WETEC
GHTU Humidity and Temperature Transducer thumbnail 1
GHTU Humidity and Temperature Transducer thumbnail 2

Senseca

GHTU Humidity and Temperature Transducer


Humidity and temperature transducer for wall, duct and open-air use.

  • Humidity 0.0…100.0 %RH and temperature -40.0…+120.0 °C
  • 4…20 mA standard output (0…1 V or 0…10 V optional), with a 4-digit LCD alternating humidity and temperature
  • ABS housing, IP65, with min/max memory and adjustable offset and slope
Quantity
1

Overview

The GHTU is a humidity and temperature transducer built with the latest microprocessor technology. It measures 0.0…100.0 %RH (temperature compensated) and -40.0…+120.0 °C, and comes in wall, wall/duct, duct, cable and open-air versions. It has a 4-digit LCD and standard signal outputs, in an ABS housing (IP65).

Typical applications

Different versions let the GHTU be used in almost all applications.

  • Wall mounting (GHTU-1R-MP, GHTU-1K-MP)
  • Duct (channel) mounting (GHTU-1K-MP, GHTU-2K-MP)
  • Open-air measurement with the heat absorption hat / weather protection shield (GHTU-SHUT-MP)
  • Wall mounting with a separated sensor tube on a 1 m cable and high-humidity sensor (GHTU-KABEL-MP)

Key Feature

Precise and flexible humidity and temperature measurement with a built-in display.

  • Temperature compensated humidity measurement, with automatic temperature compensation
  • 4…20 mA standard output; 0…1 V and 0…10 V options, including galvanically isolated versions
  • 4-digit LCD display (approx. 10 mm high) alternating humidity and temperature
  • Min/max value memory, offset and slope adjustable, output signal scaleable
  • Optional UNI display: show wet bulb temperature, dew point temperature, enthalpy, atmospheric humidity or absolute humidity instead of the humidity value
  • Optional high-humidity sensor; optional encapsulated PC board for use where condensation may occur
  • Fitting lengths of 50, 220, 300, 400 or 500 mm

Benefits

  • Flexible mounting: wall, duct, separated probe and weather shield versions support room, duct and open-air use
  • Wide sensor range: measures 0.0…100.0 %RH and sensor temperatures from -40.0…+120.0 °C
  • Process-ready output: 4-20 mA is standard; 0-1 V and 0-10 V options are available
  • Protected housing: ABS housing and EN 175301-803/A elbow-type plug are rated IP65
  • On-site adjustment: min/max memory, offset and slope adjustment, and scalable output support setup

Specifications

SPECIFICATIONS

 

Humidity measuring range

0.0…100.0 %RH (temperature compensated)

Temperature measuring range

-40.0…+120.0 °C or -40.0…+248 °F

Recommended humidity range

20.0…80.0 %RH (standard)
5.0…95.0 %RH (with option high humidity)

Display options (option UNI)

An alternative display unit can be shown instead of the humidity measuring value, selected via keyboard:
Wet bulb temperature: -27.0…+60.0 °C
Dewpoint temperature: -40.0…+60.0 °C
Enthalpy: -25.0…+999.9 kJ/kg
Atmospheric humidity: 0.0…640.0 g/kg
Absolute humidity: 0.0…200.0 g/m³

Accuracy (at 25 °C and in recommended range)

Display, humidity: ±2.5 %RH
Display, temperature: ±0.4 % of measuring value ±0.2 °C
Output signal: humidity ±0.2 % FS, temperature ±0.2 % FS

Temperature compensation

Automatically

Auxiliary energy

12…30 VDC or 18…30 VDC (for output: 0-10 V)

Reverse voltage protection

50 V, permanently

Permissible impedance (at 4-20 mA)

RA [Ω] ≤ (UV [V] - 12 V) / 0.02 A

Permissible load (at 0-1(10) V)

RL [Ω] > 3000 Ω

Display

Approx. 10 mm high, 4-digit LCD display, alternating humidity and temperature display

Working temperature

Electronics: -25…+50 °C
Sensor head and tube: -40…+100 °C, for short time up to 120 °C

Storage temperature

-25…+70 °C

Relative humidity (electronic)

0…95 %RH (non-condensing). If there is a risk of condensation due to temperature changes, use the encapsulated or lacquered types (optionally available).

 

 

DESIGN & MOUNTING

 

Models

GHTU-1R-MP: wall version, standard probe length 50 mm (Product-ID 602585)
GHTU-1K-MP: wall / channel version, standard probe length 220 mm (Product-ID 602587)
GHTU-2K-MP: channel version, standard probe length 220 mm (Product-ID 602592)
GHTU-SHUT-MP: absorption hat / weather protection (Product-ID 603896)
GHTU-KABEL-MP: wall version with cable and high humidity sensor (Product-ID 604436)

Housing

ABS (IP65)

Sensor tube

Tube 14 mm Ø, with screw-type protection cap

Design type Kabel

With separated sensor tube, sensor head (Ø14 x 68 mm) connected to device via 1 m Teflon cable. Includes option high-humidity sensor.

Design type Shut

Heat absorption hat / weather protection shield, especially designed for measurements in the open air. The measuring results are not influenced by either sun or rain.
Heat absorption hat made of plastic, Ø 110 mm, approx. 140 mm high, with a stainless steel base for wall mounting (3 fixing holes for screws with a max. shaft Ø of 5 mm). Large projection approx. 160 mm.

Electric connection

Elbow-type plug acc. to EN 175301-803/A (IP65)

Mounting

4 housing holes for wall mounting or by means of plastic tube clamps for duct mounting

Functions

Min-/max-value memory, offset and slope adjustable, output signal scaleable

 

 

ORDERING CODE

GHTU - 1 - 2 - 3 - 4 - 5 - 6 -GE

1. Design type

1R-MP = Surface design
1K-MP = Surface / duct design
2K-MP = Duct design
KABEL-MP = Surface design with cable and high humidity sensor
SHUT-MP = Weather protective shield / heat-protective hat incl. Option LACK and HO

2. Fitting length EL

-050 = 50 mm (standard for 1R)
-220 = 220 mm (standard for 1K / 2K)
-300 = 300 mm
-400 = 400 mm
-500 = 500 mm

3. Output signal

... = 4-20 mA (standard)
-AV01 = 0-1 V
-AV01G = 0-1 V (galvanically isolated)
-AV10 = 0-10 V
-AV10G = 0-10 V (galvanically isolated)

4. Options sensor

-00 = Standard sensor
-HO = High humidity sensor

5. Option

-LACK = Encapsulated PC board

6. Option

-UNI = Selectable humidity display instead of the standard humidity values

Downloads

GHTU Family Datasheet (EN) DownloadGHTU Family Datasheet (EN)