WETEC

ECD

CA6 Iron Analyzer

WaterWater Quality Measurement

Online colorimetric analyzer for automatic measurement of ferrous, ferric and total dissolved iron in water.

  • Measuring ranges: 2–250 ppb (26 mm cell) or 9–500 ppb (16 mm cell), up to 10 mg/L with internal dilution

  • Automatic calibration, validation and cleaning

  • Low reagent consumption with adjustable measurement frequency

  • Colour touchscreen interface with integrated data logger

  • Two 4–20 mA outputs and Modbus RTU RS485 communication

  • Dual-compartment enclosure with IP54 protection

Quantity
1

Overview

The ECD CA-6 Iron Analyzer is a compact online colorimetric analyzer designed for the automatic measurement of ferrous iron (Fe²⁺), ferric iron (Fe³⁺) and total dissolved iron in water. It uses differential photometric absorbance with the ferrozine method at 572 nm, offering measuring ranges of 2–250 ppb with a 26 mm reaction cell or 9–500 ppb with a 16 mm reaction cell, extendable up to 10 mg/L with an internal dilution module.

The analyzer features automatic calibration, validation and cleaning, with programmable analysis intervals and low reagent consumption. It incorporates a colour touchscreen interface, an integrated data logger with USB download, two 4–20 mA outputs and Modbus RTU RS485 communication. Its rugged, epoxy powder-coated steel cabinet provides IP54 protection, with separate liquid and electrical compartments.

Typical Applications

  • Iron monitoring in drinking water and surface water

  • Iron removal process and residual coagulant monitoring

  • Industrial wastewater and effluent monitoring

  • Boiler feed water monitoring

  • Cooling water monitoring

  • Corrosion control

Key Features

  • Online measurement of ferrous iron (Fe²⁺), ferric iron (Fe³⁺) and total dissolved iron

  • Differential photometric absorbance using the ferrozine method at 572 nm

  • Measuring ranges of 2–250 ppb (26 mm cell) or 9–500 ppb (16 mm cell), extendable up to 10 mg/L with internal dilution

  • Programmable analysis frequency with an 8–10 minute analysis cycle

  • Heated colorimetric reaction cell

  • Automatic calibration, validation and cleaning

  • Single-reagent configuration

  • Colour touchscreen interface with multilingual support

  • Integrated data logger with USB download

  • Two 4–20 mA outputs and Modbus RTU RS485 communication

  • Four programmable SPDT alarm relays

  • Loss of sample and low reagent alarms

  • Single or dual sample stream configurations, expandable up to four streams with an external sequencer

  • Epoxy powder-coated steel cabinet with IP54 protection and separate liquid and electrical compartments

  • Standard 100–240 VAC, 50/60 Hz power supply with optional 24 VDC

Benefits

  • Reduced operating costs through low reagent consumption

  • Reduced operator intervention and downtime through automated maintenance functions

  • Easy operation and process configuration through a user-friendly interface

  • Simplified installation with a factory-tested, ready-to-install system

  • Reduced measurement interference from sample colour, turbidity and refractive index variations

  • Flexible monitoring of multiple sample streams to suit different process requirements

  • Complete separation of liquid-handling components and electronics

Specifications

Measurement

Measured parameters

Ferrous iron (Fe²⁺), ferric iron (Fe³⁺), total dissolved iron

Measurement units

ppb, ppm, mg/L

Measuring principle

Differential photometric absorbance

Analytical method

Ferrozine colorimetric method

Measurement wavelength

572 nm

Measuring range

2–250 ppb (26 mm cell); 9–500 ppb (16 mm cell); up to 10 mg/L with internal dilution

Reproducibility

±1 ppb or ±5%, whichever is greater (26 mm cell); ±5 ppb or ±5%, whichever is greater (16 mm cell)

Analysis frequency

Freely programmable, batch near-continuous analysis

Cycle time

8–10 minutes, including conditioning before analysis and rinsing after measurement

Reaction cell

Heated colorimetric cell

 

 

Reagent and Automatic Maintenance

Reagent configuration

Single reagent

Reagent consumption

2.5 L (0.66 US gal) for the 16 mm cell; 5 L (1.32 US gal) for the 26 mm cell, every 60 days at a 15-minute analysis frequency

Automatic maintenance functions

Calibration, validation and cleaning with freely selectable intervals

 

 

Sample and Drain

Sample pressure

Pressure-free from overflow vessel

Sample temperature

+5…+50 °C (+41…+122 °F)

Sample flow rate

80–500 mL/min

Sample connection

6 mm (¼ in)

Drain condition

Pressure-free, atmospheric drain

Drain connection

12 mm (½ in)

Number of sample streams

1 stream; 2 streams with integrated switching valve; 3 or 4 streams with external sequencer

 

 

Display and Data Management

User interface

Colour graphical touchscreen

Display information

Measured values and status information

Language support

Multiple languages

Data logger

Integrated data logger with USB download

 

 

Outputs, Communication and Alarms

Analogue outputs

2 × 4–20 mA for measured data

Digital communication

Modbus RTU RS485

Alarm relays

4 programmable SPDT potential-free relays

Alarm functions

Loss of sample and low reagent alarms

Digital input

Remote start/stop

 

 

Power Supply

Standard power supply

100–240 VAC, 50/60 Hz

Optional power supply

24 VDC

Power consumption

Maximum 80 VA

 

 

Environmental Conditions

Operating temperature

+5…+45 °C (+41…+113 °F)

Operating humidity

10–90%, non-condensing

Operating environment

Indoor use; outdoor installation requires a protective cabinet or shelter (not included)

 

 

Mechanical Specifications and Installation

Cabinet material

Cold-rolled steel

Cabinet finish

Epoxy powder-coated

Enclosure design

Separate liquid and electrical compartments with two lockable doors

Dimensions (H × W × D)

606 × 380 × 209 mm (23.6 × 15.0 × 8.2 in), as stated in the datasheet

Weight

Approximately 20 kg (44 lb)

Ingress protection

IP54

Standard installation

Wall mounting

Optional installation

Benchtop support or panel mounting

Drawing

CA6 Iron Analyzer Drawing

Downloads

CA6 - Iron Ferrozine Analyzer Brochure DownloadCA6 - Iron Ferrozine Analyzer Brochure