ECD
CA6 Iron Analyzer
Online colorimetric analyzer for automatic measurement of ferrous, ferric and total dissolved iron in water.
Measuring ranges: 2–250 ppb (26 mm cell) or 9–500 ppb (16 mm cell), up to 10 mg/L with internal dilution
Automatic calibration, validation and cleaning
Low reagent consumption with adjustable measurement frequency
Colour touchscreen interface with integrated data logger
Two 4–20 mA outputs and Modbus RTU RS485 communication
Dual-compartment enclosure with IP54 protection
Overview
The ECD CA-6 Iron Analyzer is a compact online colorimetric analyzer designed for the automatic measurement of ferrous iron (Fe²⁺), ferric iron (Fe³⁺) and total dissolved iron in water. It uses differential photometric absorbance with the ferrozine method at 572 nm, offering measuring ranges of 2–250 ppb with a 26 mm reaction cell or 9–500 ppb with a 16 mm reaction cell, extendable up to 10 mg/L with an internal dilution module.
The analyzer features automatic calibration, validation and cleaning, with programmable analysis intervals and low reagent consumption. It incorporates a colour touchscreen interface, an integrated data logger with USB download, two 4–20 mA outputs and Modbus RTU RS485 communication. Its rugged, epoxy powder-coated steel cabinet provides IP54 protection, with separate liquid and electrical compartments.
Typical Applications
Iron monitoring in drinking water and surface water
Iron removal process and residual coagulant monitoring
Industrial wastewater and effluent monitoring
Boiler feed water monitoring
Cooling water monitoring
Corrosion control
Key Features
Online measurement of ferrous iron (Fe²⁺), ferric iron (Fe³⁺) and total dissolved iron
Differential photometric absorbance using the ferrozine method at 572 nm
Measuring ranges of 2–250 ppb (26 mm cell) or 9–500 ppb (16 mm cell), extendable up to 10 mg/L with internal dilution
Programmable analysis frequency with an 8–10 minute analysis cycle
Heated colorimetric reaction cell
Automatic calibration, validation and cleaning
Single-reagent configuration
Colour touchscreen interface with multilingual support
Integrated data logger with USB download
Two 4–20 mA outputs and Modbus RTU RS485 communication
Four programmable SPDT alarm relays
Loss of sample and low reagent alarms
Single or dual sample stream configurations, expandable up to four streams with an external sequencer
Epoxy powder-coated steel cabinet with IP54 protection and separate liquid and electrical compartments
Standard 100–240 VAC, 50/60 Hz power supply with optional 24 VDC
Benefits
Reduced operating costs through low reagent consumption
Reduced operator intervention and downtime through automated maintenance functions
Easy operation and process configuration through a user-friendly interface
Simplified installation with a factory-tested, ready-to-install system
Reduced measurement interference from sample colour, turbidity and refractive index variations
Flexible monitoring of multiple sample streams to suit different process requirements
Complete separation of liquid-handling components and electronics
Specifications
Measurement | |
|---|---|
Measured parameters | Ferrous iron (Fe²⁺), ferric iron (Fe³⁺), total dissolved iron |
Measurement units | ppb, ppm, mg/L |
Measuring principle | Differential photometric absorbance |
Analytical method | Ferrozine colorimetric method |
Measurement wavelength | 572 nm |
Measuring range | 2–250 ppb (26 mm cell); 9–500 ppb (16 mm cell); up to 10 mg/L with internal dilution |
Reproducibility | ±1 ppb or ±5%, whichever is greater (26 mm cell); ±5 ppb or ±5%, whichever is greater (16 mm cell) |
Analysis frequency | Freely programmable, batch near-continuous analysis |
Cycle time | 8–10 minutes, including conditioning before analysis and rinsing after measurement |
Reaction cell | Heated colorimetric cell |
Reagent and Automatic Maintenance | |
|---|---|
Reagent configuration | Single reagent |
Reagent consumption | 2.5 L (0.66 US gal) for the 16 mm cell; 5 L (1.32 US gal) for the 26 mm cell, every 60 days at a 15-minute analysis frequency |
Automatic maintenance functions | Calibration, validation and cleaning with freely selectable intervals |
Sample and Drain | |
|---|---|
Sample pressure | Pressure-free from overflow vessel |
Sample temperature | +5…+50 °C (+41…+122 °F) |
Sample flow rate | 80–500 mL/min |
Sample connection | 6 mm (¼ in) |
Drain condition | Pressure-free, atmospheric drain |
Drain connection | 12 mm (½ in) |
Number of sample streams | 1 stream; 2 streams with integrated switching valve; 3 or 4 streams with external sequencer |
Display and Data Management | |
|---|---|
User interface | Colour graphical touchscreen |
Display information | Measured values and status information |
Language support | Multiple languages |
Data logger | Integrated data logger with USB download |
Outputs, Communication and Alarms | |
|---|---|
Analogue outputs | 2 × 4–20 mA for measured data |
Digital communication | Modbus RTU RS485 |
Alarm relays | 4 programmable SPDT potential-free relays |
Alarm functions | Loss of sample and low reagent alarms |
Digital input | Remote start/stop |
Power Supply | |
|---|---|
Standard power supply | 100–240 VAC, 50/60 Hz |
Optional power supply | 24 VDC |
Power consumption | Maximum 80 VA |
Environmental Conditions | |
|---|---|
Operating temperature | +5…+45 °C (+41…+113 °F) |
Operating humidity | 10–90%, non-condensing |
Operating environment | Indoor use; outdoor installation requires a protective cabinet or shelter (not included) |
Mechanical Specifications and Installation | |
|---|---|
Cabinet material | Cold-rolled steel |
Cabinet finish | Epoxy powder-coated |
Enclosure design | Separate liquid and electrical compartments with two lockable doors |
Dimensions (H × W × D) | 606 × 380 × 209 mm (23.6 × 15.0 × 8.2 in), as stated in the datasheet |
Weight | Approximately 20 kg (44 lb) |
Ingress protection | IP54 |
Standard installation | Wall mounting |
Optional installation | Benchtop support or panel mounting |
Drawing

Downloads
| CA6 - Iron Ferrozine Analyzer Brochure | DownloadCA6 - Iron Ferrozine Analyzer Brochure |
|---|
Related Products

ECD
CA6 Hydrazine Analyzer
The CA-6 Hydrazine Analyzer provides automatic online colorimetric measurement of hydrazine in water, featuring low reagent consumption, automatic…
Learn More
ECD
CA6 Manganese Analyzer
The CA-6 Manganese Analyzer provides automatic online colorimetric measurement of manganese in water, featuring low reagent consumption, automatic calibration…
Learn More
ECD
DCA-23 Seawater Dechlorination Analyzer
The DCA-23 monitors seawater chlorination and dechlorination from 0.00–2.00 ppm chlorine equivalent, featuring reagent-free measurement, flow-independent…
Learn More
ECD
DC80 Dechlorination Analyzer
The DC80 Dechlorination Analyzer uses a zero-shift measurement strategy, automatic flow control and pH compensation to monitor very low total chlorine levels…
Learn More