Senseca
UNICON-LF-01/02/03 Conductivity Converter
Conductivity converter for liquid purity and concentration monitoring with 4-electrode cells or special 2-electrode cells for ultra-pure water.
Programmable measuring ranges from 0…5.00 μS/cm to 0…500.0 mS/cm
Ultra-pure water measuring ranges from 0…0.500 μS/cm to 0…50.0 μS/cm
Selectable linear and non-linear temperature compensation
4…20 mA conductivity output and two electronic alarm outputs
Two-line LCD display, IP65 housing and head or field mounting
Overview
The Senseca UNICON-LF is a conductivity converter designed to measure the conductivity of liquids as an indicator of purity or concentration. It supports 4-electrode conductivity cells for high accuracy and reduced sensitivity to contamination, as well as special 2-electrode cells for ultra-pure water. Programmable measuring ranges extend from 0…5.00 μS/cm to 0…500.0 mS/cm, or from 0…0.500 μS/cm to 0…50.0 μS/cm when using an ultra-pure water cell.
The converter provides selectable linear and non-linear temperature compensation using Pt100 or Pt1000 sensors, a 4…20 mA conductivity output and two electronic alarm outputs. Depending on the model, a second conductivity measuring range and a 4…20 mA temperature output are available. It features a two-line LCD dot-matrix display, front keypad configuration and an IP65 housing for head or field mounting. An optional USP<645> function supports ultra-pure water measurement and monitoring.
Typical Applications
Conductivity measurement of groundwater, spring water and surface water
Ultra-pure water monitoring in pharmaceutical applications with optional USP<645> functionality
Conductivity monitoring of saline, dilute acid, alkaline and cleaning solutions
Analogue signal integration with process control and monitoring systems
Key Features
Compatible with 4-electrode conductivity cells and special 2-electrode cells for ultra-pure water
Programmable measuring ranges from 0…5.00 μS/cm to 0…500.0 mS/cm
Ultra-pure water measuring ranges from 0…0.500 μS/cm to 0…50.0 μS/cm
Conductivity measurement accuracy of ±0.5% of measured value ±2 digits
Linear and non-linear temperature compensation with Pt100 or Pt1000 sensors and a programmable linear coefficient of 0.000…8.000 %/°C
4…20 mA conductivity output
Two programmable electronic alarm outputs
Model-dependent second conductivity measuring range and 4…20 mA temperature output
Two-line LCD dot-matrix display with 16 characters per line
Front keypad configuration with program lockout and EEPROM write protection
14…30 V DC loop power supply
IP65 glass fibre-reinforced polyamide housing
Head or field mounting versions
Optional USP<645> measurement and monitoring functionality
Benefits
High measurement accuracy with reduced sensitivity to contamination when using 4-electrode cells
Flexible measuring ranges to suit different liquid purity and concentration requirements
Reliable conductivity readings under varying liquid temperatures through selectable temperature compensation
Easy integration with industrial control and monitoring systems
Automatic alarm signalling for readings above or below configured limits
Convenient local configuration and measurement display
Flexible installation with head or field mounting options
Protection against dust and water jets through the IP65 housing
Protection against unintended parameter changes through program lockout and EEPROM write protection
Support for pharmaceutical ultra-pure water monitoring with optional USP<645> functionality
Specifications
Power Supply | |
|---|---|
Loop voltage | 14…30 V DC |
Connection system | 2-wire |
Operating temperature | 0…+50 °C (datasheet); −15…+55 °C (operating manual) |
Electrical isolation | Conductivity output, temperature output, alarm outputs 1 and 2, measuring range switching |
Test voltage | 500 V DC |
Conductivity Measurement | |
|---|---|
Measuring principle | 4-electrode measurement; special 2-electrode cell for ultra-pure water |
Lowest / highest measuring range | 0…5.00 μS/cm / 0…500.0 mS/cm |
Lowest / highest measuring range with ultra-pure water cell | 0…0.500 μS/cm / 0…50.0 μS/cm |
Measuring units | μS/cm, mS/cm, kΩ/cm, MΩ/cm, programmable (as listed in the manufacturer's technical data) |
Decimal positions | 0…3, depending on measuring unit |
Indicating range | 500…9999 digits, depending on unit and decimal positions |
Temperature compensation | Non-linear for natural and ultra-pure water, or programmable linear compensation |
Linear compensation coefficient | 0.000…8.000 %/°C |
Cell constant | 0.080…9.999 (datasheet); 0.080…4.000 (operating manual) |
Accuracy | ±0.5% of measured value ±2 digits |
Instrument temperature coefficient | <100 ppm/°C |
Measuring rate | Approx. 3 measurements/s |
Current output | 4…20 mA |
Temperature Measurement | |
|---|---|
Temperature sensor | RTD Pt100 or Pt1000 according to DIN IEC 751 |
Measuring units | °C or °F, programmable |
Measuring range | −40.0…+160.0 °C (−40.0…+320.0 °F) |
Minimum measuring span | 25.0 °C (45.0 °F) |
Maximum measuring span | 200 °C (360 °F) |
Standard error | ±0.1% ±1 digit |
Temperature coefficient | <50 ppm/°C |
Linearisation error | ±0.1% |
Current output | 4…20 mA (Model 2) |
Load resistance | Rₐ ≤ (Uᵦ − 14 V) / 0.02 A |
Alarm Outputs | |
|---|---|
Number of alarm outputs | 2 |
Output type | Voltage-free transistor outputs |
Transistor voltage | 14…30 V DC |
Maximum current | 60 mA |
Voltage drop | <2 V |
Output protection | Short-circuit protection |
Switching function | Programmable minimum or maximum function |
Alarm assignment | Conductivity measuring range 1, conductivity measuring range 2 or temperature, depending on model and configuration |
Alarm setpoint | Programmable |
Alarm hysteresis | Programmable |
Measuring Range Switching (Model 2) | |
|---|---|
Switching method | External signal |
Input resistance | >10 kΩ |
Measuring range 1 active | 0…3 V DC |
Measuring range 2 active | 12…30 V DC |
Display and Configuration | |
|---|---|
Display type | LCD dot-matrix |
Display format | 2 lines, 16 characters per line |
Character height | 3.8 mm |
Displayed information | Conductivity, temperature and alarm output status |
Configuration method | Front membrane keypad |
Operating language | German or English |
Digital filter | Continuous averaging of the last 16 measured values when activated |
Simulation mode | Conductivity and temperature |
Configuration protection | Program lockout and EEPROM write protection |
Housing and Mechanical | |
|---|---|
Housing material | Glass fibre-reinforced polyamide |
Material designation | PA6-GF/GK 15/15 (datasheet); PA6-GF 15/15 (operating manual) |
Front foil material | Polyester |
Dimensions (W × H × D) | 100 × 100 × 60 mm |
Weight | Max. 360 g |
Housing protection | IP65 |
Terminal protection | IP20 according to BGV A3 |
Electrical connection | Screw terminals with pressure plate |
Flexible wire cross-section | 2.5 mm² |
Single wire cross-section | 4 mm² |
Cable gland | M20 × 1.5 |
Cable diameter | 6…12 mm |
Mounting Versions code | |
|---|---|
01 | Head mounting directly on the conductivity cell |
02 | Field mounting with separate connection cable |
03 | Field mounting with separate connection cable and stainless steel plug (1.4571) |
Optional USP<645> Functionality | |
|---|---|
Option code | 14 |
Function | Ultra-pure water measurement and monitoring according to USP<645> |
Monitoring stage | Stage 1 |
Temperature compensation during USP measurement | None |
Alarm contact function | Normally open (NO) or normally closed (NC) |
Alarm setpoint adjustment | 50…100% of the permitted limit value |
Alarm hysteresis | Fixed at 0.10 μS/cm |
Calibration | Dedicated USP calibration routine for the complete measuring system |
Standards and Conformity | |
|---|---|
EMC Directive | 2014/30/EU |
RoHS Directive | 2011/65/EU |
EMC standard | EN IEC 61326-1:2013 |
RoHS standard | EN IEC 63000:2018 |
Drawing
Downloads
| UNICON-LF Family Datasheet | DownloadUNICON-LF Family Datasheet |
|---|---|
| UNICON-LF Family User Manual | DownloadUNICON-LF Family User Manual |
| UNICON-LF Family CE Declaration of Conformity | DownloadUNICON-LF Family CE Declaration of Conformity |
Related Products

Senseca
LF4003 Conductivity Cell for Wells and Open Water Systems
4-electrode immersion conductivity cell for wells and open systems up to 100 m depth.
Learn More
Senseca
ORP211 Combined Redox Electrode for Polluted Wastewater Media
Combined Redox electrode for heavily polluted media in wastewater and process chemistry.
Learn MoreSenseca
UNICON-PH-01/02 pH and ORP Converter
pH and ORP converter for process monitoring with programmable ranges of −1.00 to +15.00 pH or ±1500 mV, Pt100/Pt1000 temperature compensation, 4–20 mA output…
Learn More
Senseca
CONDIX4213 Digital Conductivity Transmitter
Digital conductivity converter for water treatment and industrial liquid monitoring.
Learn More