WETEC

ECD

CA900 Sulfide Ion Analyzer


Compact online sequential sampling analyzer for the automatic measurement of total sulfide ions in water.

  • Uses a sulfide ISE with pH adjustment and acid neutralization for safe waste disposal

  • Standard 6-minute analysis cycle with minimal reagent usage (1 ml caustic and 1 ml acid per cycle)

  • Touchscreen interface with 4-20 mA outputs, standard wall mounting, and simple tubing setup

Quantity
1

Overview

The ECD CA900 Sulfide Ion Analyzer is a compact, all-in-one system for continuous, automated sulfide ion measurement in municipal water, wastewater, and industrial process streams. Utilizing a sulfide ion selective electrode (ISE) with integrated pH conditioning and post-measurement acid neutralization, it ensures precise, eco-friendly monitoring across a broad 0.02 to 500 ppm range. Featuring programmable analysis cycles, an intuitive touchscreen interface, flexible 4-20 mA outputs, and configurable alarm relays, the CA900 provides robust, low-maintenance operation for critical discharge and water quality compliance.

Typical Applications

  • Municipal Wastewater (treatment plants and monitoring)

  • Well Water

  • Refineries (petroleum/chemical wastewater streams)

  • Tanneries

  • Chemical Plants

  • Paper and Pulp Facilities

Key Features

  • Compact all-in-one measurement system – Conditions the sample, measures sulfide, neutralizes the sample, measures pH, drains, and rinses the cell

  • ISE based sulfide ion measurement – Provides a wide measurement range from 20 ppb to 500 ppm using economical, replaceable electrode cartridges

  • pH adjusted and compensated measurement – The sample is optimized for the sulfide ion measurement at high pH values.

  • Sample neutralized after measurement – The highly caustic sample pH is reduced to a safer level near pH 8.

Benefits

  • Touchscreen interface – User-friendly menus and configuration screens with separate electronics and hydraulics/liquids compartments.

  • Low reagent consumption – At 10 samples per hour the analyzer uses less than 2 gallons per month of KOH and an equivalent amount of HCl; at 2 samples per hour reagent use is about 5 months per set.

Specifications

MEASUREMENT

 

Measured parameter

Sulfide Ions (S²⁻ / Total Sulfide)

Measuring principle

Sequential sampling, Sulfide Ion Selective Electrode (ISE), pH compensated measurement, sample neutralization

Measuring range

0.02…500 mg/L (0.02…500 ppm)

Measuring temperature range

-5…50 °C (20…120 °F)

Accuracy

±20 ppb or ±5% of reading, whichever is greater

Repeatability

±2% of reading

Cycle response time

6, 12, or 30 minutes, user selectable

 

 

SAMPLE & STREAMS

 

Sample requirements

Pressure: Drawn from atmospheric pressure
Flow rate: >100 mL/min
Temperature: 10…50 °C

Hydraulic connections

Sample Inlet: ⅛ in ID tubing barb fitting
Drain Outlet: ¼ in ID tubing barb fitting

Reagents

16% Potassium Hydroxide (KOH), 7.5% Hydrochloric Acid (HCl / Muriatic Acid)

Reagent consumption

6 min cycle time: 7.2 L/month (1.9 gal) each
12 min cycle time: 3.6 L/month (0.95 gal) each
30 min cycle time: 1.5 L/month (0.4 gal) each

 

 

ELECTRICAL

 

Power requirement

100/240 VAC, 50/60 Hz, switch selectable

Analog output

Four 4…20 mA outputs

Alarms

4 configurable relays SPDT 15A 250VAC

Data logging

Configurable data recording, storage, and output

Connections

2 x 4…20 mA, Line Neutral and Ground for power. All connections are to a terminal strip, access through IP65 ½ in cable glands

 

 

PHYSICAL & ENVIRONMENTAL

 

Dimensions (L x H x D)

43 x 81 x 23 cm (17 x 32 x 9 in)

Weight

Approx. 14 kg (30 lbs)

Operating conditions

Temperature: 10…50 °C (Operating Temperature: 5…50 °C)
Humidity: 5 to 95% non-condensing humidity

Enclosure & Cabinet

Non-metallic gray hot-molded fiberglass reinforced polyester, transparent polycarbonate cover with non-metallic hinges

Mounting

Wall mounting or with optional bench support

Protection degree

NEMA 4X, IP66-11

Drawing

Downloads

ECD-CA900-Sulfide-Ion-Analyzer-I2319.pdf Datasheet (EN) DownloadECD-CA900-Sulfide-Ion-Analyzer-I2319.pdf Datasheet (EN)