ECD
S80 Dissolved Oxygen Sensors
Intelligent digital dissolved oxygen sensor with replaceable galvanic electrode cartridges.
Measurement range: 0–20 ppm (mg/L) and 0–250% saturation
Automatic temperature compensation
Two-way digital communication with onboard sensor identity and calibration data storage
Insertion/submersion or valve retractable designs
Standard 316 stainless steel housing with optional construction materials
Overview
The ECD S80 Dissolved Oxygen Sensor is an intelligent digital sensor designed for real-time dissolved oxygen measurement over a range of 0–20 ppm (mg/L) and 0–250% saturation. It features replaceable galvanic electrode cartridges and automatic temperature compensation. The sensor internally stores its identity, measurement type, serial number and calibration data, supporting two-way digital communication and automatic configuration with compatible ECD transmitters.
The S80 is available in insertion/submersion and valve retractable designs, with a standard 316 stainless steel housing and optional construction materials. The insertion version uses a ¾″ MNPT compression fitting with adjustable insertion depth, while the valve retractable version uses a 1″ ball valve with a 1″ NPT process connection, allowing sensor removal for maintenance without shutting down the process line.
Typical Applications
Dissolved oxygen concentration and percentage saturation measurement
In-line dissolved oxygen monitoring in pipelines and flow cells
Submersible dissolved oxygen monitoring in tanks
Dissolved oxygen measurement for industrial process control systems
Key Features
Dissolved oxygen measurement range of 0–20 ppm (mg/L)
Oxygen saturation measurement range of 0–250%
Replaceable galvanic electrode cartridges with PEEK bodies
Automatic temperature compensation
Two-way digital communication with compatible ECD transmitters and controllers
Built-in memory for sensor identity, measurement type, serial number and calibration data
Factory-calibrated sensor for plug-and-play installation
Insertion/submersion and valve retractable configurations
Standard 316 stainless steel housing with optional construction materials
Adjustable insertion depth with compression fittings
Flared sensor end on the valve retractable design to prevent blowout
Standard 10 ft cable with optional cable lengths and detachable cable connection
Benefits
Simplified installation and setup through factory calibration and automatic transmitter configuration
Reduced operating costs through replaceable electrode cartridges
Flexible mounting options to accommodate different process installation requirements
Reduced process downtime by allowing sensor removal without shutting down the process line with the valve retractable design
Improved compatibility with process liquids through a choice of housing materials
Easier sensor replacement with the optional detachable cable connection
Specifications
Dissolved Oxygen Measurement | |
|---|---|
Measurement principle | Galvanic electrode |
Dissolved oxygen measurement range | 0–20 ppm (mg/L) |
Oxygen saturation range | 0–250% |
Pressure range | 0–50 psig at 80 °C |
Temperature compensation | Automatic, 0–100 °C |
Temperature accuracy | ±0.2 °C |
Electrode Cartridges | |
|---|---|
Electrode design | Replaceable cartridge |
Electrode body material | PEEK |
Membrane thickness options | 2 mil and 5 mil |
Two-tang electrode | 2005622 (Ordering code: 040) |
Full-crown electrode | 2005622.FC (Ordering code: 041) |
Thick-membrane electrode | 2005623 (Ordering code: 042) |
Digital Communication and Calibration | |
|---|---|
Communication | Two-way digital communication |
Compatible transmitter | ECD T80 Transmitter |
Compatible controller | ECD LQ800 Controller |
Stored sensor information | Sensor identity, measurement type, serial number and calibration data |
Transmitter configuration | Automatic sensor recognition and configuration |
Calibration | Factory calibrated; periodic calibration required |
Mechanical | |
|---|---|
Sensor configurations | Insertion/submersion and valve retractable |
Standard insertion sensor dimensions | ¾″ OD × 10″ length |
Standard valve retractable sensor dimensions | ¾″ OD × 17″ length |
Standard housing material | 316 stainless steel |
Optional metallic housing materials | Titanium and Hastelloy C-22 |
Polypropylene housing options | 1″ OD × 10″ or 17″ length, with stainless steel solution ground |
PVDF (Kynar) housing option | 1″ OD × 10″ length |
Standard O-ring material | Viton |
Optional O-ring materials | Ethylene propylene, Viton 75, Kalrez and CV75 |
Valve retractable safety design | Flanged blowout protector |
Process Connections | |
|---|---|
Standard insertion process connection | ¾″ MNPT compression fitting |
Insertion depth | Adjustable |
Stainless steel gland fitting options | Nylon, Teflon or stainless steel ferrule |
Other ¾″ MNPT gland fitting materials | Hastelloy, titanium, polypropylene, PVDF and Teflon |
1″ MNPT gland fitting materials | Nylon, polypropylene and PVDF |
Valve retractable process connection | 1″ ball valve with 1″ NPT process connection |
Valve retraction assembly material options | 316 stainless steel, Hastelloy, titanium, PVDF and polypropylene |
Cable and Connector | |
|---|---|
Standard cable length | 10 ft |
Optional cable lengths | 20, 30, 40, 50 and 100 ft; other lengths available upon consultation with the manufacturer |
Standard cable configuration | Integral cable |
Optional cable configuration | Detachable cable |
Detachable connector type | Straight (axial) |
Detachable connector protection rating | IP68 |
Downloads
| S80 - Sensors Manual | DownloadS80 - Sensors Manual |
|---|
Related Products

ECD
T80 Universal Transmitter
The ECD T80 Universal Transmitter provides continuous industrial process monitoring with single- or dual-channel operation, automatic S80 sensor recognition…
Learn More
ECD
S88 Hazardous Location Sensors
The S88 pH Sensors offer digital pH measurement from 0–14 pH, built-in calibration data storage, replaceable electrode cartridges, and insertion, submersion…
Learn More
ECD
S80 Ion Selective Sensors
The S80 Ion Selective Sensors measure specific ions in solution, featuring digital communication, onboard calibration storage, replaceable electrodes, and…
Learn More
ECD
S80 Inductive Conductivity Sensors
The S80 Inductive Conductivity Sensors feature non-contact conductivity measurement, chemical-resistant PVDF construction, intelligent digital communication…
Learn More