Senseca
PHIX Digital pH and ORP Transmitter
Digital pH/ORP converter for combined electrodes with S7/S8 or VarioPin connection heads.
- Measuring range: −1…+15 pH or −1500…+1500 mV
- RS485 interface with Modbus RTU protocol
- Temperature measurement −50…+200 °C with Pt1000-equipped electrodes
- Temperature-compensated glass impedance measurement from 0…1 GΩ
- Guided 1-point and 2-point pH calibration
- Storage of up to 5 calibration results with timestamps
- IP67 protection
Overview
The Senseca PHIX is a digital converter designed for potentiometric measurement of pH or oxidation-reduction potential (ORP) using combined electrodes. It mounts directly onto S7/S8 or VarioPin electrode connection heads, eliminating the need for special electrode cables. The converter communicates via RS485 with Modbus RTU protocol, enabling integration into existing networks and MULTICON systems.
PHIX supports process temperature measurement with Pt1000-equipped electrodes, glass impedance monitoring, guided 1-point and 2-point pH calibration, and manual adjustment of zero point and slope. It stores up to five calibration results with timestamps and records total, temperature-related and pH-related operating times for sensor stress evaluation. Configuration and calibration are supported through SENSware and MULTICON-H. The converter features a PVC-U housing with a PMMA window, IP67 protection and a 4.7…28 V DC power supply.
Typical Applications
pH or ORP measurement in aqueous solutions
Process temperature monitoring with Pt1000-equipped electrodes
Digital integration of conventional electrodes into monitoring and control systems
Electrode condition assessment through glass impedance monitoring
Sensor maintenance evaluation using calibration history and operating time records
Key Features
Direct screw-on connection to combined electrodes with S7/S8 or VarioPin heads
Measuring range −1…+15 pH or −1500…+1500 mV
Temperature measurement −50…+200 °C with Pt1000-equipped electrodes
Temperature-compensated glass impedance measurement from 0…1 GΩ
RS485 half-duplex interface with Modbus RTU protocol and selectable baud rates from 1200 to 19200 baud
Guided 1-point and 2-point pH calibration with manual zero point and slope adjustment also available
Storage of up to 5 calibration results with timestamps
Recording of total, temperature-related and pH-related operating times
Configuration via SENSware, MULTICON-H or Modbus RTU
LED indication for power supply, bus communication and addressing status
8-pin M12 electrical connector
PVC-U housing with PMMA window and IP67 protection
4.7…28 V DC power supply
Benefits
Simplified installation without complex special electrode cables
Easy integration into existing digital monitoring and control systems
Reduced susceptibility to interference during signal transmission
Easier identification of electrode condition through diagnostic monitoring
Simplified maintenance through guided calibration procedures
Improved calibration history traceability through timestamped records
Better sensor stress evaluation through operating time records
Flexible device configuration through compatible software and controllers
Protection against dust and water ingress for the converter electronics
Specifications
Power Supply | |
|---|---|
Supply voltage | 4.7…28 V DC |
Current consumption | Max. 60 mA |
Conformity | CE |
Input – pH/ORP | |
|---|---|
Measuring principle | Potentiometric |
Measuring range | −1…+15 pH / −1500…+1500 mV |
Accuracy | ±0.5% of measured value |
Temperature coefficient | <100 ppm/°C |
Zero point | pH = 7.00 (ΔpH = ±3.0) |
Slope | 30…80 mV/pH |
ORP adjustment | ±200 mV |
Temperature | |
|---|---|
Temperature sensor | Pt1000, integrated into the combination electrode |
Measuring range | −50…+200 °C |
Accuracy | ±0.5% of measured value |
pH temperature compensation | Automatic using measured temperature or based on manually entered temperature |
Glass Impedance | |
|---|---|
Measuring range | 0…1 GΩ (temperature compensated) |
Standard error | ±20% |
Environmental Conditions | |
|---|---|
Ambient/storage temperature | −10…+60 °C |
Condensation | Not permitted |
Electrode Connection | |
|---|---|
Electrode connector | S8 (for S7/S8 electrodes) or VP (VarioPin) |
Materials | |
|---|---|
Housing | PVC-U |
Window | Acrylic glass (PMMA) |
Electrical Connection | |
|---|---|
Connector design | 8-pole M12 × 1 round plug, IP67 |
Connector material | Nickel-plated brass |
Communication Interface | |
|---|---|
Interface | RS485, half-duplex |
Protocol | Modbus RTU |
Baud rates | 1200, 2400, 4800, 9600, 19200 baud |
Mechanical | |
|---|---|
Total weight | Approx. 160 g |
Protection class | IP67 |
Calibration | |
|---|---|
1-point calibration | Zero point correction |
2-point calibration | Zero point and slope correction |
Manual adjustment | Zero point and slope |
Calibration memory | Up to 5 completed calibration results |
Stored calibration data | Timestamp, zero point and slope |
Diagnostics and Operating Time | |
|---|---|
Additional diagnostic parameters | Internal temperature and supply voltage |
Operating time records | Total, temperature-related and pH-related operating times |
Optical signalling | Power supply, bus communication and addressing status |
Configuration | |
|---|---|
Configuration software | SENSware |
Compatible controller | MULTICON-H |
Programming adapter | EYY220 |
Drawing

Downloads
| PHIX Family Datasheet | DownloadPHIX Family Datasheet |
|---|---|
| PHIX Family User Manual | DownloadPHIX Family User Manual |
| Modbus-Sensors Family Interface Description | DownloadModbus-Sensors Family Interface Description |
Related Products
Senseca
UNICON-LF-01/02/03 Conductivity Converter
Conductivity converter for liquid purity and concentration monitoring.
Learn More
Senseca
LF4003 Conductivity Cell for Wells and Open Water Systems
4-electrode immersion conductivity cell for wells and open systems up to 100 m depth.
Learn More
Senseca
ORP211 Combined Redox Electrode for Polluted Wastewater Media
Combined Redox electrode for heavily polluted media in wastewater and process chemistry.
Learn MoreSenseca
UNICON-PH-01/02 pH and ORP Converter
pH and ORP converter for process monitoring with programmable ranges of −1.00 to +15.00 pH or ±1500 mV, Pt100/Pt1000 temperature compensation, 4–20 mA output…
Learn More