ECD
Triton DO90 ppb Dissolved Oxygen Analyzer
Trace dissolved oxygen analyzer for continuous monitoring in aqueous systems, particularly boiler feedwater.
Measuring range: 0.001…20.00 mg/L with ppm/ppb auto-ranging
Lead-silver galvanic measuring principle
Process temperature range: −5…+50 °C
Replaceable electrode cartridge
Digital signal output
Overview
The Triton® DO90 is an analyzer designed for continuous measurement of trace dissolved oxygen levels in aqueous systems, particularly boiler feedwater. It uses a lead-silver galvanic sensor with a durable PFA Teflon® membrane, covering a measuring range of 0.001…20.00 mg/L with ppm/ppb auto-ranging and a process temperature range of −5…+50 °C.
The sensor incorporates internal signal conditioning, calibration and diagnostic functions, providing a digital signal to the Model T80 Transmitter. Its replaceable electrode cartridge simplifies maintenance by eliminating the need for electrolyte and membrane replacement kits. The analyzer is available as separate components, a complete panel-mounted system or a kit for installation on an existing water panel.
Typical Applications
Trace dissolved oxygen measurement in aqueous systems
Boiler feedwater monitoring
Steam cycle monitoring to support corrosion prevention
Key Features
Lead-silver galvanic sensor with a PFA Teflon® membrane
Measuring range of 0.001…20.00 mg/L with ppm/ppb auto-ranging
Process temperature range of −5…+50 °C
Replaceable electrode cartridge
Digital signal transmission via a shielded 4-wire cable
100 kΩ NTC temperature sensor with compensation from 0…50 °C
Internal signal conditioning and diagnostic functions
Air and zero calibration
AISI 316 Ti stainless steel sensor body
Response time of t90: 30 seconds and t99: 90 seconds
Compatible with the Model T80 Transmitter
Available as separate components, a complete panel-mounted system or a kit without the panel
Benefits
Wide measurement coverage for trace and higher dissolved oxygen concentrations
Long-term measurement stability with minimal drift
Continuous in-process monitoring eliminates grab sampling errors
Simplified maintenance without the need for electrolyte or membrane replacement kits
Easy calibration using air and zero calibration methods
Noise-free digital signal transmission
Flexible installation to accommodate new or existing water panels
Specifications
Measurement | |
|---|---|
Measured parameter | Dissolved oxygen |
Measuring principle | Galvanic, silver cathode, lead anode, self-polarizing |
Measurement signal | Current signal proportional to oxygen partial pressure |
Measuring range | 0.001…20.00 mg/L or ppm/ppb auto-ranging |
Slope | Approximately 1.5 nA/ppb O₂ at 20 °C and 1013 hPa |
Response time | t90: 30 s; t99: 90 s |
Polarization time | <60 minutes |
Drift | <1% per month under continuous polarization |
Process Operating Conditions | |
|---|---|
Process temperature range | −5…+50 °C |
Process pressure range | 0…15 psi maximum overpressure |
Vacuum operation | Not permitted |
Minimum flow rate | 200 mL/min with flow cell (P/N 1000008-1) |
Temperature Compensation | |
|---|---|
Temperature sensor | NTC, 100 kΩ |
Temperature compensation range | 0…50 °C |
Calibration | |
|---|---|
Calibration methods | Air calibration and zero calibration |
Calibration storage | Stored in the sensor |
Electrical Connection and Signal | |
|---|---|
Signal output | Digital signal |
Electrical connection | Shielded 4-wire cable |
Maximum total cable length | 100 m cable extension |
Compatible transmitter | Model T80 Transmitter |
Materials of Construction | |
|---|---|
Sensor body | Flanged AISI 316 Ti stainless steel |
Electrode body | PEEK |
Membrane | PFA Teflon® |
Cathode | Silver |
Anode | Lead |
Membrane thickness | Approximately 2 mil |
Flow cell wetted parts | 316L stainless steel |
Process Connection | |
|---|---|
Tube fittings | ¼″ 316 stainless steel |
Flow cell mounting | Wall mounting using clamp-style holders or panel mounting using two 10 × 32 threaded ports |
System Configuration | |
|---|---|
Available configurations | Separate components, complete panel-mounted system or kit without panel |
Panel dimensions | 17″ × 12″ |
Power supply options (panel-mounted system) | Loop powered or 110/220 VAC |
Shipping Weight | |
|---|---|
Model DO90 sensor with 10 ft cable | 0.7 kg |
Stainless steel flow cell | 2.0 kg |
Drawing

Downloads
| DO90 – Trace Dissolved Oxygen Sensor Brochure | DownloadDO90 – Trace Dissolved Oxygen Sensor Brochure |
|---|
Related Products

ECD
DCA-23 Seawater Dechlorination Analyzer
The DCA-23 monitors seawater chlorination and dechlorination from 0.00–2.00 ppm chlorine equivalent, featuring reagent-free measurement, flow-independent…
Learn More
ECD
DC80 Dechlorination Analyzer
The DC80 Dechlorination Analyzer uses a zero-shift measurement strategy, automatic flow control and pH compensation to monitor very low total chlorine levels…
Learn More
ECD
CD80 Chlorine Dioxide Analyzer
The CD80 is a reagentless, panel-mounted analyzer for continuous chlorine dioxide (ClO₂) monitoring from 0.05 to 20 ppm, featuring automatic flow control, a…
Learn More
ECD
TC80 Total Chlorine Analyzer
The TC80 is a reagentless, panel-mounted analyzer for total chlorine monitoring in water, featuring amperometric sensing, automatic pH compensation, automatic…
Learn More