FCX80 Hazardous Location Free Chlorine Analyzer
Panel-mounted free chlorine analyzer for hazardous locations with automatic pH compensation and flow control.
Free chlorine measurement: 0.05–20 ppm (standard range) or 0.01–5.00 ppm (low range)
Reagentless monitoring and EPA Method 334.0 compliance
Sample flow rate: 38–300 L/hr (10–80 gal/hr)
4–20 mA output, optional pH output, Modbus RTU and three alarm relays
X80 analyzer/transmitter with graphical LCD and 24 VDC power supply
Overview
The ECD FCX80 is a panel-mounted free chlorine analyzer designed for hazardous locations, providing reagentless monitoring of drinking water, rinse water, cooling water and other fresh water samples. It offers a standard measurement range of 0.05–20 ppm or a low range of 0.01–5.00 ppm and complies with EPA Method 334.0 for drinking water measurement.
The system incorporates a free chlorine sensor, a pH sensor, a constant-head flow control device and an X80 analyzer/transmitter. Automatic pH compensation ensures accurate compensation for samples between pH 6 and 9, while the flow control device maintains optimum sensor flow at incoming sample flow rates of 38–300 L/hr.
The X80 operates on 24 VDC and features a graphical LCD, a 4–20 mA output for free chlorine, an optional 4–20 mA output for pH, Modbus RTU communication and three alarm relays. The plug-and-play design simplifies installation, with calibration performed by DPD comparison.
Typical Applications
Free chlorine monitoring in drinking water
Free chlorine measurement in rinse water
Free chlorine monitoring in cooling water
Free chlorine analysis in other fresh water samples
Key Features
Panel-mounted, plug-and-play design for hazardous locations
Free chlorine measurement range of 0.05–20 ppm (standard) or 0.01–5.00 ppm (low range)
Reagentless amperometric free chlorine measurement
Automatic pH compensation for samples between pH 6 and 9
Constant-head flow control device supporting 38–300 L/hr sample flow
Compliance with EPA Method 334.0 for drinking water measurement
X80 analyzer/transmitter with graphical LCD and configurable line, bar or gauge displays
4–20 mA free chlorine output, optional 4–20 mA pH output, Modbus RTU and three alarm relays
24 VDC power supply
ATEX/IECEx or FM-approved X80 transmitter and sensors, depending on configuration
Non-intrusive calibration by DPD comparison
ECD digital intelligent sensors with factory calibration and data storage
Benefits
Lower operating costs through reagentless monitoring
Easy installation with a complete, ready-to-use panel-mounted system
Eliminates the need for expensive pH conditioning systems
Maintains optimum sensor flow without pressure regulators or rotameters
Reduced risk of clogging and blockages with large flow tubes and cells
Easy cleaning and routine maintenance
Suitable for drinking water monitoring in accordance with EPA Method 334.0
Supports operation in hazardous locations with appropriately approved instruments
Easy interpretation of measurements through graphical and numerical displays
Simple navigation and configuration through an intuitive menu
Non-intrusive calibration without opening the enclosure or declassifying the area
Specifications
Sensor and Measurement | |
|---|---|
Sensor type | Polarographic |
Electrode materials | Gold/Silver |
Membrane material | PTFE |
Sensor communication | Digital communication |
Free chlorine measurement range (standard range) | 0.05–20 ppm |
Free chlorine measurement range (low range) | 0.01–5.00 ppm |
pH measurement range | 0–14 pH |
Response time | T90 in 2 minutes |
Electrolyte life | Up to 12 months |
Calibration method | DPD comparison |
Flow Control and Process Connections | |
|---|---|
Flow control device | Constant-head flow control device (CFD) |
Minimum and maximum sample flow | 38–300 L/hr (10–80 gal/hr) |
Sample drain condition | Atmospheric pressure |
Inlet connection | ¼″ barb fitting (¼″ FNPT) |
Drain connection | ¾″ FNPT |
Wetted materials | PVC, PP, PVDF, PTFE, glass and 316 stainless steel |
Operating Conditions | |
|---|---|
Operating temperature (sensor and flow train) | 0–50 °C (32–122 °F) |
X80 Analyzer/Transmitter | |
|---|---|
Chlorine measurement indication | 000.0–999.9, ppb/ppm auto-ranging |
Temperature measurement range | 0–100 °C (32–212 °F) |
pH compensation | pH 5–10; accurate compensation between pH 6 and 9, with accuracy degrading rapidly above pH 9 |
Display type | Graphical LCD |
Display resolution | 128 × 64 pixels |
Display dimensions | 2.75″ × 1.5″ |
Display colour | Black on grey background or blue on white background with LED backlight |
Graphical display formats | Line, bar or gauge |
Outputs and Communication | |
|---|---|
Free chlorine analogue output | 1 × 4–20 mA, set to sensor range |
pH analogue output | 1 × 4–20 mA (optional), set to 0–14 pH |
Digital communication | Modbus RTU |
Alarm relays | 3 × SPDT, Form C |
Alarm relay rating | 250 VAC, 10 A |
Power Supply | |
|---|---|
Nominal supply voltage | 24 VDC |
Input voltage range | 18–36 VDC |
Minimum power requirement | 250 mW |
System Configuration | |
|---|---|
Mounting | Panel-mounted |
System design | Plug-and-play |
Integrated components | Free chlorine sensor, pH sensor, constant-head flow control device and X80 analyzer/transmitter |
Compliance and Approvals | |
|---|---|
Drinking water measurement compliance | EPA Method 334.0 |
Hazardous location approvals | ATEX/IECEx or FM-approved X80 transmitter and sensors, depending on model configuration |
Downloads
| FCX80 Free Chlorine Analyzer Brochure | DownloadFCX80 Free Chlorine Analyzer Brochure |
|---|
Related Products

ECD
DCA-23 Seawater Dechlorination Analyzer
The DCA-23 monitors seawater chlorination and dechlorination from 0.00–2.00 ppm chlorine equivalent, featuring reagent-free measurement, flow-independent…
Learn More
ECD
DC80 Dechlorination Analyzer
The DC80 Dechlorination Analyzer uses a zero-shift measurement strategy, automatic flow control and pH compensation to monitor very low total chlorine levels…