ECD
CA6 Manganese Analyzer
Online colorimetric analyzer for automatic measurement of manganese in water using the PAN method.
Measuring range: 1–200 ppb (26 mm cell) or 5–1000 ppb (16 mm cell), up to 50 ppm with internal dilution
Automatic calibration, validation and cleaning
Programmable analysis frequency with 8–10 minute cycle time
Low reagent consumption and colour touchscreen interface
Dual 4–20 mA outputs and Modbus RTU RS485 communication
Dual-compartment enclosure with IP54 protection
Overview
The ECD CA-6 Manganese Analyzer is a compact online colorimetric analyzer designed for automatic measurement of manganese in water using the PAN method. It operates through differential photometric absorbance at 572 nm, covering measuring ranges of 1–200 ppb with a 26 mm cell or 5–1000 ppb with a 16 mm cell, extending up to 50 ppm with internal dilution.
The analyzer features a heated reaction cell, programmable analysis frequency and automatic calibration, validation and cleaning functions. Its rugged, epoxy powder-coated steel cabinet incorporates separate liquid and electrical compartments with IP54 protection. Equipped with a colour touchscreen interface, integrated data logger, dual 4–20 mA outputs and Modbus RTU RS485 communication, the CA-6 supports water quality monitoring with low reagent consumption and reduced operator intervention.
Typical Applications
Manganese concentration monitoring in drinking water
Manganese concentration monitoring in industrial wastewater
Manganese concentration monitoring in municipal wastewater
Key Features
Automatic online sequential analysis using the PAN colorimetric method with differential photometric absorbance at 572 nm
Measuring range of 1–200 ppb (26 mm cell) or 5–1000 ppb (16 mm cell), extending up to 50 ppm with internal dilution
LED light source and heated colorimetric reaction cell
Programmable analysis frequency with 8–10 minute cycle time
Automatic calibration, validation and cleaning with selectable intervals
Colour touchscreen interface with multilingual menu
Integrated data logger with USB data download
Dual 4–20 mA outputs and Modbus RTU RS485 communication
Four programmable SPDT potential-free alarm relays
Remote start/stop digital input
Up to four sample streams with an external sequencer
Rugged epoxy powder-coated steel cabinet with separate liquid and electrical compartments
IP54 enclosure protection
Loss of sample and low reagent alarms
Benefits
Reference measurement compensates for interference from sample colour, turbidity, reagent colour and refractive index variations
Low reagent consumption minimises operating costs
Automated analysis, calibration, validation and cleaning reduce manual workload and downtime
User-friendly touchscreen simplifies operation and process configuration
Wide measuring range accommodates different manganese concentrations
Separate liquid and electrical compartments isolate electronic components from the wet section
Factory-tested system simplifies installation and commissioning
Sample loss and low reagent alarms provide timely notifications of operating issues
Flexible communication outputs facilitate integration with monitoring and control systems
Specifications
Measurement | |
|---|---|
Measured parameter | Manganese (Mn²⁺) |
Measurement units | ppb, ppm, mg/L |
Measuring principle | Differential photometric absorbance using the PAN method |
Measurement wavelength | 572 nm |
Measuring range | 1–200 ppb Mn²⁺ (26 mm cell); 5–1000 ppb Mn²⁺ (16 mm cell); up to 50 ppm Mn²⁺ with internal dilution |
Reproducibility | ±3 ppb or ±5%, whichever is greater (26 mm cell); ±10 ppb or ±5%, whichever is greater (16 mm cell) |
Analysis frequency | Freely programmable, batch near-continuous analysis |
Cycle time | 8–10 minutes, including conditioning before analysis and rinsing after measurement |
Light source | LED |
Reaction cell | Heated colorimetric reaction cell |
Sample Requirements | |
|---|---|
Sample supply condition | Pressure-free from overflow vessel |
Sample temperature | +5…+50 °C (+41…+122 °F) |
Sample flow rate | 80–500 mL/min |
Sample connection | 6 mm (¼ in.) |
Number of sample streams | 1 stream; 2 streams with integrated switching valve; 3 or 4 streams with external sequencer |
Drain | |
|---|---|
Drain condition | Pressure-free, atmospheric drain |
Drain connection | 12 mm (½ in.) |
Calibration and Maintenance | |
|---|---|
Calibration, validation and cleaning | Automatic, with freely selectable intervals |
Reagent consumption | 1.2 L (0.32 US gal) per reagent for the 16 mm cell; 2.5 L (0.66 US gal) per reagent for the 26 mm cell, every 30 days at a 15-minute analysis frequency |
User Interface and Data Management | |
|---|---|
Display | Colour graphic touchscreen displaying measured values and status information |
Language support | Multiple languages |
Data logger | Integrated data logger with USB download |
Outputs, Communication and Alarms | |
|---|---|
Analogue outputs | 2 × 4–20 mA outputs for measured data |
Digital communication | Modbus RTU RS485 |
Alarm relays | 4 programmable SPDT potential-free relays |
Alarm functions | Loss of sample and low reagent alarms |
Digital input | Remote start/stop |
Power Supply | |
|---|---|
Supply voltage | 100–240 VAC, 50/60 Hz (standard); 24 VDC (optional) |
Power consumption | Maximum 80 VA |
Operating Conditions | |
|---|---|
Operating temperature | +5…+45 °C (+41…+113 °F) |
Humidity | 10–90%, non-condensing |
Operating environment | Indoor use; outdoor installation requires a protective cabinet or shelter (not included) |
Mechanical | |
|---|---|
Cabinet material | Cold-rolled steel |
Surface coating | Epoxy powder coating |
Enclosure configuration | Separate liquid and electrical compartments |
Access doors | Two lockable doors |
Dimensions (H × W × D) | 606 × 380 × 209 mm (23.6 × 15.0 × 8.2 in.) |
Weight | Approximately 20 kg (44 lbs) |
Ingress protection | IP54 |
Installation | |
|---|---|
Mounting configuration | Wall mounting (standard); benchtop support or panel mounting (optional) |
Drawing

Downloads
| CA6 - Manganese Analyzer Brochure | DownloadCA6 - Manganese Analyzer Brochure |
|---|
Related Products

ECD
CA6 Hydrazine Analyzer
The CA-6 Hydrazine Analyzer provides automatic online colorimetric measurement of hydrazine in water, featuring low reagent consumption, automatic…
Learn More
ECD
CA6 Iron Analyzer
The CA-6 Iron Analyzer provides automatic online colorimetric measurement of iron in water, featuring low reagent consumption, automatic calibration…
Learn More
ECD
DCA-23 Seawater Dechlorination Analyzer
The DCA-23 monitors seawater chlorination and dechlorination from 0.00–2.00 ppm chlorine equivalent, featuring reagent-free measurement, flow-independent…
Learn More
ECD
DC80 Dechlorination Analyzer
The DC80 Dechlorination Analyzer uses a zero-shift measurement strategy, automatic flow control and pH compensation to monitor very low total chlorine levels…
Learn More