Overview
The ECD CA-6 Nitrite Analyzer is a compact online colorimetric analyzer designed for automatic nitrite measurement in water. It performs sequential sampling and analysis using an LED light source and a heated colorimetric reaction cell. The analyzer determines nitrite concentration through differential photometric absorbance at 525 nm, with measuring ranges from trace µg/L up to 125 mg/L NO₂ using internal dilution.
The analyzer features a dual-compartment enclosure that separates the electronic and liquid-handling components. It incorporates automatic calibration, validation and cleaning functions, programmable analysis frequency and low reagent consumption. A colour touchscreen interface provides access to measured values, operating status and configuration settings, while an integrated data logger supports USB data download. The analyzer is equipped with two 4–20 mA outputs, Modbus RTU RS485 communication and four programmable SPDT alarm relays, with an IP54-rated enclosure.
Typical Applications
Nitrite monitoring in drinking water
Process optimisation in wastewater treatment plants
Nitrite monitoring in industrial and municipal wastewater
Mineral water quality monitoring
Nitrite monitoring in surface water
Key Features
Online sequential colorimetric measurement using differential photometric absorbance at 525 nm
Measuring range from trace µg/L up to 125 mg/L NO₂ with internal dilution
LED light source and heated colorimetric reaction cell
Automatic calibration, validation and cleaning with selectable intervals
Programmable analysis frequency with a 6–8 minute analysis cycle
Colour touchscreen interface with multilingual menu navigation
Integrated data logger with USB data download
Two 4–20 mA outputs and Modbus RTU RS485 communication
Four programmable SPDT potential-free alarm relays
Loss of sample and low reagent alarms
Up to four sample streams with an external sequencer
IP54-rated epoxy powder-coated steel cabinet with separate electronic and liquid-handling compartments
Standard wall mounting with optional benchtop or panel mounting
Benefits
Automated analysis, calibration, validation and cleaning reduce downtime and operator intervention
Low reagent consumption and adjustable analysis frequency help minimise operating costs
Reference measurements minimise interference from sample colour, turbidity and refractive index variations
Wide measuring range accommodates different nitrite concentrations
User-friendly touchscreen simplifies operation and configuration
Integrated data logging enables convenient data retrieval
Optional multiple-stream configurations allow monitoring of different sampling points
Separate compartments isolate electronics from liquid-handling components
Factory-tested design simplifies installation and commissioning
Loss of sample and low reagent alarms alert operators to operating issues
Specifications
Measurement | |
|---|---|
Measured parameter | NO₂ / N-NO₂ |
Measurement units | ppb, ppm, mg/L |
Measuring principle | Differential photometric absorbance (diazotization) |
Measurement wavelength | 525 nm |
Light source | LED |
Measuring range (26 mm cell) | 5–500 ppb N-NO₂ (1.6 ppm NO₂ at upper limit) |
Measuring range (16 mm cell) | 0.02–1 ppm N-NO₂ (3.2 ppm NO₂ at upper limit) |
Measuring range with internal dilution | Up to 40 ppm N-NO₂ / 125 ppm NO₂ |
Reproducibility (up to 150 ppb, 26 mm cell) | ±5 ppb or ±5%, whichever is greater |
Reproducibility (≥150 ppb to 600 ppb) | ±10 ppb |
Reproducibility (≥600 µg/L, 16 mm cell) | ±20 ppb or ±5%, whichever is greater |
Analysis and Operation | |
|---|---|
Analysis frequency | Freely programmable, batch near-continuous analysis |
Cycle time | 6–8 minutes, including conditioning before analysis and rinsing after measurement |
Reaction cell | Heated colorimetric reaction cell |
Calibration | Automatic, with selectable intervals |
Validation | Automatic, with selectable intervals |
Cleaning | Automatic, with selectable intervals |
Sample and Drain Requirements | |
|---|---|
Sample pressure | Pressure-free from overflow vessel |
Sample temperature | 5–50 °C (41–122 °F) |
Sample flow rate | 80–500 mL/min |
Sample connection | 6 mm (¼ in.) |
Drain condition | Pressure-free, atmospheric drain |
Drain connection | 12 mm (½ in.) |
Number of sample streams | 1 or 2 with integrated switching valve; 3 or 4 with external sequencer |
Reagent Consumption | |
|---|---|
Reagent consumption (16 mm cell) | 0.7 L (0.18 US gal) per reagent every 30 days at a 15-minute analysis frequency |
Reagent consumption (26 mm cell) | 1 L (0.26 US gal) per reagent every 30 days at a 15-minute analysis frequency |
Display and Data Management | |
|---|---|
User interface | Colour graphic touchscreen |
Display information | Measured values and status information |
Language support | Multiple languages |
Data logging | Integrated data logger |
Data download | USB |
Outputs, Communication and Alarms | |
|---|---|
Analogue outputs | 2 × 4–20 mA for measured data |
Digital communication | Modbus RTU RS485 |
Alarm relays | 4 programmable SPDT potential-free relays |
Alarm functions | Loss of sample and low reagent |
Digital input | Remote start/stop |
Power Supply | |
|---|---|
Standard power supply | 100–240 VAC, 50/60 Hz |
Optional power supply | 24 VDC |
Power consumption | Maximum 80 VA |
Environmental Conditions | |
|---|---|
Operating temperature | 5–45 °C (41–113 °F) |
Relative humidity | 10–90%, non-condensing |
Operating environment | Indoor use; outdoor installation requires a protective cabinet or shelter (not included) |
Mechanical and Installation | |
|---|---|
Cabinet construction | Epoxy powder-coated cold-rolled steel |
Enclosure design | Separate electronic and liquid-handling compartments |
Cabinet doors | Two lockable doors |
Dimensions (H × W × D) | 606 × 380 × 209 mm / 23.6 × 15.0 × 8.2 in. |
Weight | Approximately 20 kg (44 lbs) |
Ingress protection | IP54 |
Standard installation | Wall mounting |
Optional installation | Benchtop support or panel mounting |
Drawing

Downloads
| CA6 - Nitrite Analyzer Brochure | DownloadCA6 - Nitrite Analyzer Brochure |
|---|
Related Products

ECD
CA6 Total Phosphate Analyzer
The CA-6 Total Phosphate Analyzer provides automated online colorimetric measurement of total phosphate in water, featuring sequential sampling, a touchscreen…
Learn More
ECD
CA6 Nickel Analyzer
The CA-6 Nickel Analyzer uses online colorimetric technology for automated nickel measurement in water, featuring programmable analysis cycles and a…
Learn More
ECD
CA6 Phosphate Analyzer
The CA-6 Phosphate Analyzer provides automatic online phosphate measurement in water using colorimetric analysis, featuring low reagent consumption, automatic…
Learn More
ECD
CA6 Hydrazine Analyzer
The CA-6 Hydrazine Analyzer provides automatic online colorimetric measurement of hydrazine in water, featuring low reagent consumption, automatic…
Learn More