranges: 0…1.0, 0…10.0 or 0…50.0 mg/L (higher ranges require the optional dilution module)
LED light source and quartz reaction cell
4…20 mA analog output and two configurable alarm relays
Touchscreen interface and adjustable analysis cycle time
Rugged epoxy powder-coated cold-rolled steel cabinet
Overview
The ECD CA-6 Ammonia Analyzer is an online sequential sampling analyzer designed for automated measurement of ammonia concentrations in water using colorimetric technology. It features a heated reaction cell and uses an LED light source and photometer to measure absorbance through a quartz cell. Ammonia concentration is calculated from reference and reading measurements. Available measuring ranges are 0…1.0, 0…10.0 and 0…50.0 mg/L, with the latter two requiring an optional dilution module.
The analyzer features a touchscreen interface, programmable analysis sequences and adjustable cycle time to minimize reagent consumption. It provides a 4…20 mA analog output, two configurable alarm relays, and alarm functions for sample loss and low reagent levels. Its epoxy powder-coated cold-rolled steel cabinet contains separate electrical and liquid compartments with lockable doors. The analyzer supports wall mounting or installation with an optional bench support.
Typical Applications
Online monitoring of ammonia concentrations in water
Analog signal integration with external monitoring and control systems
Key Features
Online sequential sampling with differential absorbance colorimetric analysis
Ammonia measuring ranges of 0…1.0, 0…10.0 and 0…50.0 mg/L (the latter two require an optional dilution module)
LED light source, photometer and heated quartz reaction cell
Touchscreen interface with programmable analysis sequences
Adjustable analysis cycle time
4…20 mA analog output
Two configurable alarm relays, with loss-of-sample and low-reagent alarm functions
Separate electrical and liquid compartments with lockable doors
Epoxy powder-coated cold-rolled steel cabinet
Factory-calibrated analyzer
Wall mounting or optional bench support
Operating temperature of 5…50 °C
Power supply of 110…220 Vac, 50…60 Hz, 80 VA
Benefits
Automated sampling and analysis reduce manual measurement requirements
Reference measurement compensates for interference from sample color and turbidity
Factory calibration simplifies initial setup
User-friendly touchscreen interface simplifies operation and configuration
Adjustable analysis cycle time helps minimize reagent consumption
Low maintenance requirements reduce routine servicing needs
Separate electrical and liquid compartments keep electronic and fluid-handling components isolated
Sample loss and low reagent alarms help identify operational issues
Analog output enables integration with external monitoring and control systems
Specifications
Measurement | |
|---|---|
Measured parameter | Ammonia |
Measuring principle | Colorimetric differential absorbance |
Sampling method | Online sequential sampling |
Light source | Long-life LED |
Detector | Photometer |
Reaction cell | Heated quartz reaction cell |
Measuring range (Model CA6-02-A) | 0…1.0 mg/L |
Measuring range (Model CA6-02-B) | 0…10.0 mg/L |
Measuring range (Model CA6-02-C) | 0…50.0 mg/L |
Dilution module | Required for ranges B and C |
Dilution ratios | 10:1 or 50:1 with optional dilution module |
Response time | Dependent on the specific colorimetric measurement |
Repeatability | ±2% on absorbance value with turbidity < 80 NTU |
Drift | ±2% per month on the absorbance measurement |
Calibration | Factory calibrated |
Analysis and Reagent | |
|---|---|
Analysis configuration | Programmable according to installed accessories and number of reagent micropumps |
Number of reagents | Up to four (CA-6 series capability) |
Analysis cycle time | Adjustable |
Measurements per analysis cycle | Typically one |
Additional cycle | User-defined calibration or cleaning sequence |
Reagent consumption | Approximately 2,500 tests per litre of reagent, dependent on the specific colorimetric measurement |
Display and Control | |
|---|---|
User interface | Touchscreen |
Display information | Measured parameter, operating status and reagent volume percentage |
Menu options | RUN, DISPLAY, PROGRAM, SERVICE and HELP |
Outputs and Alarms | |
|---|---|
Analog output | 4…20 mA |
Alarm relays | Two configurable relays |
Alarm functions | Loss of sample and low reagent levels |
Sample Requirements and Connections | |
|---|---|
Inlet sample pressure | Atmospheric |
Outlet sample pressure | Atmospheric |
Sample flow rate for fast-loop reservoir | 100…500 mL/min |
Fast-loop reservoir connection | Flexible tubing, ¼″ OD |
Waste tubing | ⅜″ OD |
Electrical | |
|---|---|
Power supply | 110…220 Vac |
Frequency | 50…60 Hz |
Power rating | 80 VA |
Environmental | |
|---|---|
Operating temperature | 5…50 °C |
Mechanical | |
|---|---|
Cabinet construction | Epoxy powder-coated cold-rolled steel |
Enclosure configuration | Separate electrical and liquid compartments with lockable doors |
Dimensions (L × H × D) | 380 × 600 × 210 mm (15 × 24 × 8.25″) |
Weight | Approximately 17 kg (37.5 lbs) |
Mounting | Wall mounting or optional bench support |
Drawing

Downloads
| CA6 - Ammonia Analyzer Brochure | DownloadCA6 - Ammonia Analyzer Brochure |
|---|
Related Products

ECD
DCA-23 Seawater Dechlorination Analyzer
The DCA-23 monitors seawater chlorination and dechlorination from 0.00–2.00 ppm chlorine equivalent, featuring reagent-free measurement, flow-independent…
Learn More
ECD
DC80 Dechlorination Analyzer
The DC80 Dechlorination Analyzer uses a zero-shift measurement strategy, automatic flow control and pH compensation to monitor very low total chlorine levels…
Learn More
ECD
CD80 Chlorine Dioxide Analyzer
The CD80 is a reagentless, panel-mounted analyzer for continuous chlorine dioxide (ClO₂) monitoring from 0.05 to 20 ppm, featuring automatic flow control, a…
Learn More
ECD
TC80 Total Chlorine Analyzer
The TC80 is a reagentless, panel-mounted analyzer for total chlorine monitoring in water, featuring amperometric sensing, automatic pH compensation, automatic…
Learn More