ECD
CA6 Hydrazine Analyzer
Compact online colorimetric analyzer for automatic measurement of hydrazine in water.
Measuring range: 0–500 ppb (µg/L) without dilution
Automatic calibration, validation and cleaning
Dual 4–20 mA outputs and Modbus RTU RS485 communication
Colour touchscreen interface with integrated data logger
Dual-compartment enclosure with IP54 protection
Overview
The ECD CA-6 Hydrazine Analyzer is a compact online colorimetric analyzer designed for the automatic measurement of hydrazine in water. It uses differential photometric absorbance with a heated reaction cell, covering a measuring range of 0–500 ppb (µg/L) without dilution and a 10-minute analysis cycle.
The analyzer features a dual-compartment enclosure that separates the electrical and liquid components, a colour touchscreen interface with an integrated data logger, and automatic calibration, validation and cleaning functions. It provides two 4–20 mA outputs, Modbus RTU RS485 communication and four programmable SPDT alarm relays. The IP54-rated enclosure supports wall mounting as standard, with optional benchtop or panel mounting configurations.
Typical Applications
Hydrazine monitoring in power generation plants
Cooling water monitoring
Hydrazine measurement in water-steam cycles
Boiler feedwater monitoring
Control and optimisation of oxygen scavenger systems
Key Features
Differential photometric absorbance measurement at 430 nm
Hydrazine measuring range of 0–500 ppb (µg/L) without dilution
Reproducibility of ±1 ppb or ±3%, whichever is greater
Programmable analysis frequency with a 10-minute analysis cycle
Heated colorimetric reaction cell
Single-reagent configuration
Automatic calibration, validation and cleaning
Colour touchscreen interface with multilingual support
Integrated data logger with USB data download
Two 4–20 mA outputs and Modbus RTU RS485 communication
Four programmable SPDT potential-free alarm relays
Remote start/stop digital input
Single or dual sample stream configuration, with three or four streams supported by an external sequencer
Rugged, epoxy powder-coated cold-rolled steel cabinet with separate electrical and liquid compartments
IP54 enclosure protection
Standard wall mounting with optional benchtop or panel mounting
Benefits
Reduced operating costs through low reagent consumption
Reduced maintenance requirements, downtime and operator intervention
Easy operation and flexible process configuration
Simplified installation with a factory-tested, ready-to-install system
Improved protection of electronic components from liquid-handling components
Convenient access to stored measurement data
Flexible integration with external process control and monitoring systems
Notification of sample loss and low reagent levels
Specifications
Measurement | |
|---|---|
Measured parameter | Hydrazine (N₂H₄) |
Measurement units | ppb, ppm, mg/L |
Measuring principle | Differential photometric absorbance |
Measuring wavelength | 430 nm |
Measuring range | 0–500 ppb (µg/L) without dilution |
Reproducibility | ±1 ppb or ±3%, whichever is greater |
Analysis frequency | Freely programmable, batch near-continuous analysis |
Cycle time | 10 minutes, including conditioning before analysis and rinsing after measurement |
Reaction cell | Heated colorimetric reaction cell |
Reagent configuration | Single reagent |
Sample Requirements | |
|---|---|
Sample pressure | Pressure-free from overflow vessel |
Sample temperature | 5–45 °C (41–113 °F) |
Sample flow rate | 80–500 mL/min |
Sample connection | 6 mm (¼ in.) |
Number of sample streams | 1 or 2 (integrated switching valve for dual streams); 3 or 4 with external sequencer |
Drain Requirements | |
|---|---|
Drain type | Pressure-free, atmospheric drain |
Drain connection | 12 mm (½ in.) |
Calibration, Validation and Cleaning | |
|---|---|
Calibration, validation and cleaning | Automatic |
Calibration, validation and cleaning intervals | Freely selectable |
Display and Data Management | |
|---|---|
User interface | Colour graphic touchscreen |
Display information | Measured values and status information |
Language support | Multiple languages |
Data logger | Integrated |
Data download | USB |
Outputs, Communication and Alarms | |
|---|---|
Analogue outputs | 2 × 4–20 mA for measured data |
Digital communication | Modbus RTU RS485 |
Alarm relays | 4 programmable SPDT potential-free relays |
Alarm functions | Loss of sample and low reagent levels |
Digital input | Remote start/stop |
Power Supply | |
|---|---|
Standard power supply | 100–240 VAC, 50/60 Hz |
Optional power supply | 24 VDC |
Power consumption | Maximum 80 VA |
Environmental Conditions | |
|---|---|
Operating temperature | 5–45 °C (41–113 °F) |
Humidity | 10–90%, non-condensing |
Operating environment | Indoor use; outdoor installation requires a protective cabinet or shelter (not included) |
Mechanical | |
|---|---|
Cabinet material | Cold-rolled steel |
Cabinet finish | Epoxy powder coating |
Enclosure configuration | Separate electrical and liquid compartments |
Doors | Two lockable doors |
Dimensions (H × W × D) | 606 × 380 × 209 mm / 23.6 × 15.0 × 8.2 in. |
Weight | Approximately 20 kg (44 lbs) |
Ingress protection | IP54 |
Installation | |
|---|---|
Standard installation | Wall mounting |
Optional installation | Benchtop support or panel mounting |
Drawing

Downloads
| CA6 - Hydrazine Analyzer Brochure | DownloadCA6 - Hydrazine Analyzer Brochure |
|---|
Related Products

ECD
CA6 Iron Analyzer
The CA-6 Iron Analyzer provides automatic online colorimetric measurement of iron in water, featuring low reagent consumption, automatic calibration…
Learn More
ECD
CA6 Manganese Analyzer
The CA-6 Manganese Analyzer provides automatic online colorimetric measurement of manganese in water, featuring low reagent consumption, automatic calibration…
Learn More
ECD
DCA-23 Seawater Dechlorination Analyzer
The DCA-23 monitors seawater chlorination and dechlorination from 0.00–2.00 ppm chlorine equivalent, featuring reagent-free measurement, flow-independent…
Learn More
ECD
DC80 Dechlorination Analyzer
The DC80 Dechlorination Analyzer uses a zero-shift measurement strategy, automatic flow control and pH compensation to monitor very low total chlorine levels…
Learn More