WETEC

ECD

CA6 Colorimetric Analyzer

WaterWater Quality Measurement

Online sequential sampling colorimetric analyzer for measuring analyte concentrations in water.

  • LED light source and quartz reaction cell

  • Programmable analysis cycles with adjustable cycle time

  • 4–20 mA analog output and two configurable alarm relays

  • Touchscreen interface with user-friendly menu structure

  • Epoxy powder-coated steel cabinet with separate electrical and liquid compartments

Quantity
1

Overview

The ECD CA-6 Colorimetric Analyzer is an online sequential sampling analyzer designed to measure analyte concentrations in water. It uses an LED light source, a quartz reaction cell and a photometer to determine concentrations through differential absorbance measurements taken before and after the addition of color-forming reagents.

The analyzer features programmable analysis sequences, an adjustable cycle time and a touchscreen interface for process configuration. Its epoxy powder-coated steel cabinet consists of two separate lockable compartments housing the electrical and liquid components. It provides a 4–20 mA analog output and two configurable alarm relays, with measurement ranges depending on the selected analyte and model.

Typical Applications

  • Online measurement of metals such as aluminum, copper, iron, manganese, nickel and zinc

  • Monitoring of nutrients such as ammonia, nitrite and phosphate

  • Measurement of water chemistry parameters such as chlorine, chloride, hardness, silica and sulfate

  • Integration with process control and monitoring systems through a 4–20 mA analog output

Key Features

  • Online sequential sampling using colorimetric technology

  • LED light source, quartz reaction cell and photometer

  • Differential absorbance measurement using reference and reading values

  • Programmable analysis sequences with adjustable cycle time

  • Configurable analysis using up to four reagents

  • Touchscreen interface with user-friendly menu structure

  • User-defined extra cycles for calibration or cleaning

  • 4–20 mA analog output

  • Two configurable alarm relays, with loss of sample and low reagent alarm functions

  • Separate lockable electrical and liquid compartments

  • Rugged epoxy powder-coated cold-rolled steel cabinet

  • Operating temperature range of 5…50 °C

  • Power supply of 110–220 Vac, 50–60 Hz, 80 VA

  • Factory calibrated

  • Standard wall mounting with optional benchtop support

  • Optional dilution module with 10:1 or 50:1 dilution ratios

Benefits

  • Automated sequential sampling and analysis for online water monitoring

  • Reference measurement compensates for sample color, turbidity and optical variations

  • Easy installation and start-up with factory calibration

  • Simplified operation and process configuration through a touchscreen interface

  • Minimized reagent consumption through adjustable analysis cycle time

  • Low maintenance requirements

  • Notification of sample loss and low reagent levels

  • Flexible installation with standard wall mounting or optional benchtop support

  • Extended measurement ranges through the optional dilution module

Specifications

Measurement

Measuring method

Photometric differential absorbance

Sampling method

Online sequential sampling

Light source

Long-life LED

Optical detector

Photometer

Reaction cell

Quartz reaction cell

Response time

Dependent on the specific colorimetric measurement

Repeatability

±2% on absorbance value at turbidity below 80 NTU

Drift

±2% per month on absorbance measurement

Calibration

Factory calibrated

 

 

Measuring Parameters and Ranges

Model

Range A

Range B

Range C

Aluminum

CA6-01-X

0–1.00 mg/L

0–10.0 mg/L

0–50.0 mg/L

Ammonia

CA6-02-X

0–1.0 mg/L

0–10.0 mg/L

0–50.0 mg/L

Chloride

CA6-03-X

0–3.0 mg/L

0–30.0 mg/L

0–150.0 mg/L

Chlorine (free-total)

CA6-04-X

0–3.0 mg/L

0–30.0 mg/L

0–150.0 mg/L

Chromium VI

CA6-05-X

0–1.0 mg/L

0–10.0 mg/L

0–50.0 mg/L

Copper

CA6-06-X

0–5.0 mg/L

0–50.0 mg/L

0–250.0 mg/L

Cyanide (free)

CA6-07-X

0–200 µg/L

0–2.0 mg/L

0–10.0 mg/L

Hardness

CA6-08-X

0–1.0 mg/L

0–10.0 mg/L

0–50.0 mg/L

Iron

CA6-09-X

0–0.1 mg/L

0–1.0 mg/L

0–5.0 mg/L

Manganese

CA6-10-X

0–100 µg/L

0–1.0 mg/L

0–5.0 mg/L

Nickel

CA6-11-X

0–3.0 mg/L

0–30.0 mg/L

0–150.0 mg/L

Nitrite

CA6-13-X

0–600 µg/L

0–6.0 mg/L

0–30.0 mg/L

Hydrazine

CA6-14-X

0–500 µg/L

0–5.0 mg/L

0–20.0 mg/L

Phosphate

CA6-15-X

0–5.0 mg/L

0–50.0 mg/L

0–200 mg/L

Total phosphorus

CA6-16-X

0–2.0 mg/L

0–20.0 mg/L

0–100 mg/L

Silica

CA6-17-X

0–1.0 mg/L

0–10.0 mg/L

0–20.0 mg/L

Sulfate

CA6-18-X

0–50 mg/L

0–500 mg/L

0–2500 mg/L

Zinc

CA6-20-X

0–1.0 mg/L

0–20.0 mg/L

0–100 mg/L

Phosphate & total phosphorus

CA6-21-A

1.0–100 mg/L

—

—

Phenoyl*

CA6-22-X

1.0–100 mg/L

0–200.0 mg/L

—

 

 

Analysis Configuration

Measurement sequence

Reference measurement and reading measurement

Number of reagents

Up to four

Analysis cycle time

Adjustable

Measurements per analysis cycle

Typically one

Extra cycle

User-defined calibration or cleaning sequence

Reagent consumption

Approximately 2,500 tests per litre of reagent, depending on the specific colorimetric measurement

Dilution module (optional)

10:1 or 50:1 dilution ratios

 

 

Sample Requirements and Connections

Inlet sample pressure

Atmospheric

Outlet sample pressure

Atmospheric

Sample flow for fast loop reservoir

100–500 mL/min

Sample connection

Flexible tubing with ¼ inch outer diameter, connected to the fast loop reservoir

Waste tubing

⅜ inch outer diameter

 

 

Outputs and Alarms

Analog output

4–20 mA

Alarm relays

Two configurable relays

Alarm functions

Loss of sample and low reagent levels

 

 

User Interface

Interface

Touchscreen

Display information

Measured parameter, operating status and reagent volume percentage

Menu options

RUN, DISPLAY, PROGRAM, SERVICE and HELP

 

 

Electrical

Power supply

110–220 Vac

Frequency

50–60 Hz

Power rating

80 VA

 

 

Environmental

Operating temperature

5…50 °C

 

 

Mechanical

Cabinet material

Cold-rolled steel

Cabinet finish

Epoxy powder-coated

Enclosure configuration

Two separate compartments with lockable doors

Upper compartment

Electrical enclosure

Lower compartment

Liquids enclosure

Dimensions (L × H × D)

380 × 600 × 210 mm (15 × 24 × 8.25 inches)

Weight

Approximately 17 kg (37.5 lbs)

Mounting

Standard wall mounting or optional benchtop support

Drawing

CA6 Colorimetric Analyzer Drawing

Downloads

CA6 - Colorimetric Analyzer Brochure DownloadCA6 - Colorimetric Analyzer Brochure