CA6 Chromium VI Analyzer
Online colorimetric analyzer for automatic measurement of hexavalent chromium (Cr(VI)) in water.
Measuring ranges: 0.5–300 ppb (26 mm cell) and 0.01–1 ppm (16 mm cell), up to 50 ppm with internal dilution
Programmable analysis frequency with 6–8 minute cycle time
Automatic calibration, validation and cleaning
Dual 4–20 mA outputs and Modbus RTU RS485 communication
Colour touchscreen interface, integrated data logger and IP54 enclosure
Overview
The ECD CA-6 Chromium VI Analyzer is an online colorimetric analyzer designed for automatic measurement of hexavalent chromium (Cr(VI)) in water. It uses differential photometric absorbance at 525 nm based on the reaction of chromium(VI) with 1,5-diphenylcarbazide in an acidic medium. The analyzer covers measuring ranges of 0.5–300 ppb with a 26 mm reaction cell and 0.01–1 ppm with a 16 mm reaction cell, extending up to 50 ppm with internal dilution.
The analyzer features a heated reaction cell, programmable analysis frequency to minimise reagent consumption, and automatic calibration, validation and cleaning. It is equipped with a colour touchscreen interface, an integrated data logger with USB download, two 4–20 mA outputs and Modbus RTU RS485 communication. The IP54-rated dual-compartment enclosure separates the electrical and liquid-handling components.
Typical Applications
Hexavalent chromium monitoring in drinking water
Industrial wastewater discharge monitoring
Industrial process optimisation
Surface water quality monitoring
Key Features
Automatic online colorimetric measurement of hexavalent chromium (Cr(VI))
Differential photometric absorbance at 525 nm using 1,5-diphenylcarbazide (DPC)
Measuring ranges of 0.5–300 ppb (26 mm cell) and 0.01–1 ppm (16 mm cell), up to 50 ppm with internal dilution
Heated reaction cell
Programmable analysis frequency with 6–8 minute cycle time
Automatic calibration, validation and cleaning
Colour touchscreen interface with multilingual support
Integrated data logger with USB download
Two 4–20 mA outputs and Modbus RTU RS485 communication
Four programmable SPDT potential-free alarm relays
Remote start/stop via digital input
Up to four sample streams with an external sequencer
IP54-rated dual-compartment enclosure separating electrical and liquid-handling components
Standard 100–240 VAC (50/60 Hz) power supply with optional 24 VDC
Benefits
Automated measurement reduces the need for routine manual analysis
Wide measuring range accommodates different chromium concentration levels
Low reagent consumption and adjustable analysis intervals help reduce operating costs
Automatic calibration, validation and cleaning minimise downtime and operator intervention
User-friendly touchscreen simplifies operation and configuration
Separate compartments isolate electronic components from liquid-handling components
Factory-tested design simplifies installation and commissioning
Optional multi-stream configurations support monitoring at multiple sampling points
Specifications
Measurement | |
|---|---|
Measured parameter | Hexavalent chromium (Cr(VI)) |
Measurement units | ppb, ppm, mg/L |
Measuring principle | Differential photometric absorbance |
Colorimetric reagent | 1,5-Diphenylcarbazide (DPC) |
Measurement wavelength | 525 nm |
Light source | LED |
Measuring range (26 mm cell) | 0.5–300 ppb Cr(VI) |
Measuring range (16 mm cell) | 0.01–1 ppm Cr(VI) |
Extended measuring range | Up to 50 ppm Cr(VI) with internal dilution |
Reproducibility (up to 50 ppb) | ±1 ppb or ±5%, whichever is greater |
Reproducibility (≥50 ppb to 300 ppb) | ±2 ppb ±5%, whichever is greater (26 mm cell) |
Reproducibility (≥300 ppb) | ±5 ppb or ±5%, whichever is greater (16 mm cell) |
Analysis | |
|---|---|
Analysis method | Sequential sampling and colorimetric analysis |
Analysis frequency | Freely programmable, batch near-continuous analysis |
Cycle time | 6–8 minutes, including conditioning before analysis and rinsing after measurement |
Reaction cell | Heated colorimetric cell |
Calibration, validation and cleaning | Automatic, with freely selectable intervals |
Reagent Consumption | |
|---|---|
Reagent consumption (16 mm cell) | 0.7 L (0.18 US gal) of each reagent per 30 days at 15-minute analysis intervals |
Reagent consumption (26 mm cell) | 1 L (0.26 US gal) of each reagent per 30 days at 15-minute analysis intervals |
Sample Requirements | |
|---|---|
Sample supply | Pressure-free from overflow vessel |
Sample temperature | 5–50 °C (41–122 °F) |
Sample flow rate | 80–500 mL/min |
Sample connection | 6 mm (¼ in.) |
Drain Requirements | |
|---|---|
Drain type | Pressure-free, atmospheric drain |
Drain connection | 12 mm (½ in.) |
Sample Stream Configuration | |
|---|---|
Single-stream configuration | 1 sample stream |
Dual-stream configuration | 2 sample streams with integrated switching valve |
Multiple-stream configuration | 3 or 4 sample streams with external sequencer |
Display and Data Management | |
|---|---|
User interface | Colour graphic touchscreen |
Display information | Measured values and status information |
Language support | Multiple languages |
Data logging | Integrated data logger with USB download |
Outputs, Communication and Alarms | |
|---|---|
Analogue outputs | 2 × 4–20 mA for measured data |
Digital communication | Modbus RTU RS485 |
Alarm relays | 4 programmable SPDT potential-free relays |
Digital input | Remote start/stop |
Alarm functions | Loss of sample and low reagent alarms |
Electrical | |
|---|---|
Power supply | 100–240 VAC, 50/60 Hz (standard); 24 VDC (optional) |
Power consumption | Maximum 80 VA |
Environmental | |
|---|---|
Operating temperature | 5–45 °C (41–113 °F) |
Operating humidity | 10–90%, non-condensing |
Operating environment | Indoor use; outdoor installation requires a protective cabinet or shelter (not included) |
Physical | |
|---|---|
Dimensions (H × W × D) | 606 × 380 × 209 mm (23.6 × 15.0 × 8.2 in.) |
Weight | Approximately 20 kg (44 lbs) |
Cabinet material | Cold-rolled steel |
Cabinet finish | Epoxy powder coating |
Enclosure configuration | Separate electrical and liquid-handling compartments |
Cabinet doors | Two lockable doors |
Ingress protection | IP54 |
Installation | |
|---|---|
Standard mounting | Wall mounting |
Optional mounting | Benchtop support or panel mounting |
Factory preparation | Factory tested |
Drawing

Downloads
| CA6 - Chromium Analyzer Brochure | DownloadCA6 - Chromium Analyzer Brochure |
|---|
Related Products

ECD
DCA-23 Seawater Dechlorination Analyzer
The DCA-23 monitors seawater chlorination and dechlorination from 0.00–2.00 ppm chlorine equivalent, featuring reagent-free measurement, flow-independent…
Learn More
ECD
DC80 Dechlorination Analyzer
The DC80 Dechlorination Analyzer uses a zero-shift measurement strategy, automatic flow control and pH compensation to monitor very low total chlorine levels…
Learn More
ECD
CD80 Chlorine Dioxide Analyzer
The CD80 is a reagentless, panel-mounted analyzer for continuous chlorine dioxide (ClO₂) monitoring from 0.05 to 20 ppm, featuring automatic flow control, a…
Learn More
ECD
TC80 Total Chlorine Analyzer
The TC80 is a reagentless, panel-mounted analyzer for total chlorine monitoring in water, featuring amperometric sensing, automatic pH compensation, automatic…
Learn More