Senseca
PH111 Combined pH Electrode for Drinking Water and Clear Media
Combined pH electrode with glass shaft for drinking water and lightly polluted liquids.
Measurement range of 0–14 pH
Gel-filled Ag/AgCl reference system with salt reserve
Single ceramic diaphragm and large ball membrane
Response time of less than 30 s (pH 4–7)
Working temperature of 100 °C and maximum pressure of 6 bar
120 mm installation length and 12 mm diameter
PG13.5 process connection and S8 industrial plug-in head
Overview
The Senseca pH111 is a combined pH electrode designed for pH measurement in drinking water, clear media and lightly polluted liquids. It covers a measurement range of 0–14 pH and features a glass shaft, a gel-filled Ag/AgCl reference system, a single ceramic diaphragm and a large ball membrane. The reference system incorporates a salt reserve to support good electrode service life.
The electrode has a working temperature of 100 °C, a maximum pressure of 6 bar and a response time of less than 30 seconds between pH 4 and 7. It requires a minimum media conductivity of 50 µS/cm and a minimum immersion depth of 25 mm. The electrode has a 120 mm installation length, 12 mm diameter, PG13.5 process connection and S8 industrial plug-in head, with compatible S7/S8 connection cables available as accessories.
Typical Applications
pH monitoring in drinking water
pH measurement in clear liquids
pH measurement in lightly polluted liquids
Key Features
pH measurement range of 0–14 pH
Glass shaft and electrolyte suitable for drinking water applications
Ag/AgCl reference system with gel electrolyte and salt reserve
Single ceramic diaphragm
Large ball membrane
Response time of less than 30 s (pH 4–7)
Working temperature of 100 °C
Maximum pressure of 6 bar
Minimum media conductivity of 50 µS/cm
Minimum immersion depth of 25 mm
120 mm installation length and 12 mm diameter
PG13.5 process connection
S8 industrial plug-in head
Benefits
Versatile pH measurement in drinking water, clear and lightly polluted liquids
Good electrode service life supported by the salt reserve
Fast measurement response for timely pH readings
Compatibility with PG13.5 process fittings for standard installation
Flexible connection options with different S7/S8 cable lengths
Specifications
Measurement |
|
|---|---|
Sensor type |
Combined pH electrode |
Measuring principle |
Potentiometric |
Measurement range |
0–14 pH |
Application scope |
Universal, lightly polluted liquids and drinking water |
Response time |
< 30 s (pH 4–7) |
Minimum conductivity of media |
50 µS/cm |
Temperature sensor |
None |
Electrode Construction |
|
|---|---|
Shaft material |
Glass |
Reference system |
Ag/AgCl |
Reference electrolyte |
3 mol/L KCl gel |
Salt reserve |
Included |
Diaphragm |
1 × ceramic |
Membrane shape |
Large ball |
Operating Conditions |
|
|---|---|
Working temperature |
100 °C |
Maximum pressure |
6 bar |
Mechanical and Installation |
|
|---|---|
Diameter |
12 mm |
Installation length |
120 mm |
Minimum immersion depth |
25 mm |
Process connection |
PG13.5 |
Installation position |
120° |
Electrical Connection |
|
|---|---|
Electrical connection |
S8 industrial plug-in head |
Storage Conditions |
|
|---|---|
Storage temperature |
−5 to +30 °C |
Storage solution |
3 mol/L KCl solution |
Recommended storage duration |
Storage longer than one year is not recommended |
Scope of Delivery |
|
|---|---|
Included accessory |
KCl-filled electrode storage tube with sealing cap |
Model Optional Accessories |
|
|---|---|
S7-02 |
Connection cable with S7/S8 plug and pigtail, 2 m |
S7-05 |
Connection cable with S7/S8 plug and pigtail, 5 m |
S7-10 |
Connection cable with S7/S8 plug and pigtail, 10 m |
EA2200 |
Installation adaptor for PVC-U piping |
DFA32 |
Flow fitting for EA2200 |
EA2630 |
Installation fitting, G¾ or G1A, PVDF |
EA2650 |
Installation fitting, G½A, G¾A or G1A, stainless steel |
EA2730 |
Installation fitting for milk tube piping, PVDF |
EA2090 |
Installation fitting for open basins |
DFG70 |
Flow-through fitting for up to 3 sensors, PC/PVC-U |
Compliance and Standards |
|
|---|---|
European directive |
2011/65/EU (RoHS) |
Harmonised standard |
EN IEC 63000:2018 |
Downloads
| PH111 Family Datasheet | DownloadPH111 Family Datasheet |
|---|---|
| PH-Redox-Electrodes Family Operating Instructions | DownloadPH-Redox-Electrodes Family Operating Instructions |
| PH111 Family CE Declaration of Conformity | DownloadPH111 Family CE Declaration of Conformity |
Related Products

Senseca
DLF4 Data Logger for Autonomous Sensor Data Transmission
Autonomous data logger for up to 7 sensors with analog, RS485 Modbus RTU and digital inputs.
Learn More
Senseca
HLF4 Autonomous Data Logger for Water Level Monitoring in 2" Pipes
Autonomous data logger for water level monitoring in pipes from 2".
Learn More
Senseca
PH101 Combined pH Electrode for Clear and Lightly Polluted Liquids
Combined pH electrode for clear to lightly polluted liquids.
Learn MoreSenseca
UNICON-LF-01/02/03 Conductivity Converter
Conductivity converter for liquid purity and concentration monitoring.
Learn More