Senseca
UB1 Paddle Flow Switch for Water Lines DN 25 to DN 200
Paddle flow switch for water lines DN 25 to DN 200
Monitors flow in water lines from DN 25 to DN 200 with a paddle supported by a metal bellows that actuates a microswitch (changeover contact).
Transparent cover for visual function checks, and a switching point that is adjusted on site by moving the microswitch.
PN 16, medium temperature -20...+140 °C (no superheated steam), IP 65, DIN 43650-A plug, R 1" male thread or DN 32 flange, in brass or stainless steel construction.
Overview
The Senseca UB1 is a paddle flow switch for water lines from DN 25 to DN 200. A paddle supported by a metal bellows actuates a microswitch, giving plant operators a clear switching function with a switching point that can be verified and adjusted on site.
The transparent cover allows a visual function check. The device is rated for PN 16 and medium temperatures from -20...+140 °C (no superheated steam), is available with an R 1" external thread or a flange connection, and has IP 65 ingress protection.
Typical applications
Flow monitoring in water lines from DN 25 to DN 200
Flow assurance in water pipework where the switching point must be checked and adjusted on site
Installation around industrial pipework
Media containing ferritic particles
Key Features
Paddle supported by a metal bellows that actuates a microswitch
For nominal widths DN 25...200
Microswitch with gold-plated contacts for small currents and silver-plated contacts for larger currents
Visual function control through the transparent cover
Suitable for media with ferritic particles
Adjustable switching point: open the cover, loosen the microswitch fixing screw, move the switch and tighten the screw again
Plug DIN 43650-A / ISO 4400, or optionally a 4-pole M12x1 connector (max. 4 A)
R 1" male thread, or flange DIN 2527 DN 32 PN 16 with sealing surface DIN 2526 form C
Brass or stainless steel construction, with PC and PA non-medium-contact parts
Options: signal lamp, double contact, aluminium hood with IP 67, opaque cover, switching ranges for oil, special values, TÜV certification
Benefits
Wide pipe coverage: one family covers DN 25...200 water lines
Visible function check: the transparent cover lets users check switching behavior directly at the device
Industrial conditions: rated for PN 16 bar and medium temperatures from -20...+140 °C, excluding superheated steam
Contact flexibility: gold and silver microswitch coatings support small and larger currents up to 250 V AC/DC
Protected housing: ingress protection IP 65 supports installation around industrial pipework
Specifications
INPUT |
|
Switch | Micro switch |
Measuring principle | Paddle supported by a metal bellows, triggering a microswitch |
Nominal width | DN 25...200 (DN 125...200 on request) |
Switching range | 1.2...34 m³/h |
Qmax. recommended | Up to 75 m³/h |
Media | Water (oils and aggressive media available on request) |
MEASURING ACCURACY |
|
Tolerance | ±15 % of full scale value |
OUTPUTS |
|
Wiring | Changeover (no. 0.371) |
Switching voltage / current, 250 V AC/DC | Max. 6 A ohmic; 1.5 A inductive |
Switching voltage / current, 125 V AC/DC | Max. 6 A ohmic; 2 A inductive |
Switching voltage / current, 24 V DC | Max. 6 A ohmic; 5 A inductive |
Switching voltage / current, 12 V DC | Max. 6 A ohmic; 6 A inductive |
Minimum load | 4 V, 1 mA |
Contacts | Gold-plated for small currents, silver-plated for larger currents |
Load note | Electrical details apply to ohmic loads; capacitive and inductive loads must be operated using a protective circuit |
ENVIRONMENTAL |
|
Pressure resistance | PN 16 bar |
Medium temperature | -20...+140 °C (no superheated steam) |
Ambient temperature | -20...+70 °C |
Protection class | 2 - safety insulation |
Ingress protection | IP 65 |
MATERIALS |
|
Medium-contact, brass construction | CW614N nickelled, 1.4305, 1.4310, 1.4541, FVMQ |
Medium-contact, stainless steel construction | 1.4305, 1.4310, 1.4541, FKM |
Non-medium-contact materials | PC, PA |
MECHANICAL |
|
Process connection | Male thread R 1" (DIN 2999), or installation flange DIN 2527 DN 32 PN 16, sealing surface as per DIN 2526 form C |
Electrical connection | Plug DIN 43650-A / ISO 4400, optionally for round plug connector M12x1, 4-pole (max. 4 A) |
Cover | Transparent cover for visual function control |
Adjusting the switching point | Open the cover, loosen the screw slightly on the micro switch fixing, push the switch into the desired position and retighten the screw |
Weight | Threaded type 1.3 kg; flanged type 2.5 kg |
Installation location | Standard: horizontal flow; switching unit not recommended underneath; other installation positions are possible, but the installation position affects the switching point and range |
Installation notes | Straight calming section of 10 x DN in inlet and outlet; install a filter if the media are dirty; paddle fixing is unsecured, so fit a bolted fixing for critical conditions (e.g. vibration); the flange seal is not included in the scope of delivery |
OPTIONS |
|
Process connection | Threaded connection DN 25 - R 1"; flange DN 32 |
Connection material | Brass; stainless steel |
Options | Signal lamp red or red/green in the plug DIN 43650-A; double contact; aluminium hood with IP 67; opaque cover; switching ranges for oil; special values; TÜV certification |
Ordering information | Specify direction of flow, medium and switching range; for oils, state viscosity, temperature and designation (e.g. ISO VG 68) |
Drawing

Downloads
| Senseca-UB1-Datasheet-EN | DownloadSenseca-UB1-Datasheet-EN |
|---|
Related Products

Senseca
UR1-G Flow Switch for Water with Internal Thread
Compact paddle flow switch for water lines with reed switch, DN 10.50 internal threads G 3/8.G 2, PN 25 and medium temperatures up to 110 °C.
Learn More
Senseca
TMS500 Thermal Flow Meter for Compressed Air and Gases
Thermal flow meter for compressed air and gases, measuring flow, consumption and velocity in pipes up to DN 300, with 4.20 mA, RS 485 Modbus RTU and pulse…
Learn More
Senseca
EFK2 Flow Switch for Liquid Media Monitoring
Flow switch for liquid media monitoring, measuring water flow velocity at 2.150 or 3.300 cm/s with calorimetric sensing, adjustable switching point, LED…
Learn More
Senseca
RRI-032 Paddle Wheel Flow Transmitter for DN 32 Pipes
Flow transmitter for liquid flow monitoring in pipes from DN 32, with a measuring range of 0.3.6 m/s and PNP/NPN, NAMUR, 4.20 mA, 0.10 V, frequency or limit…
Learn More