WETEC
PHIX Digital pH and ORP Transmitter thumbnail 1
PHIX Digital pH and ORP Transmitter thumbnail 2
PHIX Digital pH and ORP Transmitter thumbnail 3
PHIX Digital pH and ORP Transmitter thumbnail 4

Senseca

PHIX Digital pH and ORP Transmitter

WaterProcess & Automation

Digital pH/ORP converter for combined electrodes with S7/S8 or VarioPin connection heads.

- Measuring range: −1…+15 pH or −1500…+1500 mV

- RS485 interface with Modbus RTU protocol

- Temperature measurement −50…+200 °C with Pt1000-equipped electrodes

- Temperature-compensated glass impedance measurement from 0…1 GΩ

- Guided 1-point and 2-point pH calibration

- Storage of up to 5 calibration results with timestamps

- IP67 protection

Quantity
1

Overview

The Senseca PHIX is a digital converter designed for potentiometric measurement of pH or oxidation-reduction potential (ORP) using combined electrodes. It mounts directly onto S7/S8 or VarioPin electrode connection heads, eliminating the need for special electrode cables. The converter communicates via RS485 with Modbus RTU protocol, enabling integration into existing networks and MULTICON systems.

PHIX supports process temperature measurement with Pt1000-equipped electrodes, glass impedance monitoring, guided 1-point and 2-point pH calibration, and manual adjustment of zero point and slope. It stores up to five calibration results with timestamps and records total, temperature-related and pH-related operating times for sensor stress evaluation. Configuration and calibration are supported through SENSware and MULTICON-H. The converter features a PVC-U housing with a PMMA window, IP67 protection and a 4.7…28 V DC power supply.

Typical Applications

  • pH or ORP measurement in aqueous solutions

  • Process temperature monitoring with Pt1000-equipped electrodes

  • Digital integration of conventional electrodes into monitoring and control systems

  • Electrode condition assessment through glass impedance monitoring

  • Sensor maintenance evaluation using calibration history and operating time records

Key Features

  • Direct screw-on connection to combined electrodes with S7/S8 or VarioPin heads

  • Measuring range −1…+15 pH or −1500…+1500 mV

  • Temperature measurement −50…+200 °C with Pt1000-equipped electrodes

  • Temperature-compensated glass impedance measurement from 0…1 GΩ

  • RS485 half-duplex interface with Modbus RTU protocol and selectable baud rates from 1200 to 19200 baud

  • Guided 1-point and 2-point pH calibration with manual zero point and slope adjustment also available

  • Storage of up to 5 calibration results with timestamps

  • Recording of total, temperature-related and pH-related operating times

  • Configuration via SENSware, MULTICON-H or Modbus RTU

  • LED indication for power supply, bus communication and addressing status

  • 8-pin M12 electrical connector

  • PVC-U housing with PMMA window and IP67 protection

  • 4.7…28 V DC power supply

Benefits

  • Simplified installation without complex special electrode cables

  • Easy integration into existing digital monitoring and control systems

  • Reduced susceptibility to interference during signal transmission

  • Easier identification of electrode condition through diagnostic monitoring

  • Simplified maintenance through guided calibration procedures

  • Improved calibration history traceability through timestamped records

  • Better sensor stress evaluation through operating time records

  • Flexible device configuration through compatible software and controllers

  • Protection against dust and water ingress for the converter electronics

Specifications

Power Supply

Supply voltage

4.7…28 V DC

Current consumption

Max. 60 mA

Conformity

CE

 

 

Input – pH/ORP

Measuring principle

Potentiometric

Measuring range

−1…+15 pH / −1500…+1500 mV

Accuracy

±0.5% of measured value

Temperature coefficient

<100 ppm/°C

Zero point

pH = 7.00 (ΔpH = ±3.0)

Slope

30…80 mV/pH

ORP adjustment

±200 mV

 

 

Temperature

Temperature sensor

Pt1000, integrated into the combination electrode

Measuring range

−50…+200 °C

Accuracy

±0.5% of measured value

pH temperature compensation

Automatic using measured temperature or based on manually entered temperature

 

 

Glass Impedance

Measuring range

0…1 GΩ (temperature compensated)

Standard error

±20%

 

 

Environmental Conditions

Ambient/storage temperature

−10…+60 °C

Condensation

Not permitted

 

 

Electrode Connection

Electrode connector

S8 (for S7/S8 electrodes) or VP (VarioPin)

 

 

Materials

Housing

PVC-U

Window

Acrylic glass (PMMA)

 

 

Electrical Connection

Connector design

8-pole M12 × 1 round plug, IP67

Connector material

Nickel-plated brass

 

 

Communication Interface

Interface

RS485, half-duplex

Protocol

Modbus RTU

Baud rates

1200, 2400, 4800, 9600, 19200 baud

 

 

Mechanical

Total weight

Approx. 160 g

Protection class

IP67

 

 

Calibration

1-point calibration

Zero point correction

2-point calibration

Zero point and slope correction

Manual adjustment

Zero point and slope

Calibration memory

Up to 5 completed calibration results

Stored calibration data

Timestamp, zero point and slope

 

 

Diagnostics and Operating Time

Additional diagnostic parameters

Internal temperature and supply voltage

Operating time records

Total, temperature-related and pH-related operating times

Optical signalling

Power supply, bus communication and addressing status

 

 

Configuration

Configuration software

SENSware

Compatible controller

MULTICON-H

Programming adapter

EYY220

Drawing

PHIX Digital pH and ORP Transmitter Drawing

Downloads

PHIX Family Datasheet DownloadPHIX Family Datasheet
PHIX Family User Manual DownloadPHIX Family User Manual
Modbus-Sensors Family Interface Description DownloadModbus-Sensors Family Interface Description