WETEC

ECD

CA6 Ammonia Analyzer

WaterWater Quality Measurement

Online sequential colorimetric analyzer for automated ammonia measurement in water.

  • Measuring ranges: 0…1.0, 0…10.0 or 0…50.0 mg/L (higher ranges require the optional dilution module)

  • LED light source and quartz reaction cell

  • 4…20 mA analog output and two configurable alarm relays

  • Touchscreen interface and adjustable analysis cycle time

  • Rugged epoxy powder-coated cold-rolled steel cabinet

Quantity
1

Overview

The ECD CA-6 Ammonia Analyzer is an online sequential sampling analyzer designed for automated measurement of ammonia concentrations in water using colorimetric technology. It features a heated reaction cell and uses an LED light source and photometer to measure absorbance through a quartz cell. Ammonia concentration is calculated from reference and reading measurements. Available measuring ranges are 0…1.0, 0…10.0 and 0…50.0 mg/L, with the latter two requiring an optional dilution module.

The analyzer features a touchscreen interface, programmable analysis sequences and adjustable cycle time to minimize reagent consumption. It provides a 4…20 mA analog output, two configurable alarm relays, and alarm functions for sample loss and low reagent levels. Its epoxy powder-coated cold-rolled steel cabinet contains separate electrical and liquid compartments with lockable doors. The analyzer supports wall mounting or installation with an optional bench support.

Typical Applications

  • Online monitoring of ammonia concentrations in water

  • Analog signal integration with external monitoring and control systems

Key Features

  • Online sequential sampling with differential absorbance colorimetric analysis

  • Ammonia measuring ranges of 0…1.0, 0…10.0 and 0…50.0 mg/L (the latter two require an optional dilution module)

  • LED light source, photometer and heated quartz reaction cell

  • Touchscreen interface with programmable analysis sequences

  • Adjustable analysis cycle time

  • 4…20 mA analog output

  • Two configurable alarm relays, with loss-of-sample and low-reagent alarm functions

  • Separate electrical and liquid compartments with lockable doors

  • Epoxy powder-coated cold-rolled steel cabinet

  • Factory-calibrated analyzer

  • Wall mounting or optional bench support

  • Operating temperature of 5…50 °C

  • Power supply of 110…220 Vac, 50…60 Hz, 80 VA

Benefits

  • Automated sampling and analysis reduce manual measurement requirements

  • Reference measurement compensates for interference from sample color and turbidity

  • Factory calibration simplifies initial setup

  • User-friendly touchscreen interface simplifies operation and configuration

  • Adjustable analysis cycle time helps minimize reagent consumption

  • Low maintenance requirements reduce routine servicing needs

  • Separate electrical and liquid compartments keep electronic and fluid-handling components isolated

  • Sample loss and low reagent alarms help identify operational issues

  • Analog output enables integration with external monitoring and control systems

Specifications

Measurement

Measured parameter

Ammonia

Measuring principle

Colorimetric differential absorbance

Sampling method

Online sequential sampling

Light source

Long-life LED

Detector

Photometer

Reaction cell

Heated quartz reaction cell

Measuring range (Model CA6-02-A)

0…1.0 mg/L

Measuring range (Model CA6-02-B)

0…10.0 mg/L

Measuring range (Model CA6-02-C)

0…50.0 mg/L

Dilution module

Required for ranges B and C

Dilution ratios

10:1 or 50:1 with optional dilution module

Response time

Dependent on the specific colorimetric measurement

Repeatability

±2% on absorbance value with turbidity < 80 NTU

Drift

±2% per month on the absorbance measurement

Calibration

Factory calibrated

 

 

Analysis and Reagent

Analysis configuration

Programmable according to installed accessories and number of reagent micropumps

Number of reagents

Up to four (CA-6 series capability)

Analysis cycle time

Adjustable

Measurements per analysis cycle

Typically one

Additional cycle

User-defined calibration or cleaning sequence

Reagent consumption

Approximately 2,500 tests per litre of reagent, dependent on the specific colorimetric measurement

 

 

Display and Control

User interface

Touchscreen

Display information

Measured parameter, operating status and reagent volume percentage

Menu options

RUN, DISPLAY, PROGRAM, SERVICE and HELP

 

 

Outputs and Alarms

Analog output

4…20 mA

Alarm relays

Two configurable relays

Alarm functions

Loss of sample and low reagent levels

 

 

Sample Requirements and Connections

Inlet sample pressure

Atmospheric

Outlet sample pressure

Atmospheric

Sample flow rate for fast-loop reservoir

100…500 mL/min

Fast-loop reservoir connection

Flexible tubing, ¼″ OD

Waste tubing

⅜″ OD

 

 

Electrical

Power supply

110…220 Vac

Frequency

50…60 Hz

Power rating

80 VA

 

 

Environmental

Operating temperature

5…50 °C

 

 

Mechanical

Cabinet construction

Epoxy powder-coated cold-rolled steel

Enclosure configuration

Separate electrical and liquid compartments with lockable doors

Dimensions (L × H × D)

380 × 600 × 210 mm (15 × 24 × 8.25″)

Weight

Approximately 17 kg (37.5 lbs)

Mounting

Wall mounting or optional bench support

Drawing

CA6 Ammonia Analyzer Drawing

Downloads

CA6 - Ammonia Analyzer Brochure DownloadCA6 - Ammonia Analyzer Brochure