WETEC

ECD

S80 Dissolved Oxygen Sensors


Intelligent digital dissolved oxygen sensor with replaceable galvanic electrode cartridges.

  • Measurement range: 0–20 ppm (mg/L) and 0–250% saturation

  • Automatic temperature compensation

  • Two-way digital communication with onboard sensor identity and calibration data storage

  • Insertion/submersion or valve retractable designs

  • Standard 316 stainless steel housing with optional construction materials

Quantity
1

Overview

The ECD S80 Dissolved Oxygen Sensor is an intelligent digital sensor designed for real-time dissolved oxygen measurement over a range of 0–20 ppm (mg/L) and 0–250% saturation. It features replaceable galvanic electrode cartridges and automatic temperature compensation. The sensor internally stores its identity, measurement type, serial number and calibration data, supporting two-way digital communication and automatic configuration with compatible ECD transmitters.

The S80 is available in insertion/submersion and valve retractable designs, with a standard 316 stainless steel housing and optional construction materials. The insertion version uses a ¾″ MNPT compression fitting with adjustable insertion depth, while the valve retractable version uses a 1″ ball valve with a 1″ NPT process connection, allowing sensor removal for maintenance without shutting down the process line.

Typical Applications

  • Dissolved oxygen concentration and percentage saturation measurement

  • In-line dissolved oxygen monitoring in pipelines and flow cells

  • Submersible dissolved oxygen monitoring in tanks

  • Dissolved oxygen measurement for industrial process control systems

Key Features

  • Dissolved oxygen measurement range of 0–20 ppm (mg/L)

  • Oxygen saturation measurement range of 0–250%

  • Replaceable galvanic electrode cartridges with PEEK bodies

  • Automatic temperature compensation

  • Two-way digital communication with compatible ECD transmitters and controllers

  • Built-in memory for sensor identity, measurement type, serial number and calibration data

  • Factory-calibrated sensor for plug-and-play installation

  • Insertion/submersion and valve retractable configurations

  • Standard 316 stainless steel housing with optional construction materials

  • Adjustable insertion depth with compression fittings

  • Flared sensor end on the valve retractable design to prevent blowout

  • Standard 10 ft cable with optional cable lengths and detachable cable connection

Benefits

  • Simplified installation and setup through factory calibration and automatic transmitter configuration

  • Reduced operating costs through replaceable electrode cartridges

  • Flexible mounting options to accommodate different process installation requirements

  • Reduced process downtime by allowing sensor removal without shutting down the process line with the valve retractable design

  • Improved compatibility with process liquids through a choice of housing materials

  • Easier sensor replacement with the optional detachable cable connection

Specifications

Dissolved Oxygen Measurement

Measurement principle

Galvanic electrode

Dissolved oxygen measurement range

0–20 ppm (mg/L)

Oxygen saturation range

0–250%

Pressure range

0–50 psig at 80 °C

Temperature compensation

Automatic, 0–100 °C

Temperature accuracy

±0.2 °C

 

 

Electrode Cartridges

Electrode design

Replaceable cartridge

Electrode body material

PEEK

Membrane thickness options

2 mil and 5 mil

Two-tang electrode

2005622 (Ordering code: 040)

Full-crown electrode

2005622.FC (Ordering code: 041)

Thick-membrane electrode

2005623 (Ordering code: 042)

 

 

Digital Communication and Calibration

Communication

Two-way digital communication

Compatible transmitter

ECD T80 Transmitter

Compatible controller

ECD LQ800 Controller

Stored sensor information

Sensor identity, measurement type, serial number and calibration data

Transmitter configuration

Automatic sensor recognition and configuration

Calibration

Factory calibrated; periodic calibration required

 

 

Mechanical

Sensor configurations

Insertion/submersion and valve retractable

Standard insertion sensor dimensions

¾″ OD × 10″ length

Standard valve retractable sensor dimensions

¾″ OD × 17″ length

Standard housing material

316 stainless steel

Optional metallic housing materials

Titanium and Hastelloy C-22

Polypropylene housing options

1″ OD × 10″ or 17″ length, with stainless steel solution ground

PVDF (Kynar) housing option

1″ OD × 10″ length

Standard O-ring material

Viton

Optional O-ring materials

Ethylene propylene, Viton 75, Kalrez and CV75

Valve retractable safety design

Flanged blowout protector

 

 

Process Connections

Standard insertion process connection

¾″ MNPT compression fitting

Insertion depth

Adjustable

Stainless steel gland fitting options

Nylon, Teflon or stainless steel ferrule

Other ¾″ MNPT gland fitting materials

Hastelloy, titanium, polypropylene, PVDF and Teflon

1″ MNPT gland fitting materials

Nylon, polypropylene and PVDF

Valve retractable process connection

1″ ball valve with 1″ NPT process connection

Valve retraction assembly material options

316 stainless steel, Hastelloy, titanium, PVDF and polypropylene

 

 

Cable and Connector

Standard cable length

10 ft

Optional cable lengths

20, 30, 40, 50 and 100 ft; other lengths available upon consultation with the manufacturer

Standard cable configuration

Integral cable

Optional cable configuration

Detachable cable

Detachable connector type

Straight (axial)

Detachable connector protection rating

IP68

Downloads

S80 - Sensors Manual DownloadS80 - Sensors Manual